CLASS II ONGOING Device recall · Reported 2 Jul 2025 · FDA Enforcement Report

BD Bactec Mgit 960 Pza Kit recalled by Becton Dickinson &

Becton Dickinson is recalling BD Bactec Mgit 960 Pza Kit, distributed nationwide in the US. The recall was initiated on 20 May 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesCatalog No. 245128; UDI-DI: (01) 0038290245128; Lot No. 4150998, 4177908, 4178500, 4262031, 4262036, 4262037, 4262039, 4262040, 4262044, 4284438, 4284441, 4284443, 4284445, 4284449, 4304789, 4304796, 4347228, 4347230, 4352110, 4362418, 5007456, 5020153, 5050670, 5050671, 5050672, 5050675, 5050678.
Quantity recalled12,256 eaches

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2023-2025
FDA event ID
96910
Recalling firm
Becton Dickinson, Sparks, MD
Classification
Class II
Status
Ongoing
Recall initiated
20 May 2025
FDA classified
25 Jun 2025
Reported
2 Jul 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Sterility

Timeline

20 May 2025Recall initiated by company
25 Jun 2025Classified by FDA+36 days
2 Jul 2025Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

BD has received additional complaints of intermittent false resistance results for PZA during susceptibility testing of Mycobacterium tuberculosis isolates.

A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.

Product as listed by the FDA

BD BACTEC MGIT 960 PZA Kit. IVD qualitative procedure for susceptibility testing of Mycobacterium tuberculosis, from culture to pyrazinamide (PZA).

Where it was distributed

Worldwide distribution - US Nationwide and the countries of AR, AU, BD, BE, BN, BR, CA, CN, CR, DO, GT, HK, HT, ID, IN, JP, KH, KR, MO, MX, MY, NP, NZ, PE, PH, PK, SG, SV, TH, TW, VN.

Nationwide ArkansasCaliforniaIdahoIndianaMissouri

Questions about this recall

Why was it recalled?

BD has received additional complaints of intermittent false resistance results for PZA during susceptibility testing of Mycobacterium tuberculosis isolates.

Which lots are affected?

Catalog No. 245128; UDI-DI: (01) 0038290245128; Lot No. 4150998, 4177908, 4178500, 4262031, 4262036, 4262037, 4262039, 4262040, 4262044, 4284438, 4284441, 4284443, 4284445, 4284449, 4304789, 4304796, 4347228, 4347230, 4352110, 4362418, 5007456, 5020153, 5050670, 5050671, 5050672, 5050675, 5050678.

Where was it sold?

Worldwide distribution - US Nationwide and the countries of AR, AU, BD, BE, BN, BR, CA, CN, CR, DO, GT, HK, HT, ID, IN, JP, KH, KR, MO, MX, MY, NP, NZ, PE, PH, PK, SG, SV, TH, TW, VN.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Becton Dickinson