CLASS I Device recall · Reported 11 Jul 2012 · FDA Enforcement Report

AirLife Isothermal Breathing Circuit recalled by Carefusion 2200

Carefusion 2200 is recalling AirLife Isothermal Breathing Circuit, Infant Respiratory Circuit, Heated, distributed nationwide in the US. The recall was initiated on 29 May 2012 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesproduct codes 10127-4H1, 10331N-4S2, 10351-4H2, 10387-4S2, 10555-4S2, 10706-4S2, 10790-4S2, 10814-4S2, 10849-4S2, 1175-4H2, 1208-4H1, 1553-4H1, 1815-409, 1998-4H1, 2119-4S2G, 2120-4S2G, 4319-4H2, 4326-4H2, 4652-5014677-4H2, 4875-4H2, 5665-4H2, 5682-409, 5732-4H1, 5868-4H2, 6002-4H2, 6040-501, 6074-4H1, 6198-4H2, 6274-H12, 6301-H12, 6313-501, 6459-4H2; 6485-4H2, 6588-4H2, 6942-4H2, 7100-4S2, 7147-4S2, 7156-4S2, 7184-4S2, 7206-4S2, 7213-4S2, 7354-4S2, 7395-4S2, 7407-4S2, 7441-4S2, 7483-4S2, 7526-4S2, 7575-4S2, 7582-4S2, 7588-4S2, 9053-4S2, 9054-4S2, 9055-4S2, 9056-4S2, 9058-4S2, RC41-12027, RC41-18039, RC41-18052, RC41-18060, RC51-12074, RC51-18004, RT4851-00, RT4851-12 and RT4851-18; includes all lot numbers of the affected product codes manufactured prior to February 3, 2012, lot #0000414257 and any lot numbers beginning with the letter Y.
Quantity recalledN/A

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1899-2012
FDA event ID
61943
Recalling firm
Carefusion 2200, McGaw Park, IL
Classification
Class I
Status
Terminated
Recall initiated
29 May 2012
FDA classified
29 Jun 2012
Reported
11 Jul 2012
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Packaging Defects

Timeline

29 May 2012Recall initiated by company
29 Jun 2012Classified by FDA+31 days
11 Jul 2012Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The Y adapter within the breathing circuit may spontaneously crack, causing a leak in the closed ventilation system. This leak could lead to a decrease in the tidal volume delivered to the ventilated patient.

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

AirLife Isothermal Breathing Circuit, Infant Respiratory Circuit, Heated; an Rx device for single patient use; CareFusion, 22745 Savi Ranch Parkway, Yorba Linda, CA 92887 USA, Made in Mexico Respiratory breathing circuits are used with a ventilator when mechanical ventilation is administered to a patient.

Where it was distributed

Worldwide distribution: USA (nationwide) including Puerto Rico and internationally to: Canada, Egypt, France, Germany, Great Britain, Italy, Kuwait, Mexico, Portugal, Saudi Arabia, Spain, Taiwan and United Arab Emirates.

Nationwide Puerto Rico

Questions about this recall

Why was it recalled?

The Y adapter within the breathing circuit may spontaneously crack, causing a leak in the closed ventilation system. This leak could lead to a decrease in the tidal volume delivered to the ventilated patient.

Which lots are affected?

product codes 10127-4H1, 10331N-4S2, 10351-4H2, 10387-4S2, 10555-4S2, 10706-4S2, 10790-4S2, 10814-4S2, 10849-4S2, 1175-4H2, 1208-4H1, 1553-4H1, 1815-409, 1998-4H1, 2119-4S2G, 2120-4S2G, 4319-4H2, 4326-4H2, 4652-5014677-4H2, 4875-4H2, 5665-4H2, 5682-409, 5732-4H1, 5868-4H2, 6002-4H2, 6040-501, 6074-4H1, 6198-4H2, 6274-H12, 6301-H12, 6313-501, 6459-4H2; 6485-4H2, 6588-4H2, 6942-4H2, 7100-4S2, 7147-4S2, 7156-4S2, 7184-4S2, 7206-4S2, 7213-4S2, 7354-4S2, 7395-4S2, 7407-4S2, 7441-4S2, 7483-4S2, 7526-4S2, 7575-4S2, 7582-4S2, 7588-4S2, 9053-4S2, 9054-4S2, 9055-4S2, 9056-4S2, 9058-4S2, RC41-12027, RC41-18039, RC41-18052, RC41-18060, RC51-12074, RC51-18004, RT4851-00, RT4851-12 and RT4851-18; includes all lot numbers of the affected product codes manufactured prior to February 3, 2012, lot #0000414257 and any lot numbers beginning with the letter Y.

Where was it sold?

Worldwide distribution: USA (nationwide) including Puerto Rico and internationally to: Canada, Egypt, France, Germany, Great Britain, Italy, Kuwait, Mexico, Portugal, Saudi Arabia, Spain, Taiwan and United Arab Emirates.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

2 other products · FDA event 61943

This product is part of a recall event; the main page for the event is AirLife Isothermal Breathing Circuit Accessory recalled.

More recalls from Carefusion 2200

All 112

Other CPAP & Ventilators recalls

All 552