Atellica CH Iron_2 (Iron_2) recalled by Siemens Healthcare Diagnostics
Siemens Healthcare Diagnostics is recalling Atellica CH Iron_2 (Iron_2) is for in vitro diagnostic use in the quantitative, distributed nationwide in the US. The recall was initiated on 28 Jun 2022 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1498-2022
- FDA event ID
- 90559
- Recalling firm
- Siemens Healthcare Diagnostics, Tarrytown, NY
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 28 Jun 2022
- FDA classified
- 4 Aug 2022
- Reported
- 10 Aug 2022
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Potency & Assay
- Device type
- Test Kits & Diagnostics
Timeline
| 28 Jun 2022 | Recall initiated by company | |
| 4 Aug 2022 | Classified by FDA | +37 days |
| 10 Aug 2022 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Falsely Elevated Atellica CH Microalbumin_2 (¿ALB_2) Results due to Reagent Carryover from the Iron_2 Assay. This defect occurs only when a microalbumin test is run immediately after an iron test.
Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.
Product as listed by the FDA
Atellica CH Iron_2 (Iron_2) is for in vitro diagnostic use in the quantitative determination of iron in human serum and (lithium heparin) plasma samples.
Where it was distributed
Worldwide distribution - US Nationwide and the countries of AR, AU, AT, BS, BH, BD, BE, BR, BG, BF, CA, CL, CO, HR, CW, CZ, DK, EC, EG, EE, FI, FR, DE, GR, GP, HK, HU, IN, IR, IQ, IE, IL, IT, JP, JO, KE, KW, LV, LY, LT, MY, MX, MA, NL, NZ, NO, OM, CN, PK, PY, PE, PH, PL, PT, QA, KR, RO, RU, SA, RS, SG, SK, ZA, ES, SE, CH, TW, TH, TR, AE, UG, GB, UY, VA, VN.
Nationwide ArkansasCaliforniaColoradoDelawareIllinoisIndianaMassachusettsVirginia
Questions about this recall
Why was it recalled?
Falsely Elevated Atellica CH Microalbumin_2 (¿ALB_2) Results due to Reagent Carryover from the Iron_2 Assay. This defect occurs only when a microalbumin test is run immediately after an iron test.
Which lots are affected?
Unique Device Identification (UDI): 00630414596402 All lots
Where was it sold?
Worldwide distribution - US Nationwide and the countries of AR, AU, AT, BS, BH, BD, BE, BR, BG, BF, CA, CL, CO, HR, CW, CZ, DK, EC, EG, EE, FI, FR, DE, GR, GP, HK, HU, IN, IR, IQ, IE, IL, IT, JP, JO, KE, KW, LV, LY, LT, MY, MX, MA, NL, NZ, NO, OM, CN, PK, PY, PE, PH, PL, PT, QA, KR, RO, RU, SA, RS, SG, SK, ZA, ES, SE, CH, TW, TH, TR, AE, UG, GB, UY, VA, VN.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Siemens Healthcare Diagnostics
All 694| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Syva EMIT 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide |
| Device | CLASS II | Beckman Coulter, Emit 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide |
| Device | CLASS III | Syva Emit Disopyramide Assay Enzyme Immunoassay recalled | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide |
| Device | CLASS II | Atellica CH Urine Albumin recalled by Siemens Healthcare Diagnostics | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide |
| Device | CLASS III | Syva Emit tox Serum Tricyclic Antidepressant recalled | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide |
Other Test Kits & Diagnostics recalls
All 2,656| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | STA Staclot Protein recalled by Diagnostica Stago | Diagnostica Stago | Potency & Assay | Nationwide |
| Device | CLASS II | DxC 500i Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | DxC 500 AU Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | Syva EMIT 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide |
| Device | CLASS II | Beckman Coulter, Emit 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide |