CLASS II Device recall · Reported 16 May 2018 · FDA Enforcement Report
Advia Salicylate (SAL) reagent chemistry assay recalled
Siemens Healthcare Diagnostics is recalling Advia Salicylate (SAL) reagent chemistry assay, distributed in District of Columbia, Georgia, Puerto Rico. The recall was initiated on 29 Jun 2017 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1648-2018
- FDA event ID
- 79225
- Recalling firm
- Siemens Healthcare Diagnostics, Tarrytown, NY
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 29 Jun 2017
- FDA classified
- 4 May 2018
- Reported
- 16 May 2018
- Type
- Voluntary: Firm initiated
- Customers notified by
- Reason category
- Potency & Assay
- Device type
- Test Kits & Diagnostics
Timeline
| 29 Jun 2017 | Recall initiated by company | |
| 4 May 2018 | Classified by FDA | +309 days |
| 16 May 2018 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine causing falsely elevated or falsely depressed results.
Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.
Product as listed by the FDA
ADVIA Salicylate (SAL) reagent chemistry assay; use in the quantitative determination of salicylate in human serum and plasma (lithium heparin). Product Number: 10327382 (Mfr: Sekisui Diagnostics)
Where it was distributed
Distributed in all 50 U.S. states, the District of Columbia, Guam, and Puerto Rico. Distributed in 108 foreign countries: Afghanistan Bolivia Albania¿¿¿¿¿¿¿¿ Algeria¿¿¿¿¿¿¿¿ Angola ARGENTINA Armenia AUSTRALIA Austria¿¿¿¿¿¿¿¿ Azerbaijan Bahamas Bahrain¿¿¿¿¿¿¿¿ Bangladesh¿¿¿¿¿ Belarus¿¿¿¿¿¿¿¿ Belgium¿¿¿¿¿¿¿¿ Bosnia¿Herzeg.¿ Brazil Bulgaria¿¿¿¿¿¿¿ Cambodia Canada Canary islands Chile China Colombia Costa Rica Croatia¿¿¿¿¿¿¿¿ Cyprus¿¿¿¿¿¿¿¿¿ Czech¿Republic¿ Denmark¿¿¿¿¿¿¿¿ Ecuador Egypt¿¿¿¿¿¿¿¿¿¿ El Salavador Estonia¿¿¿¿¿¿¿¿ Finland¿¿¿¿¿¿¿¿ France¿¿¿¿¿¿¿¿¿ Georgia Germany¿¿¿¿¿¿¿¿ Greece¿¿¿¿¿¿¿¿¿ Guatemala Honduras Hong Kong Hungary¿¿¿¿¿¿¿¿ Iceland India Indonesia Iraq Ireland¿¿¿¿¿¿¿¿ Israel¿¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ japan Jordan Kazakhstan¿¿¿¿¿ Korea Kuwait Latvia¿¿¿¿¿¿¿¿¿ Lebanon Lithuania¿¿¿¿¿¿ Luxembourg¿¿¿¿¿ Macedonia Malaysia Mali Malta¿¿¿¿¿¿¿¿¿¿ Mexico Morocco¿¿¿¿¿¿¿¿ Myanmar Nepal Netherlands¿¿¿¿ New Zealand Nicaragua Norway¿¿¿¿¿¿¿¿¿ Oman Pakistan¿¿¿¿¿¿¿ Paraguay Peru Philippines Poland¿¿¿¿¿¿¿¿¿ Portugal¿¿¿¿¿¿¿ Qatar¿¿¿¿¿¿¿¿¿¿ Romania¿¿¿¿¿¿¿¿ Russian¿Fed.¿¿¿ Saudi Arabia Senegal Serbia¿¿¿¿¿¿¿¿¿ Serbia¿¿¿¿¿¿¿¿¿ Singapore Slovakia¿¿¿¿¿¿¿ Slovenia¿¿¿¿¿¿¿ South Africa South Africa South Korea Spain¿¿¿¿¿¿¿¿¿¿ Sri Lanka Sweden¿¿¿¿¿¿¿¿¿ Switzerland¿¿¿¿ Taiwan¿¿¿¿¿¿¿¿¿ Tajikistan Thailand Tunisia¿¿¿¿¿¿¿¿ Turkey¿¿¿¿¿¿¿¿¿ Turkmenistan¿¿¿ United Arab Emerites¿¿¿¿¿¿ United¿Kingdom¿ Uruguay¿¿¿¿¿¿¿¿ Uzbekistan Vatikancity¿¿¿¿ Venezuela Vietnam Yemen
Questions about this recall
Is Advia Salicylate (SAL) reagent chemistry assay recalled?
Yes. Siemens Healthcare Diagnostics recalled Advia Salicylate (SAL) reagent chemistry assay on 29 Jun 2017. The recall is Class II and its status is Terminated. Recall number Z-1648-2018.
Why was it recalled?
Various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine causing falsely elevated or falsely depressed results.
Which lots are affected?
All lots
Where was it sold?
Distributed in all 50 U.S. states, the District of Columbia, Guam, and Puerto Rico. Distributed in 108 foreign countries: Afghanistan Bolivia Albania¿¿¿¿¿¿¿¿ Algeria¿¿¿¿¿¿¿¿ Angola ARGENTINA Armenia AUSTRALIA Austria¿¿¿¿¿¿¿¿ Azerbaijan Bahamas Bahrain¿¿¿¿¿¿¿¿ Bangladesh¿¿¿¿¿ Belarus¿¿¿¿¿¿¿¿ Belgium¿¿¿¿¿¿¿¿ Bosnia¿Herzeg.¿ Brazil Bulgaria¿¿¿¿¿¿¿ Cambodia Canada Canary islands Chile China Colombia Costa Rica Croatia¿¿¿¿¿¿¿¿ Cyprus¿¿¿¿¿¿¿¿¿ Czech¿Republic¿ Denmark¿¿¿¿¿¿¿¿ Ecuador Egypt¿¿¿¿¿¿¿¿¿¿ El Salavador Estonia¿¿¿¿¿¿¿¿ Finland¿¿¿¿¿¿¿¿ France¿¿¿¿¿¿¿¿¿ Georgia Germany¿¿¿¿¿¿¿¿ Greece¿¿¿¿¿¿¿¿¿ Guatemala Honduras Hong Kong Hungary¿¿¿¿¿¿¿¿ Iceland India Indonesia Iraq Ireland¿¿¿¿¿¿¿¿ Israel¿¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ japan Jordan Kazakhstan¿¿¿¿¿ Korea Kuwait Latvia¿¿¿¿¿¿¿¿¿ Lebanon Lithuania¿¿¿¿¿¿ Luxembourg¿¿¿¿¿ Macedonia Malaysia Mali Malta¿¿¿¿¿¿¿¿¿¿ Mexico Morocco¿¿¿¿¿¿¿¿ Myanmar Nepal Netherlands¿¿¿¿ New Zealand Nicaragua Norway¿¿¿¿¿¿¿¿¿ Oman Pakistan¿¿¿¿¿¿¿ Paraguay Peru Philippines Poland¿¿¿¿¿¿¿¿¿ Portugal¿¿¿¿¿¿¿ Qatar¿¿¿¿¿¿¿¿¿¿ Romania¿¿¿¿¿¿¿¿ Russian¿Fed.¿¿¿ Saudi Arabia Senegal Serbia¿¿¿¿¿¿¿¿¿ Serbia¿¿¿¿¿¿¿¿¿ Singapore Slovakia¿¿¿¿¿¿¿ Slovenia¿¿¿¿¿¿¿ South Africa South Africa South Korea Spain¿¿¿¿¿¿¿¿¿¿ Sri Lanka Sweden¿¿¿¿¿¿¿¿¿ Switzerland¿¿¿¿ Taiwan¿¿¿¿¿¿¿¿¿ Tajikistan Thailand Tunisia¿¿¿¿¿¿¿¿ Turkey¿¿¿¿¿¿¿¿¿ Turkmenistan¿¿¿ United Arab Emerites¿¿¿¿¿¿ United¿Kingdom¿ Uruguay¿¿¿¿¿¿¿¿ Uzbekistan Vatikancity¿¿¿¿ Venezuela Vietnam Yemen
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Siemens Healthcare Diagnostics
All 694| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Syva EMIT 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Beckman Coulter, Emit 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Lead & Heavy Metals | Nationwide |
| Device | CLASS III | Syva Emit Disopyramide Assay Enzyme Immunoassay recalled | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide |
| Device | CLASS II | Atellica CH Urine Albumin recalled by Siemens Healthcare Diagnostics | Siemens Healthcare Diagnostics | Lead & Heavy Metals | Nationwide |
| Device | CLASS III | Syva Emit tox Serum Tricyclic Antidepressant recalled | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide |
Other Test Kits & Diagnostics recalls
All 2,656| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | STA Staclot Protein S recalled by Diagnostica Stago | Diagnostica Stago | Potency & Assay | Nationwide |
| Device | CLASS II | DxC 500i Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | DxC 500 AU Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | Syva EMIT 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Beckman Coulter, Emit 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Lead & Heavy Metals | Nationwide |