CLASS II Device recall · Reported 19 Dec 2018 · FDA Enforcement Report
AIA-900 Automated Immunoassay Analyzer performs three recalled
Tosoh Bioscience is recalling AIA-900 Automated Immunoassay Analyzer performs three methods of immunoassay: an, distributed nationwide in the US. The recall was initiated on 28 Sep 2017 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0602-2019
- FDA event ID
- 81172
- Recalling firm
- Tosoh Bioscience, Grove City, OH
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 28 Sep 2017
- FDA classified
- 13 Dec 2018
- Reported
- 19 Dec 2018
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Test Kits & Diagnostics
Timeline
| 28 Sep 2017 | Recall initiated by company | |
| 13 Dec 2018 | Classified by FDA | +441 days |
| 19 Dec 2018 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The firm has become aware of a potential issue that may affect the mixing motor drive harness and mixing home detection sensor relay harness on your AIA-900 Analyzer. It is possible that the harnesses could rub against the Plarail chain during operation and result in premature wear. If observed, this issue will cause the following system error to be reported D.lane Mix home overrun error [4093] . If observed, this issue will cause the AIA-900 Analyzer not to function and the D.lane Mix home overrun error [4093] will be reported by the system. Replacement of the motor drive harness and mixing home detection sensor relay harness will mitigate the potential for premature wear and effectively prevent this issue from causing the system error. If the sensor relay harness fails, the AIA-900 analyzer will cease to function and will be unavailable for use. If this issue occurs, there may be a delay in testing and reporting patient test results. Tosoh is aware of two complaints associated with the unexpected wearing of the harness. There have been no user or patient injuries reported that are related to this device malfunction.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
AIA-900 Automated Immunoassay Analyzer performs three methods of immunoassay: an immunoenzymetric (IEMA) or sandwich immunoassay, a competitive binding (EIA) immunoassay, and a two-step immunoenzymetric immunoassay.
Where it was distributed
Nationwide distribution. Foreign distribution to Chile, Colombia, Ecuador, Guatemala, Honduras, Panama, Peru, Puerto Rico, Uruguay, and Venezuela.
Questions about this recall
Why was it recalled?
The firm has become aware of a potential issue that may affect the mixing motor drive harness and mixing home detection sensor relay harness on your AIA-900 Analyzer. It is possible that the harnesses could rub against the Plarail chain during operation and result in premature wear. If observed, this issue will cause the following system error to be reported D.lane Mix home overrun error [4093] . If observed, this issue will cause the AIA-900 Analyzer not to function and the D.lane Mix home overrun error [4093] will be reported by the system. Replacement of the motor drive harness and mixing home detection sensor relay harness will mitigate the potential for premature wear and effectively prevent this issue from causing the system error. If the sensor relay harness fails, the AIA-900 analyzer will cease to function and will be unavailable for use. If this issue occurs, there may be a delay in testing and reporting patient test results. Tosoh is aware of two complaints associated with the unexpected wearing of the harness. There have been no user or patient injuries reported that are related to this device malfunction.
Which lots are affected?
All active analyzer serial numbers
Where was it sold?
Nationwide distribution. Foreign distribution to Chile, Colombia, Ecuador, Guatemala, Honduras, Panama, Peru, Puerto Rico, Uruguay, and Venezuela.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Tosoh Bioscience
All 74| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | ST Aia-Pack Testosterone, IN Vitro Diagnostic Use Only recalled | Tosoh Bioscience | Potency & Assay | Nationwide |
| Device | CLASS II | Thyroglobulin Antibody (TgAB) Calibrator recalled by Tosoh Bioscience | Tosoh Bioscience | Potency & Assay | Nationwide |
| Device | CLASS II | Aia- Pack Hcg Calibrator Set-In Vitro Diagnostic for Bhcg recalled | Tosoh Bioscience | Potency & Assay | Nationwide |
| Device | CLASS II | Aia-pack Bhcg Calibration Verification Test Set-In Vitro recalled | Tosoh Bioscience | Potency & Assay | Nationwide |
| Device | CLASS II | AIA-900 Analyzer recalled by Tosoh Bioscience | Tosoh Bioscience | Potency & Assay | Nationwide |
Other Test Kits & Diagnostics recalls
All 2,663| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | DxC 500 AU Clinical Chemistry Analyzers: AU w/out ISE recalled | Beckman Coulter | Device Malfunction | Nationwide |
| Device | CLASS II | DxC 500i Clinical Analyzer Modules: DxC 500 AU Module w/ISE recalled | Beckman Coulter | Device Malfunction | Nationwide |
| Device | CLASS II | McKesson Consult hCG Urine Cassette recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |
| Device | CLASS II | Systems/units containing affected recalled | Beckman Coulter Mishima K.K | Device Malfunction | Nationwide |
| Device | CLASS II | Fisher Scientific Sure-Vue Urine hCG recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |