CLASS II ONGOING Device recall · Reported 26 Mar 2025 · FDA Enforcement Report
BD Pyxis: Anesthesia Station ES recalled by CareFusion 303
CareFusion 303 is recalling BD Pyxis: Anesthesia Station ES, distributed nationwide in the US. The recall was initiated on 6 Feb 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1375-2025
- FDA event ID
- 96202
- Recalling firm
- CareFusion 303, San Diego, CA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 6 Feb 2025
- FDA classified
- 17 Mar 2025
- Reported
- 26 Mar 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Medical Software
Timeline
| 6 Feb 2025 | Recall initiated by company | |
| 17 Mar 2025 | Classified by FDA | +39 days |
| 26 Mar 2025 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Sync timeout may occur between the server and automated dispensing device, due to a server software error which may occur if syncing a large amount of data that takes excessive time, which may cause the dispensing devices to become disconnected from the ES server; dispensing devices will not receive updated information, which may lead to incorrect therapy, and delay in accessing medication.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
BD Pyxis: Anesthesia Station ES, REF: 327, MedStation ES, REF: 323, MedStation ES Tower, REF: 352, CII Safe ES, REF: 1116-00, CII Safe ES Desktop PC, REF: 1156-00, CII Safe ES Desktop PC, Wired SCNR, REF: 107-254-01, CII Safe ES Desktop PC, W-less SCNR, REF: 107-255-01, Used with Enterprise Server (ES), REF: 1115-00 with affected Enterprise Server software.
Where it was distributed
Worldwide - US Nationwide distribution including in the states of KY, OH, NY, MO, NC, PA, FL, TN, CT, LA, MN, VA, MD, CA, TX, NE, IA, OK, MI, PR, SD, IN, ME, KS, IL, NM, WA, AR, HI, MS, GA, WY, WI, VT, AL, NV, NJ, DC, AZ, MT, AK, OR, CO, NH, MA, SC, ID, UT, ND and the countries of TW, CA, AE, MX, SG, AU, NZ, GB, CH, BH.
Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMaryland
Questions about this recall
Why was it recalled?
Sync timeout may occur between the server and automated dispensing device, due to a server software error which may occur if syncing a large amount of data that takes excessive time, which may cause the dispensing devices to become disconnected from the ES server; dispensing devices will not receive updated information, which may lead to incorrect therapy, and delay in accessing medication.
Which lots are affected?
BD Pyxis Enterprise Server software versions: 1.7.4.128, 1.7.4.134, 1.7.4.136, 1.7.4.137, All 1.7.1, All 1.7.2, All 1.7.3. REF/UDI-DI: 327/10885403477836, 323/10885403512667, 352/10885403512674, 1116-00/10885403512605
Where was it sold?
Worldwide - US Nationwide distribution including in the states of KY, OH, NY, MO, NC, PA, FL, TN, CT, LA, MN, VA, MD, CA, TX, NE, IA, OK, MI, PR, SD, IN, ME, KS, IL, NM, WA, AR, HI, MS, GA, WY, WI, VT, AL, NV, NJ, DC, AZ, MT, AK, OR, CO, NH, MA, SC, ID, UT, ND and the countries of TW, CA, AE, MX, SG, AU, NZ, GB, CH, BH.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from CareFusion 303
All 143| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | BD Pyxis MedStation ES Tower recalled by CareFusion 303 | CareFusion 303 | Temperature Abuse | Nationwide |
| Device | CLASS II | BD Pyxis MedBank Mini recalled by CareFusion 303 | CareFusion 303 | Temperature Abuse | Nationwide |
| Device | CLASS II | BD Pyxis MedStation recalled by CareFusion 303 | CareFusion 303 | Temperature Abuse | Nationwide |
| Device | CLASS II | BD Pyxis Anesthesia Station ES recalled by CareFusion 303 | CareFusion 303 | Temperature Abuse | Nationwide |
| Device | CLASS II | BD Pyxis MedStation ES recalled by CareFusion 303 | CareFusion 303 | Temperature Abuse | Nationwide |
Other Medical Software recalls
All 908| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Volara System Volara, HC, NA recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Enterprise Archive V8 recalled by GE Healthcare | GE Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Veriqa RT MonteCarlo 3D v3 recalled by PTW Freiburg | PTW Freiburg | Device Malfunction | Nationwide |
| Device | CLASS II | Centricity Enterprise Archive V2, V3, and V4 recalled by GE Healthcare | GE Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Et Control recalled by Datex-Ohmeda | Datex-Ohmeda | Device Malfunction | Nationwide |