CLASS II ONGOING Device recall · Reported 30 Apr 2025 · FDA Enforcement Report

Amplatzer TorqVue LP Delivery System (TVLP) recalled by Abbott

Abbott is recalling Amplatzer TorqVue LP Delivery System (TVLP), distributed nationwide in the US. The recall was initiated on 17 Mar 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModels: 1. 9-TVLP4F90/060; UDI-DI 00811806011561 Lots 9114256 9125090 9157327 9211450 10005333 10080103 10092488 10092519 10212158 10248201 10265245 10333989 10334851 10353324 2. 9-TVLP4F90/080; UDI-DI 00811806011578 Lots 9206150 10040010 10068340 10227816 10341801
Quantity recalled4,850

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1643-2025
FDA event ID
96545
Recalling firm
Abbott, Plymouth, MN
Classification
Class II
Status
Ongoing
Recall initiated
17 Mar 2025
FDA classified
21 Apr 2025
Reported
30 Apr 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

17 Mar 2025Recall initiated by company
21 Apr 2025Classified by FDA+35 days
30 Apr 2025Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Device may have a small breach in the proximal end of the shaft under the strain relief of the delivery system. Breach is not visibly detectable and may lead to prolonged procedure, blood loss, or air ingress with potential for air embolism.

Product as listed by the FDA

Amplatzer TorqVue LP Delivery System (TVLP), Model nos. 9-TVLP4F90/060 and 9-TVLP4F90/080

Where it was distributed

Worldwide - US Nationwide distribution.

Nationwide

Questions about this recall

Why was it recalled?

Device may have a small breach in the proximal end of the shaft under the strain relief of the delivery system. Breach is not visibly detectable and may lead to prolonged procedure, blood loss, or air ingress with potential for air embolism.

Which lots are affected?

Models: 1. 9-TVLP4F90/060; UDI-DI 00811806011561 Lots 9114256 9125090 9157327 9211450 10005333 10080103 10092488 10092519 10212158 10248201 10265245 10333989 10334851 10353324 2. 9-TVLP4F90/080; UDI-DI 00811806011578 Lots 9206150 10040010 10068340 10227816 10341801

Where was it sold?

Worldwide - US Nationwide distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 96545
ReportedClassRecallCompanyReasonWhere
CLASS IIAmplatzer TorqVue LP Catheter (Tvlpc) recalled by AbbottAbbottOtherNationwide

More recalls from Abbott

All 21

Other Other Medical Devices recalls

All 13,969