CLASS II Device recall · Reported 14 Sep 2016 · FDA Enforcement Report
ArjoHuntleigh Sara Combilizer recalled by ARJOHUNTLEIGH POLSKA Sp. z.o.o
ARJOHUNTLEIGH POLSKA Sp. z.o.o is recalling ArjoHuntleigh Sara Combilizer, distributed in 13 states. The recall was initiated on 29 Jun 2016 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2718-2016
- FDA event ID
- 74506
- Recalling firm
- ARJOHUNTLEIGH POLSKA Sp. z.o.o, KOMORNIKI
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 29 Jun 2016
- FDA classified
- 2 Sep 2016
- Reported
- 14 Sep 2016
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Other Medical Devices
Timeline
| 29 Jun 2016 | Recall initiated by company | |
| 2 Sep 2016 | Classified by FDA | +65 days |
| 14 Sep 2016 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
ArjoHuntleigh has concluded that a device tipping hazard may occur on the current Sara Combilizer lifting mechanism, providing a potential for risk to the patient or the caregiver. The tipping phenomenon has been found during internal device testing. To date, there have been no adverse events or customer complaints related to the tipping phenomenon. The tipping hazard may occur when fully loaded Sara Combilizer in "stretcher mode" is lifted to the highest position (approximately 984 mm = 38 3/4") and tilted backwards to the angle of 25 degrees¿-28 degrees¿. Although such scenario has been defined as highly unlikely to occur during use with patients, as a commitment to quality, ArjoHuntleigh is providing customers with information on action needed. The Sara Combilizer may stay in use until the upgrade will be performed, on the condition that the tilt angle in the device ''stretcher mode" will be no greater than minus 15 degrees¿ and the device is not loaded over 330 lbs (150 kg).
Product as listed by the FDA
ArjoHuntleigh Sara Combilizer; Intended to facilitate early immobilization, rehabilitation, and care of patients.
Where it was distributed
CO, AL, MA, NC, MI, NY, FL, MD, CA, VA, CT, TX, and IL; Australia, Austria, Denmark, Germany, Israel, Japan, Norway, Sweden, Switzerland, UAE, and UK.
AlabamaCaliforniaColoradoConnecticutFloridaIllinoisMassachusettsMarylandMichiganNorth CarolinaNew YorkTexasVirginia
Questions about this recall
Is ArjoHuntleigh Sara Combilizer recalled?
Yes. ARJOHUNTLEIGH POLSKA Sp. z.o.o recalled ArjoHuntleigh Sara Combilizer on 29 Jun 2016. The recall is Class II and its status is Terminated. Recall number Z-2718-2016.
Why was it recalled?
ArjoHuntleigh has concluded that a device tipping hazard may occur on the current Sara Combilizer lifting mechanism, providing a potential for risk to the patient or the caregiver. The tipping phenomenon has been found during internal device testing. To date, there have been no adverse events or customer complaints related to the tipping phenomenon. The tipping hazard may occur when fully loaded Sara Combilizer in "stretcher mode" is lifted to the highest position (approximately 984 mm = 38 3/4") and tilted backwards to the angle of 25 degrees¿-28 degrees¿. Although such scenario has been defined as highly unlikely to occur during use with patients, as a commitment to quality, ArjoHuntleigh is providing customers with information on action needed. The Sara Combilizer may stay in use until the upgrade will be performed, on the condition that the tilt angle in the device ''stretcher mode" will be no greater than minus 15 degrees¿ and the device is not loaded over 330 lbs (150 kg).
Which lots are affected?
Device Listing Number E101987, Model Number APA1000-XX; Serial numbers P0359671 , P0360103 , P0338345 , P0354548 , P0357690 , P0359672 , P0359673 , P0348548 , P0349629 , P0361336 , P0361337 , P0361875 , P0361876 , P0361877 , P0360104 , P0360105 , P0346000 , P0373592 , P0357691 , P0349634 , P0354549 , P0349627 , P0339257 , P0321458 , P0299662 , P0325930 , P0325931 , P0325933 , P0310854 , P0344175 , P0356357 , P0356362 , P0356363 , P0360990 , P0375493 , P0298175 , P0309635 , P0310848 , P0323622 , P0323624 , P0325167 , 1404000118 , P0296131 , P0296315 , P0298465 , P0319583 , P0333285 , P0333288 , P0339264 , P0343583 , P0343584 , P0343585 , P0343684 , P0344174 , P0353246 , P0364628 , 1404000112 , P0333284 , P0333290 , P0294982 , P0294983 , P0295762 , P0298158 , P0309472 , P0317162 , P0325970 , P0331310 , 1404000113 , 1404000119 , 1404000120 , P0325967 , P0332385 , P0333283 , P0339574 , P0296130 , P0296318 , P0321460 , P0332383 , 1404000115 , P0331312 , P0331311 , P0331318 , P0331319 , P0325968 , P0295846 , P0294981 , P0294985 , P0295847 , P0332386 , P0294984 , P0332384 , 1404000116 , 1404000117 , P0321461 , 1404000114 , P0294986 , P0298176 , P0323011 , P0361422 , P0361423 , P0363456 , P0357692 , P0360102 , P0328718 , P0328720 , P0328724 .
Where was it sold?
CO, AL, MA, NC, MI, NY, FL, MD, CA, VA, CT, TX, and IL; Australia, Austria, Denmark, Germany, Israel, Japan, Norway, Sweden, Switzerland, UAE, and UK.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from ARJOHUNTLEIGH POLSKA Sp. z.o.o
All 10| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Arjo Tenor mobile passive patient lift recalled | ARJOHUNTLEIGH POLSKA Sp. z.o.o | Device Malfunction | Nationwide |
| Device | CLASS II | Arjo medical beds recalled by ARJOHUNTLEIGH POLSKA Sp. z.o.o | ARJOHUNTLEIGH POLSKA Sp. z.o.o | Other | 7 states |
| Device | CLASS II | Arjo medical beds recalled by ARJOHUNTLEIGH POLSKA Sp. z.o.o | ARJOHUNTLEIGH POLSKA Sp. z.o.o | Other | 7 states |
| Device | CLASS I | Sara Plus Active Floor Lift recalled by ARJOHUNTLEIGH POLSKA Sp. z.o.o | ARJOHUNTLEIGH POLSKA Sp. z.o.o | Other | Nationwide |
| Device | CLASS II | Arjo Citadel beds originally assembled recalled | ARJOHUNTLEIGH POLSKA Sp. z.o.o | Lead & Heavy Metals | Nationwide |
Other Other Medical Devices recalls
All 13,771| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Convenience Kits: Blood Draw Kit recalled by Medline Industries | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep Clear recalled | Medline Industries | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Ortho Aspiration Kit recalled | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep DYKA1054F recalled | Medline Industries | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Arterial Line Insertion Kit recalled | Medline Industries | Sterility | Nationwide |