CLASS II ONGOING Device recall · Reported 12 Jan 2022 · FDA Enforcement Report

Aviva FX with LiNX Gyro Power Wheelchair (with LiNX Gyro Module recalled

Invacare is recalling Aviva FX with LiNX Gyro Power Wheelchair (with LiNX Gyro Module, distributed nationwide in the US. The recall was initiated on 12 Nov 2021 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLiNX Gyro firmware version 6.1.2 UDI-DI Numbers: IFX-20C 00841447111627 IFX-20MP 00841447111658 IFX-20R 00841447111634 IFX-20SP 00841447111641 IFX20WIDE 00841447111658
Quantity recalled303 units US; 20 units OUS

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0446-2022
FDA event ID
89062
Recalling firm
Invacare, Elyria, OH
Classification
Class II
Status
Ongoing
Recall initiated
12 Nov 2021
FDA classified
3 Jan 2022
Reported
12 Jan 2022
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

12 Nov 2021Recall initiated by company
3 Jan 2022Classified by FDA+52 days
12 Jan 2022Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile, resulting in potential injury

Product as listed by the FDA

Invacare AVIVA FX with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A) Models: IFX-20R, IFX-20SP, IFX-20MP, IFX-20C

Where it was distributed

Worldwide distribution - US Nationwide and the countries of Canada, Australia, Thailand, New Zealand.

Nationwide

Questions about this recall

Why was it recalled?

Power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile, resulting in potential injury

Which lots are affected?

LiNX Gyro firmware version 6.1.2 UDI-DI Numbers: IFX-20C 00841447111627 IFX-20MP 00841447111658 IFX-20R 00841447111634 IFX-20SP 00841447111641 IFX20WIDE 00841447111658

Where was it sold?

Worldwide distribution - US Nationwide and the countries of Canada, Australia, Thailand, New Zealand.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

2 other products · FDA event 89062
ReportedClassRecallCompanyReasonWhere
CLASS IITDX SP2 with LiNX Gyro Power Wheelchair (with LiNX Gyro Module recalledInvacareOtherNationwide
CLASS IIAviva Storm RX with LiNX Gyro Power Wheelchair recalled by InvacareInvacareOtherNationwide

More recalls from Invacare

All 35

Other Wheelchairs, Beds & Lifts recalls

All 175