CLASS II ONGOING Device recall · Reported 21 Oct 2020 · FDA Enforcement Report

Advia Chemistry Systems-Fructosamine (Fruc) recalled

Siemens Healthcare Diagnostics is recalling Advia Chemistry Systems-Fructosamine (Fruc)-For in vitro diagnostic use in the, distributed in 7 states. The recall was initiated on 22 Sep 2020 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesAll lots UDI: (01)00630414558073(10)501874(17)20201128 (01)00630414558073(10)514594(17)20210221 (01)00630414558073(10)528971(17)20210630
Quantity recalledUS: 23 units; OUS: 204 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0101-2021
FDA event ID
86504
Recalling firm
Siemens Healthcare Diagnostics, Tarrytown, NY
Classification
Class II
Status
Ongoing
Recall initiated
22 Sep 2020
FDA classified
13 Oct 2020
Reported
21 Oct 2020
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Potency & Assay

Timeline

22 Sep 2020Recall initiated by company
13 Oct 2020Classified by FDA+21 days
21 Oct 2020Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Fructosamine QC and patient samples may exhibit a 60 umol/L positive bias across the measuring range compared to predicate assay reference leading to a potential adjustment of hyperglycemic therapy. Positively biased fructosamine results may prompt changes in glucose management that may affect glycemia. In extreme situations, more aggressive management of glycemia may occur and contribute to hypoglycemia.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

ADVIA Chemistry Systems-Fructosamine (FRUC)-For in vitro diagnostic use in the quantitative measurement of glycated protein (fructosamine) in human serum or plasma on the ADVIA Chemistry systems Siemens Material Number (SMN): 10361941

Where it was distributed

CA, DE, MD, NC, NJ, NY, PA Foreign: Argentina¿¿¿¿¿¿ Australia¿¿¿¿¿¿ Brazil¿¿¿¿ Canada¿¿¿¿¿ Czech¿Republic¿ Denmark¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ Netherlands¿¿¿¿ P.R.¿China¿¿¿¿¿ Paraguay¿¿¿¿¿¿¿ Slovakia¿¿¿¿¿¿¿ Turkey¿¿¿¿¿¿¿¿¿ United¿Kingdom¿ Vietnam¿¿¿¿¿¿¿¿

CaliforniaDelawareMarylandNorth CarolinaNew JerseyNew YorkPennsylvania

Questions about this recall

Why was it recalled?

Fructosamine QC and patient samples may exhibit a 60 umol/L positive bias across the measuring range compared to predicate assay reference leading to a potential adjustment of hyperglycemic therapy. Positively biased fructosamine results may prompt changes in glucose management that may affect glycemia. In extreme situations, more aggressive management of glycemia may occur and contribute to hypoglycemia.

Which lots are affected?

All lots UDI: (01)00630414558073(10)501874(17)20201128 (01)00630414558073(10)514594(17)20210221 (01)00630414558073(10)528971(17)20210630

Where was it sold?

CA, DE, MD, NC, NJ, NY, PA Foreign: Argentina¿¿¿¿¿¿ Australia¿¿¿¿¿¿ Brazil¿¿¿¿ Canada¿¿¿¿¿ Czech¿Republic¿ Denmark¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ Netherlands¿¿¿¿ P.R.¿China¿¿¿¿¿ Paraguay¿¿¿¿¿¿¿ Slovakia¿¿¿¿¿¿¿ Turkey¿¿¿¿¿¿¿¿¿ United¿Kingdom¿ Vietnam¿¿¿¿¿¿¿¿

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 86504
ReportedClassRecallCompanyReasonWhere
CLASS IIAtellica CH Systems Fructosamine (FRUC) recalledSiemens Healthcare DiagnosticsPotency & Assay7 states

More recalls from Siemens Healthcare Diagnostics

All 694

Other Test Kits & Diagnostics recalls

All 2,656