CLASS II ONGOING Device recall · Reported 16 Feb 2022 · FDA Enforcement Report
Baxter Capd System Transfer Set (Capd MiniCap Transfer Set II) recalled
Baxter Healthcare is recalling Baxter Capd System Transfer Set (Capd MiniCap Transfer Set II). The recall was initiated on 7 Jan 2022 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0579-2022
- FDA event ID
- 89327
- Recalling firm
- Baxter Healthcare, Deerfield, IL
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 7 Jan 2022
- FDA classified
- 8 Feb 2022
- Reported
- 16 Feb 2022
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Packaging Defects
- Device type
- Other Medical Devices
Timeline
| 7 Jan 2022 | Recall initiated by company | |
| 8 Feb 2022 | Classified by FDA | +32 days |
| 16 Feb 2022 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Cleaning products, containing, but not limited to, hydrogen peroxide, bleach, alcohol, or antiseptic agents and solvent, such as those containing acetone, toluene, xylene, or cyclohexanone may cause damage, such as leaking or cracking, to the MiniCap Extended Life PD transfer sets.
Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.
Product as listed by the FDA
Baxter CAPD System Transfer Set (CAPD MiniCap Transfer Set II (24cm)), Code No. T5C4482
Where it was distributed
Worldwide
Questions about this recall
Is Baxter Capd System Transfer Set (Capd MiniCap Transfer Set II) recalled?
Yes. Baxter Healthcare recalled Baxter Capd System Transfer Set (Capd MiniCap Transfer Set II) on 7 Jan 2022. The recall is Class II and its status is Ongoing. Recall number Z-0579-2022.
Why was it recalled?
Cleaning products, containing, but not limited to, hydrogen peroxide, bleach, alcohol, or antiseptic agents and solvent, such as those containing acetone, toluene, xylene, or cyclohexanone may cause damage, such as leaking or cracking, to the MiniCap Extended Life PD transfer sets.
Which lots are affected?
All lot numbers within expiry
Where was it sold?
Worldwide
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Baxter Healthcare
All 564| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Drug | CLASS I | Dextrose Injection, USP, 70 % recalled by Baxter Healthcare | Baxter Healthcare | Particulate Matter | Nationwide |
| Drug | CLASS I | 9% Sodium Chloride Injection USP, Viaflex recalled by Baxter HealthcareSodium Chloride | Baxter Healthcare | Particulate Matter | IL |
| Device | CLASS II | Baxter Novum IQ Syringe Pump, Loaner recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Baxter Novum IQ Syringe Pump recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |
| Drug | CLASS II | Clindamycin Phosphate in 9% Sodium Chloride Injection recalledClindamycin Phosphate | Baxter Healthcare | Sterility | Nationwide |
Other Other Medical Devices recalls
All 13,778| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Citric Complete Dry Citric Acid Concentrate recalled | Nipro Renal Soultions USA | Other | Nationwide |
| Device | CLASS I | Kit: Epidural/Pain TQC 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | Halyard IV Start Kit Non-Sterile recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS I | Kit: Single Shot Epidural 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | AMS Convenience Kit: Dsaek Pack - RX recalled | Windstone Medical Packaging | Device Malfunction | Nationwide |