CLASS II ONGOING Device recall · Reported 11 Dec 2024 · FDA Enforcement Report

BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) recalled by Becton Dickinson &

Becton Dickinson is recalling BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) - In-Vitro BD BBL Sensi Disc Ampicillin 2 ¿g, distributed nationwide in the US. The recall was initiated on 6 Nov 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDi-DI: (01)30382902312636 Lot No. (Exp. Date): 2339360 (12/31/2024), 3010977 (01/31/25), 3058508 (03/31/2025), 3184064 (07/02/2025), 3234190 (08/21/2025).
Quantity recalled13,190 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0621-2025
FDA event ID
95646
Recalling firm
Becton Dickinson, Sparks, MD
Classification
Class II
Status
Ongoing
Recall initiated
6 Nov 2024
FDA classified
5 Dec 2024
Reported
11 Dec 2024
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Potency & Assay

Timeline

6 Nov 2024Recall initiated by company
5 Dec 2024Classified by FDA+29 days
11 Dec 2024Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

BD identified through potency testing as part of a stability test request to monitor Ampicillin AM-2 due to decrease in potency results of 65% at 18 months, may result in falsely resistant result for ampicillin susceptibility

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) - In-Vitro BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) are used for semi-quantitative in vitro susceptibility testing by the agar disc diffusion test procedure of common, rapidly growing and certain fastidious bacterial pathogens. Catalog Number: 231263

Where it was distributed

Worldwide - US Nationwide distribution in the states of CA, GA, IL, KY, MA, MD, NE, PA and the countries of Canada, Belgium, Singapore, Malaysia.

Nationwide CaliforniaGeorgiaIllinoisKentuckyMassachusettsMarylandNebraskaPennsylvania

Questions about this recall

Why was it recalled?

BD identified through potency testing as part of a stability test request to monitor Ampicillin AM-2 due to decrease in potency results of 65% at 18 months, may result in falsely resistant result for ampicillin susceptibility

Which lots are affected?

UDi-DI: (01)30382902312636 Lot No. (Exp. Date): 2339360 (12/31/2024), 3010977 (01/31/25), 3058508 (03/31/2025), 3184064 (07/02/2025), 3234190 (08/21/2025).

Where was it sold?

Worldwide - US Nationwide distribution in the states of CA, GA, IL, KY, MA, MD, NE, PA and the countries of Canada, Belgium, Singapore, Malaysia.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Becton Dickinson