CLASS II ONGOING Device recall · Reported 18 Jun 2025 · FDA Enforcement Report
Allon 2001. Part Number recalled by Belmont Instrument
Belmont Instrument is recalling Allon 2001. Part Number: 200-00263, distributed nationwide in the US. The recall was initiated on 13 May 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1966-2025
- FDA event ID
- 96854
- Recalling firm
- Belmont Instrument, Billerica, MA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 13 May 2025
- FDA classified
- 12 Jun 2025
- Reported
- 18 Jun 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Temperature Abuse
- Device type
- Other Medical Devices
Timeline
| 13 May 2025 | Recall initiated by company | |
| 12 Jun 2025 | Classified by FDA | +30 days |
| 18 Jun 2025 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
An investigation into a report of the patient wrap (ThermoWrap), used with the Allon device, exceeding the set temperature determined that if: 1. The Human Sensor Board is disconnected from the Controller Board in the Allon device AND the temperature alarms are disabled, or 2. The Water In and Water Out and Thermostat sensors are disconnected from the Human Sensor Board AND the temperature alarms are disabled, the device will proceed to heat the water above the hardware fault limit without notifying the user of the faulty condition.
Foods that need refrigeration and were stored or shipped too warm can grow harmful bacteria without any change in look or smell.
Product as listed by the FDA
Allon 2001. Part Number: 200-00263
Where it was distributed
Worldwide - US Nationwide distribution in the states of CA, DC, LA, MA, NM, PA, TX, VA, WA and the countries of Belgium, Bolivia (Plurinational State of), Brazil, Canada, Colombia, Israel, Italy, Portugal, Romania, Sweden, Taiwan (Province of China), Thailand, United Kingdom.
Nationwide CaliforniaDistrict of ColumbiaLouisianaMassachusettsNew MexicoPennsylvaniaTexasVirginiaWashington
Questions about this recall
Is Allon 2001. Part Number: 200-00263 recalled?
Yes. Belmont Instrument recalled Allon 2001. Part Number: 200-00263 on 13 May 2025. The recall is Class II and its status is Ongoing. Recall number Z-1966-2025.
Why was it recalled?
An investigation into a report of the patient wrap (ThermoWrap), used with the Allon device, exceeding the set temperature determined that if: 1. The Human Sensor Board is disconnected from the Controller Board in the Allon device AND the temperature alarms are disabled, or 2. The Water In and Water Out and Thermostat sensors are disconnected from the Human Sensor Board AND the temperature alarms are disabled, the device will proceed to heat the water above the hardware fault limit without notifying the user of the faulty condition.
Which lots are affected?
Part Number: 200-00263. UDI-DI: 007290107581103, Serial Numbers: 99111143000002, 99111143000014, 99111627300034, 9993901600001, 9993901600002, 9993901600004, 9993901600006, 9993901600007, 9993901600009, 9995356300005, 9995356300006, 9995356300007, 9995356300008, 9996673800002, 99111143000001, 99111143000003, 99111143000004, 99111143000007, 99111143000008, 99111143000009, 99111143000011, 99111143000012, 99111143000013, 99111143000015, 99111143000018, 99111143000019, 99111288700016, 99111288700021, 99111288700035, 99111288700036, 99111627300011, 99111627300017, 99111627300022, 99111627300029, 99111627300040, 99111627300041, 99111627300045, 99111627300046, 99111627300054, 99111627300057, 99111627300064, 99111627300067, 99113168300004, 99113168300005, 99113168300008, 99113168300010, 99113168300012, 99128623400008, 99129614700002, 99129614700005, 99131046400001, 99131046400003, 99131046400004, 99131046400005, 99131046400006, 99131046400009, 99131046400011, 99131046500007, 99132732400009, 99132732400018, 99132732400019, 99134232100011, 99132121000010, 99113168300003, 9983349800008, 9991451200002, 9991451200004, 9992175000001, 9995356300010, 99111627300001, 99111627300008, 99113168300013, 99132121000014, 99132732400005, 99134232100001, 99134232100002, 99134232100003, 99134232100009, 99134232100022, 99134232100035, 99111627300042, 99111627300056, 99111627300058, 99111627300059, 99113168300014, 99111627300020, 99111627300049, 99113168300006, 99128623400003, 99132121000013, 99134232100018, 9991451200001, 99111143000006, 99111627300005, 99111627300035, 99113168300001, 99113168300002, 99113168300007, 99111627300053, 99111288700014, 99111288700017, 99111288700018, 99111288700019, 99111288700020, 99111627300016, 99111627300032, 99111143000017, 99111143000021, 99111143000022, 99111288700001, 99111288700002, 99129614700009, 9962076700011, 99132121000012, 99132121000015, 99132121000005, 99132121000007, 99111627300009, 99134232100017, 99131046400008, 9995356100003, 99113168300011, 99111627300025, 99132121000009, 99111627300061, 99111627300038, 9974083900001, 99111627300030, 9991451200006, 99111627300018, 9976534100006, 9945767700003, 9983349700008, 9991451200003, 9991451200009, 99111288700005, 99111627300031, 99128623400004, 99128623400005, 99128623400007, 99129614700001, 99129614700003, 99129614700004, 99129614700007, 99129614700008, 99131046400002, 99131046500001, 99131046500002, 99132121000001, 99128623400006, 99111143000016, 9969414000006, 99134232100013, 99134232100021, 99111627300036, 99132121000017, 99132121000018, 99132121000019, 99132121000020, 99132732400002, 99132732400006, 99132732400013, 99132732400014, 99134232100005, 99134232100010, 99132121000016, 99111627300013, 99111627300033, 99111627300043, 99111627300044, 99111627300051, 99111627300063, 99129614700006, 99111627300007, 99111627300024, 99111627300037, 99111627300047, 99111627300014, 99111627300052, 99113168300009, 99128623400001, 99128623400002, 99128623400009, 99128623400010, 99134232100008, 99111143000020, 99131046400010, 99132121000011, 99132732400007, 99132732400011, 99132732400012, 99132732400016, 99134232100034, 99132732400008, 99132732400015, 99132732400017, 99132732400020, 99134232100014, 9959060500014, 99111627300026, 99134232100004, ", 99134232100006", 9991451200005, 9995356200008, 99111143000010, 99111288700004, 99131046500003, 99131046500005, 9973536800007, 99131046500004, 99131046500008, 99131046500009, 99132121000004, 99132121000006, 99132121000008, 99134232100026, 99132732400001, 99132732400003, 99132732400004, 99134232100007, 99134232100019, 9940577300005, 99111143000005, 99111627300048, 99111627300062, 99132121000002, 9979841500003, 99132121000003, 99131046500010
Where was it sold?
Worldwide - US Nationwide distribution in the states of CA, DC, LA, MA, NM, PA, TX, VA, WA and the countries of Belgium, Bolivia (Plurinational State of), Brazil, Canada, Colombia, Israel, Italy, Portugal, Romania, Sweden, Taiwan (Province of China), Thailand, United Kingdom.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Belmont Instrument
All 8| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Straight Inflow/Bifurcated Outflow Line Kit recalled | Belmont Instrument | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Straight Inflow/Outflow Patient Line Kit recalled by Belmont Instrument | Belmont Instrument | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Hyperthermia Pump Procedure Kit recalled by Belmont Instrument | Belmont Instrument | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Disposable Set (4 sets/case) recalled by Belmont Instrument | Belmont Instrument | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Belmont Medical 3-Spike Disposable Set used recalled | Belmont Instrument | Lead & Heavy Metals | Nationwide |
Other Other Medical Devices recalls
All 13,778| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Citric Complete Dry Citric Acid Concentrate recalled | Nipro Renal Soultions USA | Other | Nationwide |
| Device | CLASS I | Kit: Epidural/Pain TQC 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | Halyard IV Start Kit Non-Sterile recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS I | Kit: Single Shot Epidural 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | AMS Convenience Kit: Dsaek Pack - RX recalled | Windstone Medical Packaging | Device Malfunction | Nationwide |