CLASS II Device recall · Reported 6 Apr 2016 · FDA Enforcement Report

30" Appx., 20 Drop Admin Set w/ Integrated ChemoLock Port recalled

ICU Medical is recalling 30" Appx., 20 Drop Admin Set w/ Integrated ChemoLock Port Drip Chamber, 2 Micron Filter, distributed in 15 states. The recall was initiated on 25 Feb 2016 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codes2770706
Quantity recalled50 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1260-2016
FDA event ID
73391
Recalling firm
ICU Medical, San Clemente, CA
Classification
Class II
Status
Terminated
Recall initiated
25 Feb 2016
FDA classified
25 Mar 2016
Reported
6 Apr 2016
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Sterility

Timeline

25 Feb 2016Recall initiated by company
25 Mar 2016Classified by FDA+29 days
6 Apr 2016Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

ICU Medical has identified an issue with a connection in an isolated number of I.V. infusion administration sets that incorporate the ChemLock additive port. In these sets, the connection between the Drip Chamber or Bag Spike and the ChemoLock port has the potential for developing a leak.

A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.

Product as listed by the FDA

30" (76 cm) Appx. 6.3 ml, 20 Drop Admin Set w/ Integrated ChemoLock Port Drip Chamber, 0.2 Micron Filter, ChemoLock, Hanger, Item No. CL3528 sterile devices intended for the infusion and withdrawal of fluids.

Where it was distributed

US: NV, OH, MO, CA, FL, VA, TX, NJ, CT, MI, AZ, MA, CO Non-US: DE, AE, AT, PL, HU, SE, IL, JP, JO, SA,

ArizonaCaliforniaColoradoConnecticutDelawareFloridaIllinoisMassachusettsMichiganMissouriNew JerseyNevadaOhioTexasVirginia

Questions about this recall

Is 30" Appx., 20 Drop Admin Set w/ Integrated ChemoLock Port Drip Chamber, 2 Micron Filter recalled?

Yes. ICU Medical recalled 30" Appx., 20 Drop Admin Set w/ Integrated ChemoLock Port Drip Chamber, 2 Micron Filter on 25 Feb 2016. The recall is Class II and its status is Terminated. Recall number Z-1260-2016.

Why was it recalled?

ICU Medical has identified an issue with a connection in an isolated number of I.V. infusion administration sets that incorporate the ChemLock additive port. In these sets, the connection between the Drip Chamber or Bag Spike and the ChemoLock port has the potential for developing a leak.

Which lots are affected?

2770706

Where was it sold?

US: NV, OH, MO, CA, FL, VA, TX, NJ, CT, MI, AZ, MA, CO Non-US: DE, AE, AT, PL, HU, SE, IL, JP, JO, SA,

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.