CLASS II Device recall · Reported 21 Nov 2018 · FDA Enforcement Report

AIA-900 Analyzer, Product recalled by Tosoh Bioscience

Tosoh Bioscience is recalling AIA-900 Analyzer, distributed nationwide in the US. The recall was initiated on 5 Oct 2018 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesAll units are affected
Quantity recalled393

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0469-2019
FDA event ID
81187
Recalling firm
Tosoh Bioscience, Grove City, OH
Classification
Class II
Status
Terminated
Recall initiated
5 Oct 2018
FDA classified
14 Nov 2018
Reported
21 Nov 2018
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

5 Oct 2018Recall initiated by company
14 Nov 2018Classified by FDA+40 days
21 Nov 2018Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Tosoh Bioscience has become aware of a potential issue with the pickup arm and pickup arm motor on the AIA-900 Analyzer. If the pickup arm or pickup arm motor malfunctions, the analyzer will cease to function and result in an inability to test patient samples. This may cause a delay in testing and reporting patient results. Tosoh has received fifty-one (51) complaints related to this issue with no serious injuries reported.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

AIA-900 Analyzer, Product code 022930, 022930R

Where it was distributed

Distributed nationwide. Foreign distribution to Canada and Latin America.

Nationwide

Questions about this recall

Why was it recalled?

Tosoh Bioscience has become aware of a potential issue with the pickup arm and pickup arm motor on the AIA-900 Analyzer. If the pickup arm or pickup arm motor malfunctions, the analyzer will cease to function and result in an inability to test patient samples. This may cause a delay in testing and reporting patient results. Tosoh has received fifty-one (51) complaints related to this issue with no serious injuries reported.

Which lots are affected?

All units are affected

Where was it sold?

Distributed nationwide. Foreign distribution to Canada and Latin America.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Tosoh Bioscience

All 74

Other Test Kits & Diagnostics recalls

All 2,663