CLASS I ONGOING Device recall · Reported 22 Apr 2026 · FDA Enforcement Report
B. Braun Hemodialysis Bloodlines recalled by B Braun Medical
B Braun Medical is recalling B. Braun Hemodialysis Bloodlines, Streamline Bloodline Set For Dialog, distributed nationwide in the US. The recall was initiated on 19 Mar 2026 and the FDA has classified it as Class I.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1799-2026
- FDA event ID
- 98552
- Recalling firm
- B Braun Medical, Bethlehem, PA
- Classification
- Class I
- Status
- Ongoing
- Recall initiated
- 19 Mar 2026
- FDA classified
- 15 Apr 2026
- Reported
- 22 Apr 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Other Medical Devices
Timeline
| 19 Mar 2026 | Recall initiated by company | |
| 15 Apr 2026 | Classified by FDA | +27 days |
| 22 Apr 2026 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.
Product as listed by the FDA
B. Braun Hemodialysis Bloodlines, STREAMLINE BLOODLINE SET FOR DIALOG. Model Number: SL-2010M2096
Where it was distributed
Worldwide distribution - US Nationwide and the country of Canada.
Questions about this recall
Is B. Braun Hemodialysis Bloodlines, Streamline Bloodline Set For Dialog recalled?
Yes. B Braun Medical recalled B. Braun Hemodialysis Bloodlines, Streamline Bloodline Set For Dialog on 19 Mar 2026. The recall is Class I and its status is Ongoing. Recall number Z-1799-2026.
Why was it recalled?
Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.
Which lots are affected?
Model Number: SL-2010M2096; UDI-DI Primary: 04046964367786; UDI-DI Unit: 04046964367779; All lots manufactured since 25MAY2025.
Where was it sold?
Worldwide distribution - US Nationwide and the country of Canada.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class I mean?
Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from B Braun Medical
All 447| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Extension Set Discofix Four Way Stopcock recalled by B Braun Medical | B Braun Medical | Other | Nationwide |
| Drug | ANNOUNCED | B. Braun 0.9% Sodium Chloride Injection USP, in a PAB Container recalledB. Braun | B. Braun Medical | Particulate Matter | Nationwide |
| Drug | CLASS I | Lactated Ringer's Injection USP Excel container recalledLactated Ringers | B Braun Medical | Particulate Matter | Nationwide |
| Device | CLASS I | Design Options Spinal Tray Spinal Needle Anesthesia recalled | B Braun Medical | Processing Deviation | Nationwide |
| Device | CLASS II | Sesk Upmc Double Catheter Tray recalled by B Braun Medical | B Braun Medical | Device Malfunction | Nationwide |
Other Other Medical Devices recalls
All 13,778| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Citric Complete Dry Citric Acid Concentrate recalled | Nipro Renal Soultions USA | Other | Nationwide |
| Device | CLASS I | Kit: Epidural/Pain TQC 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | Halyard IV Start Kit Non-Sterile recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS I | Kit: Single Shot Epidural 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | AMS Convenience Kit: Dsaek Pack - RX recalled | Windstone Medical Packaging | Device Malfunction | Nationwide |