CLASS II ONGOING Device recall · Reported 19 Jun 2024 · FDA Enforcement Report

Aeos Robotic Digital Microscope, Product recalled by Aesculap

Aesculap is recalling Aeos Robotic Digital Microscope, distributed nationwide in the US. The recall was initiated on 13 May 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesProduct Code: PV010, UDI/DI: 04046955206742; Serial Numbers: 1011, 1005, 1057, 1032, 1038, 1054, 1036, 1060, 1056, 1089, 1093, 1027, 1061, 1004, 1026, 1035, 1097, 1084, 1063
Quantity recalled19 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2139-2024
FDA event ID
94623
Recalling firm
Aesculap, Center Valley, PA
Classification
Class II
Status
Ongoing
Recall initiated
13 May 2024
FDA classified
12 Jun 2024
Reported
19 Jun 2024
Type
Voluntary: Firm initiated
Customers notified by
N/A
Reason category
Device Malfunction

Timeline

13 May 2024Recall initiated by company
12 Jun 2024Classified by FDA+30 days
19 Jun 2024Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Possibility of the robotic arm to drop more than 10cm from its original position after pressing the emergency stop button due to failure of the integrated safety mechanism.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Aeos Robotic Digital Microscope, Product Code: PV010

Where it was distributed

US Nationwide distribution in the states of AR, FL, IL, IN, KY, MI, NC, NE, OK, SD, TX.

Nationwide ArkansasFloridaIllinoisIndianaKentuckyMichiganNorth CarolinaNebraskaOklahomaSouth DakotaTexas

Questions about this recall

Why was it recalled?

Possibility of the robotic arm to drop more than 10cm from its original position after pressing the emergency stop button due to failure of the integrated safety mechanism.

Which lots are affected?

Product Code: PV010, UDI/DI: 04046955206742; Serial Numbers: 1011, 1005, 1057, 1032, 1038, 1054, 1036, 1060, 1056, 1089, 1093, 1027, 1061, 1004, 1026, 1035, 1097, 1084, 1063

Where was it sold?

US Nationwide distribution in the states of AR, FL, IL, IN, KY, MI, NC, NE, OK, SD, TX.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Aesculap

All 42

Other Other Medical Devices recalls

All 13,908