CLASS I ONGOING Device recall · Reported 13 Aug 2025 · FDA Enforcement Report
BD Alaris Pump Module recalled by CareFusion 303
CareFusion 303 is recalling BD Alaris Pump Module, distributed nationwide in the US. The recall was initiated on 8 Jul 2025 and the FDA has classified it as Class I.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2179-2025
- FDA event ID
- 97067
- Recalling firm
- CareFusion 303, San Diego, CA
- Classification
- Class I
- Status
- Ongoing
- Recall initiated
- 8 Jul 2025
- FDA classified
- 6 Aug 2025
- Reported
- 13 Aug 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Medical Software
Timeline
| 8 Jul 2025 | Recall initiated by company | |
| 6 Aug 2025 | Classified by FDA | +29 days |
| 13 Aug 2025 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
BD Alaris Pump Module Model, 8100 with Guardrails, Suite MX software versions
Where it was distributed
US Nationwide Distribution to states of: AK AL AR AS AZ CA CO CT DC DE FL GA GU HI IA ID IL IN KS KY LA MA MD ME MI MN MO MP MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV WY; and OUS (foreign) to countries of: Japan, Italy, Germany, Turkey
Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaFloridaGeorgiaIowaIllinoisKansasKentuckyLouisianaMassachusettsMarylandMichiganMinnesotaMissouri
Questions about this recall
Is BD Alaris Pump Module recalled?
Yes. CareFusion 303 recalled BD Alaris Pump Module on 8 Jul 2025. The recall is Class I and its status is Ongoing. Recall number Z-2179-2025.
Why was it recalled?
Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.
Which lots are affected?
BD Alaris" Pump Module Model 8100- All Serial Numbers/ UDI(s): 10885403222054, 10885403517723, 10885403810015, 10885403810039, 10885403810046
Where was it sold?
US Nationwide Distribution to states of: AK AL AR AS AZ CA CO CT DC DE FL GA GU HI IA ID IL IN KS KY LA MA MD ME MI MN MO MP MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV WY; and OUS (foreign) to countries of: Japan, Italy, Germany, Turkey
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class I mean?
Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from CareFusion 303
All 143| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | BD Pyxis MedStation ES Tower recalled by CareFusion 303 | CareFusion 303 | Temperature Abuse | Nationwide |
| Device | CLASS II | BD Pyxis MedBank Mini recalled by CareFusion 303 | CareFusion 303 | Temperature Abuse | Nationwide |
| Device | CLASS II | BD Pyxis MedStation recalled by CareFusion 303 | CareFusion 303 | Temperature Abuse | Nationwide |
| Device | CLASS II | BD Pyxis Anesthesia Station ES recalled by CareFusion 303 | CareFusion 303 | Temperature Abuse | Nationwide |
| Device | CLASS II | BD Pyxis MedStation ES recalled by CareFusion 303 | CareFusion 303 | Temperature Abuse | Nationwide |
Other Medical Software recalls
All 903| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Philips Brilliance recalled by Philips North America | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Philips IQon Spectral CT Product Codes recalled by Philips North America | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Philips Ingenuity CT Product Codes recalled by Philips North America | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Philips Brilliance CT 64: Brilliance CT 64 Channel Product recalled | Philips North America | Device Malfunction | MA |
| Device | ANNOUNCED | BMC Luna G3 APAP recalled by BMC MedicalBMC | BMC Medical | Device Malfunction | Nationwide |