CLASS II ONGOING Device recall · Reported 22 Nov 2023 · FDA Enforcement Report

AU/DxC AU Chemistry Transferrin recalled by Beckman Coulter Biomedical

Beckman Coulter Biomedical Division is recalling AU/DxC AU Chemistry Transferrin, distributed nationwide in the US. The recall was initiated on 6 Sep 2023 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 15099590010492, LOT: 2573, Expiration: 01 August 2024
Quantity recalled2,754

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0309-2024
FDA event ID
93206
Recalling firm
Beckman Coulter Biomedical Division, County Clare, N/A
Classification
Class II
Status
Ongoing
Recall initiated
6 Sep 2023
FDA classified
16 Nov 2023
Reported
22 Nov 2023
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Potency & Assay

Timeline

6 Sep 2023Recall initiated by company
16 Nov 2023Classified by FDA+71 days
22 Nov 2023Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The firm has identified that the Transferrin reagent does not meet the Instructions for Use (IFU) reagent on-board stability claims, which may cause control failures and/or erroneous patient samples results. The Complaint handling unit has received a total of 2 safety cases (CASE-2023-02027212 and CASE-2023-02045278) and 99 quality cases with the failure mode attributed to Transferrin reagent (OSR6152 / lot 2573) associated with reagent blank shifts for the Reagent 2 (R2), imprecision, calibration failures, erratic Quality Control (QC) recoveries, and/or erroneous high/low (serum, plasma, urine) patient results. The issue was confirmed through internal testing by Beckman Coulter on June 30, 2023. Reagent Blank shifts for R2, imprecision, calibration failures, erratic Quality Control (QC) recoveries, and/or erroneous high/low (serum, plasma, urine) patient results. The issue was confirmed through internal testing by Beckman Coulter on June 30, 2023.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

AU/DxC AU Chemistry Transferrin REF: OSR6152 Immune complexes formed in solution scatter light in proportion to their size, shape and concentration. Turbidimeters measure the reduction of incident light due to reflection, absorption, or scatter. In the procedure, the measurement of the decrease in light transmitted (increase in absorbance) through particles suspended in solution as a result of complexes formed during the antigen-antibody reaction, is the basis of this assay.

Where it was distributed

Worldwide - US Nationwide distribution including in the states of NY, CA, MS, GA, PA, SC, WA, TN, AL, AK, OR, MO, KS, FL, LA, OH, TX, NJ, MI, WV, MD, IL, RI, VA, NC, IN, OK, NV, NE, WY, KY, WI, ND, MA, NH, UT, CO, GU, PR, AR, AZ, ME and the countries of Canada, Netherlands, United Kingdom, Sweden, Germany, France, Switzerland, Tunisia, Australia, Belgium, Costa Rica, Uruguay, Turkey, South Africa, New Zealand, Italy, Thailand, Ghana, Russia, Hungary, Czech Republic, Slovakia, India, Austria, Poland, Romania, Ireland, Spain, Portugal, Mexico, Vietnam, Korea, Mozambique, Kazakhstan, China, Malaysia, Singapore, Kenya, Brazil, Namibia, Egypt, Panama.

Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoFloridaGeorgiaIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMaineMichiganMissouriMississippiNorth Carolina

Questions about this recall

Why was it recalled?

The firm has identified that the Transferrin reagent does not meet the Instructions for Use (IFU) reagent on-board stability claims, which may cause control failures and/or erroneous patient samples results. The Complaint handling unit has received a total of 2 safety cases (CASE-2023-02027212 and CASE-2023-02045278) and 99 quality cases with the failure mode attributed to Transferrin reagent (OSR6152 / lot 2573) associated with reagent blank shifts for the Reagent 2 (R2), imprecision, calibration failures, erratic Quality Control (QC) recoveries, and/or erroneous high/low (serum, plasma, urine) patient results. The issue was confirmed through internal testing by Beckman Coulter on June 30, 2023. Reagent Blank shifts for R2, imprecision, calibration failures, erratic Quality Control (QC) recoveries, and/or erroneous high/low (serum, plasma, urine) patient results. The issue was confirmed through internal testing by Beckman Coulter on June 30, 2023.

Which lots are affected?

UDI-DI: 15099590010492, LOT: 2573, Expiration: 01 August 2024

Where was it sold?

Worldwide - US Nationwide distribution including in the states of NY, CA, MS, GA, PA, SC, WA, TN, AL, AK, OR, MO, KS, FL, LA, OH, TX, NJ, MI, WV, MD, IL, RI, VA, NC, IN, OK, NV, NE, WY, KY, WI, ND, MA, NH, UT, CO, GU, PR, AR, AZ, ME and the countries of Canada, Netherlands, United Kingdom, Sweden, Germany, France, Switzerland, Tunisia, Australia, Belgium, Costa Rica, Uruguay, Turkey, South Africa, New Zealand, Italy, Thailand, Ghana, Russia, Hungary, Czech Republic, Slovakia, India, Austria, Poland, Romania, Ireland, Spain, Portugal, Mexico, Vietnam, Korea, Mozambique, Kazakhstan, China, Malaysia, Singapore, Kenya, Brazil, Namibia, Egypt, Panama.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.