CLASS II Device recall · Reported 29 Aug 2018 · FDA Enforcement Report
BD BBL GC-Lect Agar recalled by Becton Dickinson &
Becton Dickinson is recalling BD BBL GC-Lect Agar, distributed in 29 states. The recall was initiated on 21 Jun 2018 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2869-2018
- FDA event ID
- 80573
- Recalling firm
- Becton Dickinson, Sparks, MD
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 21 Jun 2018
- FDA classified
- 17 Aug 2018
- Reported
- 29 Aug 2018
- Type
- Voluntary: Firm initiated
- Customers notified by
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Reason category
- Other Reasons
- Device type
- Other Medical Devices
Timeline
| 21 Jun 2018 | Recall initiated by company | |
| 17 Aug 2018 | Classified by FDA | +57 days |
| 29 Aug 2018 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
A portion of this lot was manufactured using Gentamicin instead of Vancomycin. Gentamicin is an inhibitory agent of Neisseria organisms. This product is a selective plated medium to provide enhanced growth and recovery of N gonorrhoeae. Inhibited growth may cause a false negative result or a delayed result leading to incorrect or delayed treatment.
Product as listed by the FDA
BD BBL GC-Lect Agar; Cat. No. 297715 The device is a selective medium providing enhanced growth and recovery of Neisseria gonorrhoeae and better inhibition of contamination bacteria and fungi.
Where it was distributed
US Distribution in states of: AK, AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, ID, IN, KS, KY, MA, MI, MS, MT, NC, JH, NY, OK, OR, PA, SC, TN, TX, WA, and WI.
AlaskaAlabamaArkansasArizonaCaliforniaColoradoDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIdahoIndianaKansasKentuckyMassachusettsMichiganMississippiMontanaNorth Carolina
Questions about this recall
Why was it recalled?
A portion of this lot was manufactured using Gentamicin instead of Vancomycin. Gentamicin is an inhibitory agent of Neisseria organisms. This product is a selective plated medium to provide enhanced growth and recovery of N gonorrhoeae. Inhibited growth may cause a false negative result or a delayed result leading to incorrect or delayed treatment.
Which lots are affected?
8110781
Where was it sold?
US Distribution in states of: AK, AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, ID, IN, KS, KY, MA, MI, MS, MT, NC, JH, NY, OK, OR, PA, SC, TN, TX, WA, and WI.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Becton Dickinson
All 364| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | BD¿Spinal Tray with BD¿Whitacre Needle recalled by Becton Dickinson & | Becton Dickinson | Processing Deviation | Nationwide |
| Device | CLASS I | BD¿Spinal Tray with BD¿Whitacre Needle recalled by Becton Dickinson & | Becton Dickinson | Processing Deviation | Nationwide |
| Device | CLASS I | BD¿Spinal Tray with BD¿Quincke Needle recalled by Becton Dickinson & | Becton Dickinson | Processing Deviation | Nationwide |
| Device | CLASS I | BD¿Spinal Tray with BD¿Whitacre Needle recalled by Becton Dickinson & | Becton Dickinson | Processing Deviation | Nationwide |
| Device | CLASS I | BD¿Spinal Tray with BD¿Whitacre Needle, BD¿Quincke Needle recalled | Becton Dickinson | Processing Deviation | Nationwide |
Other Other Medical Devices recalls
All 13,778| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Citric Complete Dry Citric Acid Concentrate recalled | Nipro Renal Soultions USA | Other | Nationwide |
| Device | CLASS I | Kit: Epidural/Pain TQC 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | Halyard IV Start Kit Non-Sterile recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS I | Kit: Single Shot Epidural 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |