CLASS I ONGOING Device recall · Reported 8 Oct 2025 · FDA Enforcement Report

Atlan A350 recalled by Draeger

Draeger is recalling Atlan A350, distributed in 25 states. The recall was initiated on 9 Sep 2025 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesLot Code: UDI-DI Number: 04048675556176 Serial Numbers Distributed in the US: ASTB-0181 ASTC-0005,ASTC-0006,ASTC-0007 ASTD-0244,ASTD-0259 ASTE-0180,ASTE-0254,ASTE-0255 ASTH-0083,ASTH-0464 ASUA-0479,ASUA-0482 ASUB-0012,ASUB-0013,ASUB-0015,ASUB-0017,ASUB-0018,ASUB-0292 ASUC-0065,ASUC-0084,ASUC-0085,ASUC-0086,ASUC-0087,ASUC-0088,ASUC-0090,ASUC-0091, ASUC-0092,ASUC-0096,ASUC-0097,ASUC-0099,ASUC-0100,ASUC-0101,ASUC-0102,ASUC-0103, ASUC-0104,ASUC-0105,ASUC-0107,ASUC-0108,ASUC-0109,ASUC-0110,ASUC-0111,ASUC-0112, ASUC-0113,ASUC-0116,ASUC-0117,ASUC-0118,ASUC-0119,ASUC-0120,ASUC-0121,ASUC-0122, ASUC-0123,ASUC-0125,ASUC-0126,ASUC-0127,ASUC-0128,ASUC-0129,ASUC-0130,ASUC-0131, ASUC-0132,ASUC-0133,ASUC-0134,ASUC-0135,ASUC-0136,ASUC-0137,ASUC-0138,ASUC-0139, ASUC-0140,ASUC-0141,ASUC-0142,ASUC-0143,ASUC-0143,ASUC-0144,ASUC-0145,ASUC-0162, ASUC-0163,ASUC-0164,ASUC-0165,ASUC-0166,ASUC-0167,ASUC-0168,ASUC-0169,ASUC-0170, ASUC-0171,ASUC-0172,ASUC-0173,ASUC-0203,ASUC-0204,ASUC-0205,ASUC-0206,ASUC-0207, ASUC-0208,ASUC-0209,ASUC-0210,ASUC-0218,ASUC-0219,ASUC-0220,ASUC-0221,ASUC-0222, ASUC-0223,ASUC-0224,ASUC-0225,ASUC-0235,ASUC-0236,ASUC-0237,ASUC-0340,ASUC-0341 ASUD-0244,ASUD-0261,ASUD-0262,ASUD-0263,ASUD-0267,ASUD-0281,ASUD-0282,ASUD-0283 Serial Numbers Distributed OUS ASUA-0353,ASUA-0441,ASUA-0447,ASUA-0472,ASUA-0478,ASUA-0496,,SUA-0497,ASUA-0500, ASUB-0004,ASUB-0024,ASUB-0025,ASUB-0030,ASUB-0033,ASUB-0034,ASUB-0035,ASUB-0038, ASUB-0039,ASUB-0040,ASUB-0041,ASUB-0043,ASUB-0044,ASUB-0046,ASUB-0048,ASUB-0063, ASUB-0064,ASUB-0066,ASUB-0067,ASUB-0069,ASUB-0073,ASUB-0074,ASUB-0075,ASUB-0076, ASUB-0077,ASUB-0095,ASUB-0096,ASUB-0097,ASUB-0098,ASUB-0099,ASUB-0100,ASUB-0101, ASUB-0102,ASUB-0104,ASUB-0105,ASUB-0107,ASUB-0108,ASUB-0109,ASUB-0112,ASUB-0113, ASUB-0114,ASUB-0115,ASUB-0116,ASUB-0118,ASUB-0120,ASUB-0121,ASUB-0122,ASUB-0123, ASUB-0124,ASUB-0128,ASUB-0129,ASUB-0130,ASUB-0136,ASUB-0137,ASUB-0139,ASUB-0140, ASUB-0141,ASUB-0143,ASUB-0144,ASUB-0145,ASUB-0146,ASUB-0147,ASUB-0148,ASUB-0149, ASUB-0150,ASUB-0151,ASUB-0152,ASUB-0153,ASUB-0154,ASUB-0155,ASUB-0156,ASUB-0157, ASUB-0158,ASUB-0159,ASUB-0160,ASUB-0161,ASUB-0162,ASUB-0163,ASUB-0177,ASUB-0178, ASUB-0188,ASUB-0189,ASUB-0190,ASUB-0191,ASUB-0192,ASUB-0193,ASUB-0194,ASUB-0195, ASUB-0196,ASUB-0197,ASUB-0198,ASUB-0199,ASUB-0201,ASUB-0202,ASUB-0203,ASUB-0204, ASUB-0205,ASUB-0206,ASUB-0207,ASUB-0208,ASUB-0209,ASUB-0210,ASUB-0211,ASUB-0223, ASUB-0224,ASUB-0225,ASUB-0226,ASUB-0227,ASUB-0229,ASUB-0230,ASUB-0231,ASUB-0232, ASUB-0233,ASUB-0234,ASUB-0235,ASUB-0236,ASUB-0237,ASUB-0238,ASUB-0240,ASUB-0243, ASUB-0245,ASUB-0252,ASUB-0253,ASUB-0254,ASUB-0255,ASUB-0256,ASUB-0257,ASUB-0258, ASUB-0260,ASUB-0261,ASUB-0262,ASUB-0263,ASUB-0264,ASUB-0265,ASUB-0271,ASUB-0272, ASUB-0273,ASUB-0276,ASUB-0278,ASUC-0001ASUC-0004,ASUC-0016,ASUC-0017,ASUC-0018, ASUC-0019,ASUC-0020,ASUC-0022,ASUC-0024,ASUC-0026,,ASUC-0028,,ASUC-0030,ASUC-0031, ASUC-0032,ASUC-0033,ASUC-0034,ASUC-0035,ASUC-0036,ASUC-0039,ASUC-0061,ASUC-0062, ASUC-0063,ASUC-0226,ASUC-0227,ASUC-0228,ASUC-0229,ASUC-0230,ASUC-0232,ASUC-0233, ASUC-0234,ASUC-0247,ASUC-0248,ASUC-0249,ASUC-0250,ASUC-0251,ASUC-0252,ASUC-0253, ASUC-0254,ASUC-0255,ASUC-0256,ASUC-0257,ASUC-0262,ASUC-0263,ASUC-0264,ASUC-0265, ASUC-0266,ASUC-0267,ASUC-0269,ASUC-0272,ASUC-0273,ASUC-0274,ASUC-0275,ASUC-0276, ASUC-0277,ASUC-0278,ASUC-0279,ASUC-0284,ASUC-0286,ASUC-0287,ASUC-0288,ASUC-0289, ASUC-0290,ASUC-0291,ASUC-0292,ASUC-0293,ASUC-0294,ASUC-0295,ASUC-0296,ASUC-0300, ASUC-0301,ASUC-0302,ASUC-0303,ASUC-0304,ASUC-0305,ASUC-0306ASUC-0307,ASUC-0308, ASUC-0309,ASUC-0310,ASUC-0311,ASUC-0314,ASUC-0315,ASUC-0316,ASUC-0319,ASUC-0320, ASUC-0321,ASUC-0322,ASUC-0323,ASUC-0339,ASUC-0342,ASUC-0343,ASUC-0344,ASUC-0345, ASUC-0346,ASUC-0348,ASUC-0349,ASUC-0350,ASUC-0351,ASUC-0352,ASUC-0353,ASUC-0354, ASUC-0355,ASUC-0356,ASUC-0357,ASUC-0358,ASUC-0359,ASUC-0360,ASUC-0361,ASUC-0362, ASUC-0363,ASUC-0364,ASUC-0365,ASUC-0366,ASUC-0367,ASUC-0380,ASUC-0381,ASUC-0382, ASUC-0383,ASUC-0384,ASUC-0385,ASUC-0386,ASUC-0387,ASUC-0391,ASUC-0392,ASUC-0393, ASUC-0394,ASUC-0395,ASUC-0396,ASUC-0397,ASUC-0398,ASUC-0399,ASUC-0400,ASUC-0401, ASUC-0402,ASUC-0403,ASUC-0404,ASUC-0405,ASUC-0406,ASUC-0407,ASUC-0408,ASUC-0409, ASUC-0410,ASUC-0416,ASUD-0001,ASUD-0002,ASUD-0004,ASUD-0005,ASUD-0006,ASUD-0007, ASUD-0008,ASUD-0009,ASUD-0010,ASUD-0011,ASUD-0012,ASUD-0013,ASUD-0014,ASUD-0015, ASUD-0016,ASUD-0018,ASUD-0019,ASUD-0020,ASUD-0022,ASUD-0023,ASUD-0041,ASUD-0042, ASUD-0043,ASUD-0044,ASUD-0046,ASUD-0047,ASUD-0048,ASUD-0049,ASUD-0050,ASUD-0051, ASUD-0057,ASUD-0059,ASUD-0060,ASUD-0061,ASUD-0062,ASUD-0063,ASUD-0064,ASUD-0065, ASUD-0066,ASUD-0074,ASUD-0075,ASUD-0077,ASUD-0078,ASUD-0079,ASUD-0080,ASUD-0081, ASUD-0098,ASUD-0099,ASUD-0100,ASUD-0101,ASUD-0102,ASUD-0103,ASUD-0104,ASUD-0105, ASUD-0106,ASUD-0107,ASUD-0108,ASUD-0109,ASUD-0126,ASUD-0127,ASUD-0128,ASUD-0129, ASUD-0130,ASUD-0131,ASUD-0133,ASUD-0135,ASUD-0136,ASUD-0140,ASUD-0141,ASUD-0150, ASUD-0168,ASUD-0173,ASUD-0174,ASUD-0175,ASUD-0181,ASUD-0182,ASUD-0183,ASUD-0184, ASUD-0185,ASUD-0186,ASUD-0196,ASUD-0198,ASUD-0199,ASUD-0200,ASUD-0201,ASUD-0202, ASUD-0203,ASUD-0204,ASUD-0205,ASUD-0206,ASUD-0219,ASUD-0220,ASUD-0221,ASUD-0222, ASUD-0223,ASUD-0224,ASUD-0225,ASUD-0226,ASUD-0227,ASUD-0228,ASUD-0238,ASUD-0239, ASUD-0245,ASUD-0246,ASUD-0247,ASUD-0248,ASUD-0249,ASUD-0250,ASUD-0251,ASUD-0252, ASUD-0253,ASUD-0258,ASUD-0259,ASUD-0260,ASUD-0273,ASUD-0274
Quantity recalled530

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2616-2025
FDA event ID
97525
Recalling firm
Draeger, Telford, PA
Classification
Class I
Status
Ongoing
Recall initiated
9 Sep 2025
FDA classified
1 Oct 2025
Reported
8 Oct 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

9 Sep 2025Recall initiated by company
1 Oct 2025Classified by FDA+22 days
8 Oct 2025Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The device either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Brand Name: Atlan Product Name: Atlan A350 Model/Catalog Number: 8211500 Software Version: Not applicable Product Description: The Atlan is an anesthesia workstation intended for use in anesthetizing adults, pediatrics, and neonates. The device can be used for mechanical ventilation, manual ventilation, pressure-supported spontaneous breathing, and spontaneous breathing. The device is specified for inhalational anesthesia and/or patient ventilation in accordance with the intended use during surgical or diagnostic interventions Component: Not applicable

Where it was distributed

Devices were imported from Draegerwerk Ag and Co. KGaA and distributed in the following states: AL, AZ, CA, CT, FL, GA, ID, IL, IN, KS, KY, MA, MN, MO, NC, NE, NM, NV, NY, OH, PA, TX, UT, VA, WI. Devices were direct shipped from Draegerwerk Ag and Co. KGaA to the following countries: Argentina, Australia, Belgium, Brazil, Chili, China, Croatia, Denmark, Ecuador, France, Georgia, Germany, Greece, Hungary, India, Indonesia, Italy, Japan, Jordan, Kuwait, Lithuania, Mexico, Netherlands, Peru, Poland, Romania, Russian Federation, Saudi Arabia, Serbia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, Vietnam.

AlabamaArizonaCaliforniaConnecticutFloridaGeorgiaIdahoIllinoisIndianaKansasKentuckyMassachusettsMinnesotaMissouriNorth CarolinaNebraskaNew MexicoNevadaNew YorkOhio

Questions about this recall

Is Atlan A350 recalled?

Yes. Draeger recalled Atlan A350 on 9 Sep 2025. The recall is Class I and its status is Ongoing. Recall number Z-2616-2025.

Why was it recalled?

The device either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.

Which lots are affected?

Lot Code: UDI-DI Number: 04048675556176 Serial Numbers Distributed in the US: ASTB-0181 ASTC-0005,ASTC-0006,ASTC-0007 ASTD-0244,ASTD-0259 ASTE-0180,ASTE-0254,ASTE-0255 ASTH-0083,ASTH-0464 ASUA-0479,ASUA-0482 ASUB-0012,ASUB-0013,ASUB-0015,ASUB-0017,ASUB-0018,ASUB-0292 ASUC-0065,ASUC-0084,ASUC-0085,ASUC-0086,ASUC-0087,ASUC-0088,ASUC-0090,ASUC-0091, ASUC-0092,ASUC-0096,ASUC-0097,ASUC-0099,ASUC-0100,ASUC-0101,ASUC-0102,ASUC-0103, ASUC-0104,ASUC-0105,ASUC-0107,ASUC-0108,ASUC-0109,ASUC-0110,ASUC-0111,ASUC-0112, ASUC-0113,ASUC-0116,ASUC-0117,ASUC-0118,ASUC-0119,ASUC-0120,ASUC-0121,ASUC-0122, ASUC-0123,ASUC-0125,ASUC-0126,ASUC-0127,ASUC-0128,ASUC-0129,ASUC-0130,ASUC-0131, ASUC-0132,ASUC-0133,ASUC-0134,ASUC-0135,ASUC-0136,ASUC-0137,ASUC-0138,ASUC-0139, ASUC-0140,ASUC-0141,ASUC-0142,ASUC-0143,ASUC-0143,ASUC-0144,ASUC-0145,ASUC-0162, ASUC-0163,ASUC-0164,ASUC-0165,ASUC-0166,ASUC-0167,ASUC-0168,ASUC-0169,ASUC-0170, ASUC-0171,ASUC-0172,ASUC-0173,ASUC-0203,ASUC-0204,ASUC-0205,ASUC-0206,ASUC-0207, ASUC-0208,ASUC-0209,ASUC-0210,ASUC-0218,ASUC-0219,ASUC-0220,ASUC-0221,ASUC-0222, ASUC-0223,ASUC-0224,ASUC-0225,ASUC-0235,ASUC-0236,ASUC-0237,ASUC-0340,ASUC-0341 ASUD-0244,ASUD-0261,ASUD-0262,ASUD-0263,ASUD-0267,ASUD-0281,ASUD-0282,ASUD-0283 Serial Numbers Distributed OUS ASUA-0353,ASUA-0441,ASUA-0447,ASUA-0472,ASUA-0478,ASUA-0496,,SUA-0497,ASUA-0500, ASUB-0004,ASUB-0024,ASUB-0025,ASUB-0030,ASUB-0033,ASUB-0034,ASUB-0035,ASUB-0038, ASUB-0039,ASUB-0040,ASUB-0041,ASUB-0043,ASUB-0044,ASUB-0046,ASUB-0048,ASUB-0063, ASUB-0064,ASUB-0066,ASUB-0067,ASUB-0069,ASUB-0073,ASUB-0074,ASUB-0075,ASUB-0076, ASUB-0077,ASUB-0095,ASUB-0096,ASUB-0097,ASUB-0098,ASUB-0099,ASUB-0100,ASUB-0101, ASUB-0102,ASUB-0104,ASUB-0105,ASUB-0107,ASUB-0108,ASUB-0109,ASUB-0112,ASUB-0113, ASUB-0114,ASUB-0115,ASUB-0116,ASUB-0118,ASUB-0120,ASUB-0121,ASUB-0122,ASUB-0123, ASUB-0124,ASUB-0128,ASUB-0129,ASUB-0130,ASUB-0136,ASUB-0137,ASUB-0139,ASUB-0140, ASUB-0141,ASUB-0143,ASUB-0144,ASUB-0145,ASUB-0146,ASUB-0147,ASUB-0148,ASUB-0149, ASUB-0150,ASUB-0151,ASUB-0152,ASUB-0153,ASUB-0154,ASUB-0155,ASUB-0156,ASUB-0157, ASUB-0158,ASUB-0159,ASUB-0160,ASUB-0161,ASUB-0162,ASUB-0163,ASUB-0177,ASUB-0178, ASUB-0188,ASUB-0189,ASUB-0190,ASUB-0191,ASUB-0192,ASUB-0193,ASUB-0194,ASUB-0195, ASUB-0196,ASUB-0197,ASUB-0198,ASUB-0199,ASUB-0201,ASUB-0202,ASUB-0203,ASUB-0204, ASUB-0205,ASUB-0206,ASUB-0207,ASUB-0208,ASUB-0209,ASUB-0210,ASUB-0211,ASUB-0223, ASUB-0224,ASUB-0225,ASUB-0226,ASUB-0227,ASUB-0229,ASUB-0230,ASUB-0231,ASUB-0232, ASUB-0233,ASUB-0234,ASUB-0235,ASUB-0236,ASUB-0237,ASUB-0238,ASUB-0240,ASUB-0243, ASUB-0245,ASUB-0252,ASUB-0253,ASUB-0254,ASUB-0255,ASUB-0256,ASUB-0257,ASUB-0258, ASUB-0260,ASUB-0261,ASUB-0262,ASUB-0263,ASUB-0264,ASUB-0265,ASUB-0271,ASUB-0272, ASUB-0273,ASUB-0276,ASUB-0278,ASUC-0001ASUC-0004,ASUC-0016,ASUC-0017,ASUC-0018, ASUC-0019,ASUC-0020,ASUC-0022,ASUC-0024,ASUC-0026,,ASUC-0028,,ASUC-0030,ASUC-0031, ASUC-0032,ASUC-0033,ASUC-0034,ASUC-0035,ASUC-0036,ASUC-0039,ASUC-0061,ASUC-0062, ASUC-0063,ASUC-0226,ASUC-0227,ASUC-0228,ASUC-0229,ASUC-0230,ASUC-0232,ASUC-0233, ASUC-0234,ASUC-0247,ASUC-0248,ASUC-0249,ASUC-0250,ASUC-0251,ASUC-0252,ASUC-0253, ASUC-0254,ASUC-0255,ASUC-0256,ASUC-0257,ASUC-0262,ASUC-0263,ASUC-0264,ASUC-0265, ASUC-0266,ASUC-0267,ASUC-0269,ASUC-0272,ASUC-0273,ASUC-0274,ASUC-0275,ASUC-0276, ASUC-0277,ASUC-0278,ASUC-0279,ASUC-0284,ASUC-0286,ASUC-0287,ASUC-0288,ASUC-0289, ASUC-0290,ASUC-0291,ASUC-0292,ASUC-0293,ASUC-0294,ASUC-0295,ASUC-0296,ASUC-0300, ASUC-0301,ASUC-0302,ASUC-0303,ASUC-0304,ASUC-0305,ASUC-0306ASUC-0307,ASUC-0308, ASUC-0309,ASUC-0310,ASUC-0311,ASUC-0314,ASUC-0315,ASUC-0316,ASUC-0319,ASUC-0320, ASUC-0321,ASUC-0322,ASUC-0323,ASUC-0339,ASUC-0342,ASUC-0343,ASUC-0344,ASUC-0345, ASUC-0346,ASUC-0348,ASUC-0349,ASUC-0350,ASUC-0351,ASUC-0352,ASUC-0353,ASUC-0354, ASUC-0355,ASUC-0356,ASUC-0357,ASUC-0358,ASUC-0359,ASUC-0360,ASUC-0361,ASUC-0362, ASUC-0363,ASUC-0364,ASUC-0365,ASUC-0366,ASUC-0367,ASUC-0380,ASUC-0381,ASUC-0382, ASUC-0383,ASUC-0384,ASUC-0385,ASUC-0386,ASUC-0387,ASUC-0391,ASUC-0392,ASUC-0393, ASUC-0394,ASUC-0395,ASUC-0396,ASUC-0397,ASUC-0398,ASUC-0399,ASUC-0400,ASUC-0401, ASUC-0402,ASUC-0403,ASUC-0404,ASUC-0405,ASUC-0406,ASUC-0407,ASUC-0408,ASUC-0409, ASUC-0410,ASUC-0416,ASUD-0001,ASUD-0002,ASUD-0004,ASUD-0005,ASUD-0006,ASUD-0007, ASUD-0008,ASUD-0009,ASUD-0010,ASUD-0011,ASUD-0012,ASUD-0013,ASUD-0014,ASUD-0015, ASUD-0016,ASUD-0018,ASUD-0019,ASUD-0020,ASUD-0022,ASUD-0023,ASUD-0041,ASUD-0042, ASUD-0043,ASUD-0044,ASUD-0046,ASUD-0047,ASUD-0048,ASUD-0049,ASUD-0050,ASUD-0051, ASUD-0057,ASUD-0059,ASUD-0060,ASUD-0061,ASUD-0062,ASUD-0063,ASUD-0064,ASUD-0065, ASUD-0066,ASUD-0074,ASUD-0075,ASUD-0077,ASUD-0078,ASUD-0079,ASUD-0080,ASUD-0081, ASUD-0098,ASUD-0099,ASUD-0100,ASUD-0101,ASUD-0102,ASUD-0103,ASUD-0104,ASUD-0105, ASUD-0106,ASUD-0107,ASUD-0108,ASUD-0109,ASUD-0126,ASUD-0127,ASUD-0128,ASUD-0129, ASUD-0130,ASUD-0131,ASUD-0133,ASUD-0135,ASUD-0136,ASUD-0140,ASUD-0141,ASUD-0150, ASUD-0168,ASUD-0173,ASUD-0174,ASUD-0175,ASUD-0181,ASUD-0182,ASUD-0183,ASUD-0184, ASUD-0185,ASUD-0186,ASUD-0196,ASUD-0198,ASUD-0199,ASUD-0200,ASUD-0201,ASUD-0202, ASUD-0203,ASUD-0204,ASUD-0205,ASUD-0206,ASUD-0219,ASUD-0220,ASUD-0221,ASUD-0222, ASUD-0223,ASUD-0224,ASUD-0225,ASUD-0226,ASUD-0227,ASUD-0228,ASUD-0238,ASUD-0239, ASUD-0245,ASUD-0246,ASUD-0247,ASUD-0248,ASUD-0249,ASUD-0250,ASUD-0251,ASUD-0252, ASUD-0253,ASUD-0258,ASUD-0259,ASUD-0260,ASUD-0273,ASUD-0274

Where was it sold?

Devices were imported from Draegerwerk Ag and Co. KGaA and distributed in the following states: AL, AZ, CA, CT, FL, GA, ID, IL, IN, KS, KY, MA, MN, MO, NC, NE, NM, NV, NY, OH, PA, TX, UT, VA, WI. Devices were direct shipped from Draegerwerk Ag and Co. KGaA to the following countries: Argentina, Australia, Belgium, Brazil, Chili, China, Croatia, Denmark, Ecuador, France, Georgia, Germany, Greece, Hungary, India, Indonesia, Italy, Japan, Jordan, Kuwait, Lithuania, Mexico, Netherlands, Peru, Poland, Romania, Russian Federation, Saudi Arabia, Serbia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, Vietnam.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.