CLASS II ONGOING Device recall · Reported 7 Oct 2026 · FDA Enforcement Report
Systems/units containing affected recalled
Beckman Coulter Mishima K.K is recalling Systems/units containing affected, distributed nationwide in the US. The recall was initiated on 3 Aug 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-3332-2026
- FDA event ID
- 99588
- Recalling firm
- Beckman Coulter Mishima K.K, Sunto-Gun
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 3 Aug 2026
- FDA classified
- 30 Sep 2026
- Reported
- 7 Oct 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Test Kits & Diagnostics
Timeline
| 3 Aug 2026 | Recall initiated by company | |
| 30 Sep 2026 | Classified by FDA | +58 days |
| 7 Oct 2026 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. This causes a delay in obtaining patient results.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 REF / UDI-DI code: B96692 / 14987666542965 B96693 / 14987666542972 B12183 / 14987666540671 The AU480 Chemistry Analyzer measure analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material, and other accessories. This system is for in vitro diagnostic use only.
Where it was distributed
Worldwide - U.S. Nationwide distribution in the states of AR, CA, DE, FL, GA, GU, HI, IA, IL, IN, KS, KY, MN, MO, NC, NM, NY, OK, PA, SD, TX, and WI. The countries of Algeria, Australia, Belarus, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, China, Costa Rica, Croatia, Cyprus, Czechia, Egypt, France, Germany, Guatemala, Hungary, India, Ireland, Italy, Japan, Kazakhstan, Korea, Republic of, Latvia, Malaysia, Morocco, Myanmar, Nepal, Nigeria, Pakistan, Poland, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, South Africa, Taiwan, Province of China, Turkey, United Arab Emirates, and Viet Nam.
Nationwide ArkansasCaliforniaDelawareFloridaGeorgiaHawaiiIowaIllinoisIndianaKansasKentuckyMinnesotaMissouriNorth CarolinaNew MexicoNew YorkOklahomaPennsylvaniaSouth DakotaTexas
Questions about this recall
Why was it recalled?
Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. This causes a delay in obtaining patient results.
Which lots are affected?
Products containing affected software: Item Numbers / UDI-DI codes: Item Number: B96692 UDI-DI code: 14987666542965 Item Number: B96693 UDI-DI code: 14987666542972 Item Number: B12183 UDI-DI code: 14987666540671 Item Number: N3151400 UDI-DI code: Due to the age of the device and the fact that it has exceeded its storage period, the UDI number is not available.)
Where was it sold?
Worldwide - U.S. Nationwide distribution in the states of AR, CA, DE, FL, GA, GU, HI, IA, IL, IN, KS, KY, MN, MO, NC, NM, NY, OK, PA, SD, TX, and WI. The countries of Algeria, Australia, Belarus, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, China, Costa Rica, Croatia, Cyprus, Czechia, Egypt, France, Germany, Guatemala, Hungary, India, Ireland, Italy, Japan, Kazakhstan, Korea, Republic of, Latvia, Malaysia, Morocco, Myanmar, Nepal, Nigeria, Pakistan, Poland, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, South Africa, Taiwan, Province of China, Turkey, United Arab Emirates, and Viet Nam.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 13 Oct 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Beckman Coulter Mishima K.K
All 7| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | DxC 700 AU recalled by Beckman Coulter Mishima K.K | Beckman Coulter Mishima K.K | Device Malfunction | Nationwide |
| Device | CLASS II | Chemistry Analyzer Au5800 recalled by Beckman Coulter Mishima K.K | Beckman Coulter Mishima K.K | Device Malfunction | Nationwide |
| Device | CLASS II | Sample probe sucks a sample dispensed into a tube or cup recalled | Beckman Coulter Mishima K.K | Device Malfunction | Nationwide |
| Device | CLASS II | Sample probe sucks a sample dispensed into a tube or cup recalled | Beckman Coulter Mishima K.K | Device Malfunction | Nationwide |
| Device | CLASS II | DxC 500 AU Clinical Chemistry Analyzer recalled | Beckman Coulter Mishima K.K | Device Malfunction | Nationwide |
Other Test Kits & Diagnostics recalls
All 2,661| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | McKesson Consult hCG Urine Cassette recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |
| Device | CLASS II | Fisher Scientific Sure-Vue Urine hCG recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |
| Device | CLASS II | Vitek 2 Ast-N461 Test Kit, Gram Negative Susceptibility recalled | Biomerieux | Other | MO |
| Device | CLASS II | DASH Rapid PCR Instrument recalled by Nuclein | Nuclein | Device Malfunction | Nationwide |
| Device | CLASS II | STA Staclot Protein recalled by Diagnostica Stago | Diagnostica Stago | Potency & Assay | Nationwide |