CLASS II ONGOING Device recall · Reported 29 Jul 2026 · FDA Enforcement Report

Hotline Fluid Warming Set, single use device recalled by ICU Medical

ICU Medical is recalling Hotline Fluid Warming Set, single use device, is designed to warm blood and intravenous, distributed nationwide in the US. The recall was initiated on 24 Jun 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 30695085407007; Lot Numbers: 4293568, 4293569, 4297510, 4297509, 4326087, 4337724, 4337725, 4346349
Quantity recalled39,750 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2759-2026
FDA event ID
99289
Recalling firm
ICU Medical, Minneapolis, MN
Classification
Class II
Status
Ongoing
Recall initiated
24 Jun 2026
FDA classified
17 Jul 2026
Reported
29 Jul 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Packaging Defects

Timeline

24 Jun 2026Recall initiated by company
17 Jul 2026Classified by FDA+23 days
29 Jul 2026Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

ICU Medical became aware of an increased likelihood of leakage in specific Lots of L-70 HOTLINE Fluid Warming Sets due to potentially inadequate solvent application at the tubing to return connector bond.

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

Hotline Fluid Warming Set, single use device, is designed to warm blood and intravenous fluids and deliver them to the patient's intravenous access site at normothermic temperatures under gravity flow conditions, List Number L-70

Where it was distributed

Worldwide - US Nationwide distribution.

Nationwide

Questions about this recall

Why was it recalled?

ICU Medical became aware of an increased likelihood of leakage in specific Lots of L-70 HOTLINE Fluid Warming Sets due to potentially inadequate solvent application at the tubing to return connector bond.

Which lots are affected?

UDI-DI: 30695085407007; Lot Numbers: 4293568, 4293569, 4297510, 4297509, 4326087, 4337724, 4337725, 4346349

Where was it sold?

Worldwide - US Nationwide distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from ICU Medical

All 663

Other Other Medical Devices recalls

All 13,908