CLASS II ONGOING Device recall · Reported 15 Sep 2021 · FDA Enforcement Report

Atellica IM BR 29 (BR) Assay 50 Test Kit recalled

Siemens Healthcare Diagnostics is recalling Atellica IM BR 29 (BR) Assay 50 Test Kit- in vitro diagnostic use in the quantitative, distributed nationwide in the US. The recall was initiated on 6 Jul 2021 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesAll lots impacted. Kit lot numbers: 40476255 & 64505255 expiration date 2021-08-03, Lot numbers: 89332257 & 05694257 expiration date 2021-11-10.
Quantity recalled580 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2424-2021
FDA event ID
88431
Recalling firm
Siemens Healthcare Diagnostics, East Walpole, MA
Classification
Class II
Status
Ongoing
Recall initiated
6 Jul 2021
FDA classified
3 Sep 2021
Reported
15 Sep 2021
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Potency & Assay

Timeline

6 Jul 2021Recall initiated by company
3 Sep 2021Classified by FDA+59 days
15 Sep 2021Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Product does not meet the Instructions for Use (IFU) claimed Pack Calibration Interval of 10 days. IFU calibration updated interval from 10 days to 3 days for all current and future kit lot numbers

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

Atellica IM BR 27.29 (BR) Assay 50 Test Kit- in vitro diagnostic use in the quantitative serial determination of cancer antigen CA 27.29 in human serum and plasma (EDTA) using the Atellica IM Analyzer. SMN: 10995478

Where it was distributed

Worldwide distribution. US Nationwide, Australia, Bahamas, India, Mexico, Saudi Arabia, and Viet Nam

Nationwide

Questions about this recall

Is Atellica IM BR 29 (BR) Assay 50 Test Kit- in vitro diagnostic use in the quantitative recalled?

Yes. Siemens Healthcare Diagnostics recalled Atellica IM BR 29 (BR) Assay 50 Test Kit- in vitro diagnostic use in the quantitative on 6 Jul 2021. The recall is Class II and its status is Ongoing. Recall number Z-2424-2021.

Why was it recalled?

Product does not meet the Instructions for Use (IFU) claimed Pack Calibration Interval of 10 days. IFU calibration updated interval from 10 days to 3 days for all current and future kit lot numbers

Which lots are affected?

All lots impacted. Kit lot numbers: 40476255 & 64505255 expiration date 2021-08-03, Lot numbers: 89332257 & 05694257 expiration date 2021-11-10.

Where was it sold?

Worldwide distribution. US Nationwide, Australia, Bahamas, India, Mexico, Saudi Arabia, and Viet Nam

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Siemens Healthcare Diagnostics

All 694

Other Test Kits & Diagnostics recalls

All 2,656