CLASS II ONGOING Device recall · Reported 26 Aug 2026 · FDA Enforcement Report

DxC 500i Clinical Analyzer Modules using recalled by Beckman Coulter

Beckman Coulter is recalling Beckman Coulter DxC 500i Clinical Analyzer Modules using, distributed nationwide in the US. The recall was initiated on 15 Jul 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codes1) Part Number C63522, UDI/DI 14987666545089; 2) Part Number C63521, UDI/DI 14987666545072; 3) Part Number C13252, UDI/DI 15099590742331.

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2973-2026
FDA event ID
99436
Recalling firm
Beckman Coulter, Brea, CA
Classification
Class II
Status
Ongoing
Recall initiated
15 Jul 2026
FDA classified
17 Aug 2026
Reported
26 Aug 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Packaging Defects

Timeline

15 Jul 2026Recall initiated by company
17 Aug 2026Classified by FDA+33 days
26 Aug 2026Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU Module w/ISE, DxC 500i, Part Number C63522; 2) DxC 500 AU Module, DxC 500i, Part Number C63521; 3) Access 2 Module, DxC 500i, Part Number C13252; DxC 500 AU Clinical Chemistry Analyzers

Where it was distributed

Worldwide distribution - US Nationwide.

Nationwide

Questions about this recall

Is DxC 500i Clinical Analyzer Modules using recalled?

Yes. Beckman Coulter recalled DxC 500i Clinical Analyzer Modules using on 15 Jul 2026. The recall is Class II and its status is Ongoing. Recall number Z-2973-2026.

Why was it recalled?

a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.

Which lots are affected?

1) Part Number C63522, UDI/DI 14987666545089; 2) Part Number C63521, UDI/DI 14987666545072; 3) Part Number C13252, UDI/DI 15099590742331.

Where was it sold?

Worldwide distribution - US Nationwide.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Beckman Coulter

All 334

Other Test Kits & Diagnostics recalls

All 2,656