CLASS I ONGOING Device recall · Reported 13 Aug 2025 · FDA Enforcement Report

BD Alaris Pump Module recalled by CareFusion 303

CareFusion 303 is recalling BD Alaris Pump Module, distributed nationwide in the US. The recall was initiated on 8 Jul 2025 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesBD Alaris" Pump Module Model 8100- All Serial Numbers/ UDI(s): 10885403222054, 10885403517723, 10885403810015, 10885403810039, 10885403810046
Quantity recalled33,162,680 units total

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2179-2025
FDA event ID
97067
Recalling firm
CareFusion 303, San Diego, CA
Classification
Class I
Status
Ongoing
Recall initiated
8 Jul 2025
FDA classified
6 Aug 2025
Reported
13 Aug 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction
Device type
Medical Software

Timeline

8 Jul 2025Recall initiated by company
6 Aug 2025Classified by FDA+29 days
13 Aug 2025Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

BD Alaris Pump Module Model, 8100 with Guardrails, Suite MX software versions

Where it was distributed

US Nationwide Distribution to states of: AK AL AR AS AZ CA CO CT DC DE FL GA GU HI IA ID IL IN KS KY LA MA MD ME MI MN MO MP MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV WY; and OUS (foreign) to countries of: Japan, Italy, Germany, Turkey

Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaFloridaGeorgiaIowaIllinoisKansasKentuckyLouisianaMassachusettsMarylandMichiganMinnesotaMissouri

Questions about this recall

Is BD Alaris Pump Module recalled?

Yes. CareFusion 303 recalled BD Alaris Pump Module on 8 Jul 2025. The recall is Class I and its status is Ongoing. Recall number Z-2179-2025.

Why was it recalled?

Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.

Which lots are affected?

BD Alaris" Pump Module Model 8100- All Serial Numbers/ UDI(s): 10885403222054, 10885403517723, 10885403810015, 10885403810039, 10885403810046

Where was it sold?

US Nationwide Distribution to states of: AK AL AR AS AZ CA CO CT DC DE FL GA GU HI IA ID IL IN KS KY LA MA MD ME MI MN MO MP MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV WY; and OUS (foreign) to countries of: Japan, Italy, Germany, Turkey

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from CareFusion 303

All 143

Other Medical Software recalls

All 903