CLASS II ONGOING Device recall · Reported 23 Sep 2026 · FDA Enforcement Report

Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system recalled

Datex-Ohmeda is recalling Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, distributed nationwide in the US. The recall was initiated on 30 Jul 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSerial Numbers: ANAS01357 ANAP00609 ANAQ00988 ANAS00586 ANAR00543 ANAP01088 ANAM00443
Quantity recalled7 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-3191-2026
FDA event ID
99649
Recalling firm
Datex-Ohmeda, Madison, WI
Classification
Class II
Status
Ongoing
Recall initiated
30 Jul 2026
FDA classified
16 Sep 2026
Reported
23 Sep 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

30 Jul 2026Recall initiated by company
16 Sep 2026Classified by FDA+48 days
23 Sep 2026Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000

Where it was distributed

Worldwide - US Nationwide distribution.

Nationwide

Questions about this recall

Why was it recalled?

Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.

Which lots are affected?

Serial Numbers: ANAS01357 ANAP00609 ANAQ00988 ANAS00586 ANAR00543 ANAP01088 ANAM00443

Where was it sold?

Worldwide - US Nationwide distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.