CLASS II ONGOING Device recall · Reported 16 Feb 2022 · FDA Enforcement Report
Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp recalled
Baxter Healthcare is recalling Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp. The recall was initiated on 7 Jan 2022 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0576-2022
- FDA event ID
- 89327
- Recalling firm
- Baxter Healthcare, Deerfield, IL
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 7 Jan 2022
- FDA classified
- 8 Feb 2022
- Reported
- 16 Feb 2022
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Packaging Defects
- Device type
- Other Medical Devices
Timeline
| 7 Jan 2022 | Recall initiated by company | |
| 8 Feb 2022 | Classified by FDA | +32 days |
| 16 Feb 2022 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Cleaning products, containing, but not limited to, hydrogen peroxide, bleach, alcohol, or antiseptic agents and solvent, such as those containing acetone, toluene, xylene, or cyclohexanone may cause damage, such as leaking or cracking, to the MiniCap Extended Life PD transfer sets.
Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.
Product as listed by the FDA
Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp, REF R5C4482E
Where it was distributed
Worldwide
Questions about this recall
Is Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp recalled?
Yes. Baxter Healthcare recalled Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp on 7 Jan 2022. The recall is Class II and its status is Ongoing. Recall number Z-0576-2022.
Why was it recalled?
Cleaning products, containing, but not limited to, hydrogen peroxide, bleach, alcohol, or antiseptic agents and solvent, such as those containing acetone, toluene, xylene, or cyclohexanone may cause damage, such as leaking or cracking, to the MiniCap Extended Life PD transfer sets.
Which lots are affected?
All lot numbers within expiry
Where was it sold?
Worldwide
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Baxter Healthcare
All 562| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Baxter Novum IQ Syringe Pump, Loaner recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Baxter Novum IQ Syringe Pump recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |
| Drug | CLASS II | Clindamycin Phosphate in 9% Sodium Chloride Injection recalledClindamycin Phosphate | Baxter Healthcare | Sterility | Nationwide |
| Drug | CLASS II | Vancomycin Injection, USP in 5% Dextrose per (5 mg/mL) recalledVancomycin Hydrochloride | Baxter Healthcare | Sterility | Nationwide |
| Drug | CLASS II | Clindamycin Injection USP in 5% Dextrose, 900 mg recalledClindamycin Phosphate | Baxter Healthcare | Sterility | Nationwide |
Other Other Medical Devices recalls
All 13,771| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Convenience Kits: Blood Draw Kit recalled by Medline Industries | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep Clear recalled | Medline Industries | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Ortho Aspiration Kit recalled | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep DYKA1054F recalled | Medline Industries | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Arterial Line Insertion Kit recalled | Medline Industries | Sterility | Nationwide |