CLASS II Device recall · Reported 5 Jan 2022 · FDA Enforcement Report

Atellica CH ¿2-Microglobulin (B2M) -M) Assay recalled

Siemens Healthcare Diagnostics is recalling Atellica CH ¿2-Microglobulin (B2M) -M) Assay for in vitro diagnostic use in the, distributed nationwide in the US. The recall was initiated on 1 Dec 2021 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLot Number: 211446 UDI: 00630414595566
Quantity recalledUS: 46 units OUS: 794 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0440-2022
FDA event ID
89211
Recalling firm
Siemens Healthcare Diagnostics, Tarrytown, NY
Classification
Class II
Status
Terminated
Recall initiated
1 Dec 2021
FDA classified
29 Dec 2021
Reported
5 Jan 2022
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Potency & Assay

Timeline

1 Dec 2021Recall initiated by company
29 Dec 2021Classified by FDA+28 days
5 Jan 2022Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

¿2 Microglobulin (B2M) Reagent Lot 211446 Bias with Quality Control and Patient Sample Results. Positive bias on patient samples of up to +20.9% at a B2M concentration of approximately 1.5 mg/L and a negative bias on patient samples of up to -10.5% at a B2M concentration of approximately 15 mg/L

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

Atellica CH ¿2-Microglobulin (B2M) -M) Assay for in vitro diagnostic use in the quantitative determination of ¿2-microglobulin in human serum and plasma Siemens Material Number (SMN): 11097635

Where it was distributed

Worldwide distribution - US Nationwide distribution in the states of CA, CO, MA, MD, MS, NC, NE, NH, OH, VA, WI and the countries of Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Czech Republic, Denmark, Ecuador, France, Germany, Greece, India, Iran, Israel, Italy, Japan, Mexico, Morocco, Oman, P.R. China, Pakistan, Philippines, Portugal, Slovakia, South Africa, Spain, Switzerland, Turkey, United Kingdom, Uruguay.

Nationwide CaliforniaColoradoMassachusettsMarylandMississippiNorth CarolinaNebraskaNew HampshireOhioVirginiaWisconsin

Questions about this recall

Why was it recalled?

¿2 Microglobulin (B2M) Reagent Lot 211446 Bias with Quality Control and Patient Sample Results. Positive bias on patient samples of up to +20.9% at a B2M concentration of approximately 1.5 mg/L and a negative bias on patient samples of up to -10.5% at a B2M concentration of approximately 15 mg/L

Which lots are affected?

Lot Number: 211446 UDI: 00630414595566

Where was it sold?

Worldwide distribution - US Nationwide distribution in the states of CA, CO, MA, MD, MS, NC, NE, NH, OH, VA, WI and the countries of Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Czech Republic, Denmark, Ecuador, France, Germany, Greece, India, Iran, Israel, Italy, Japan, Mexico, Morocco, Oman, P.R. China, Pakistan, Philippines, Portugal, Slovakia, South Africa, Spain, Switzerland, Turkey, United Kingdom, Uruguay.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.