CLASS I ONGOING Device recall · Reported 6 May 2026 · FDA Enforcement Report

AIDBAGs are first aid kits of convenience composed recalled

North American Rescue is recalling AIDBAGs are first aid kits of convenience composed of individually labeled devices, distributed in California, Washington. The recall was initiated on 27 Mar 2026 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesKit REF/UDI-DI/Lot: 85-4661/00842209132089/85-4661122024; 85-4742/00842209133536/85-4742123024. Meter: UDI-DI: 20612479197217, all lots.
Quantity recalled21

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1947-2026
FDA event ID
98697
Recalling firm
North American Rescue, Greer, SC
Classification
Class I
Status
Ongoing
Recall initiated
27 Mar 2026
FDA classified
29 Apr 2026
Reported
6 May 2026
Type
Voluntary: Firm initiated
Reason category
Device Malfunction

Timeline

27 Mar 2026Recall initiated by company
29 Apr 2026Classified by FDA+33 days
6 May 2026Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Kits contain the McKesson TRUE Metrix PRO Professional Monitoring Blood Glucose System that has been recalled for an issue with the software design of the E-5 Error Code where the meter displays an E-5 error code for a very high blood glucose event (> 600 mg/dL) or when there is a test strip error, which can lead to operator confusion when evaluating patients and a delay in appropriate treatment. This is a downstream recall of Trividia Health Class I, Recall Event ID 98317.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

AIDBAGs are first aid kits of convenience composed of individually labeled devices designed as specialty first aid kits designed for the appropriate level provider to evaluate and treat patients at the point-of-injury / point-of-illness. The original manufacturer labeling, including single use indications, is maintained for each device, which may be used independently and at different times. These specific kits contain the McKesson TRUE Metrix PRO Professional Monitoring Blood Glucose System that were recalled as part of RES 98317 with (510(k): K140100).

Where it was distributed

US: CA, WA

CaliforniaWashington

Questions about this recall

Why was it recalled?

Kits contain the McKesson TRUE Metrix PRO Professional Monitoring Blood Glucose System that has been recalled for an issue with the software design of the E-5 Error Code where the meter displays an E-5 error code for a very high blood glucose event (> 600 mg/dL) or when there is a test strip error, which can lead to operator confusion when evaluating patients and a delay in appropriate treatment. This is a downstream recall of Trividia Health Class I, Recall Event ID 98317.

Which lots are affected?

Kit REF/UDI-DI/Lot: 85-4661/00842209132089/85-4661122024; 85-4742/00842209133536/85-4742123024. Meter: UDI-DI: 20612479197217, all lots.

Where was it sold?

US: CA, WA

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.