CLASS II ONGOING Device recall · Reported 4 Nov 2026 · FDA Enforcement Report

Alphenix Interventional Angiography System recalled

Canon Medical System USA is recalling Alphenix Interventional Angiography System, distributed nationwide in the US. The recall was initiated on 12 Oct 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesDI-04987670100208/
Quantity recalled3 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0290-2027
FDA event ID
99889
Recalling firm
Canon Medical System USA, Tustin, CA
Classification
Class II
Status
Ongoing
Recall initiated
12 Oct 2026
FDA classified
27 Oct 2026
Reported
4 Nov 2026
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Device Malfunction

Timeline

12 Oct 2026Recall initiated by company
27 Oct 2026Classified by FDA+15 days
4 Nov 2026Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Due to software issue, X-ray interventional system may experience internal error if F-Store or imagine acquisition is performed while clinical analysis function is use.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000F

Where it was distributed

US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, HI, IL, LA, MA, MI, MN, MS, NC, NV, OH, OR, TX.

Nationwide AlabamaArkansasArizonaCaliforniaFloridaGeorgiaHawaiiIllinoisLouisianaMassachusettsMichiganMinnesotaMississippiNorth CarolinaNevadaOhioOregonTexas

Questions about this recall

Why was it recalled?

Due to software issue, X-ray interventional system may experience internal error if F-Store or imagine acquisition is performed while clinical analysis function is use.

Which lots are affected?

DI-04987670100208/

Where was it sold?

US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, HI, IL, LA, MA, MI, MN, MS, NC, NV, OH, OR, TX.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 10 Nov 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

2 other products · FDA event 99889

More recalls from Canon Medical System USA

All 54

Other Other Medical Devices recalls

All 13,968