CLASS II Device recall · Reported 8 Apr 2020 · FDA Enforcement Report
Berchtold Chromophare Ceiling Mounted Surgical Light System recalled
Stryker Communications is recalling Berchtold Chromophare Ceiling Mounted Surgical Light System, distributed nationwide in the US. The recall was initiated on 11 Oct 2019 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1602-2020
- FDA event ID
- 85083
- Recalling firm
- Stryker Communications, Flower Mound, TX
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 11 Oct 2019
- FDA classified
- 27 Mar 2020
- Reported
- 8 Apr 2020
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Surgical Instruments
Timeline
| 11 Oct 2019 | Recall initiated by company | |
| 27 Mar 2020 | Classified by FDA | +168 days |
| 8 Apr 2020 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The firm has become aware that there is a likelihood that a missing snap ring or spanner nut may result in an insufficient mount force that causes the surgical light to fall. If the suspension of the surgical light falls, it may potentially cause serious injuries to Health Care Providers and/or patients.
Product as listed by the FDA
Berchtold Chromophare Ceiling Mounted Surgical Light System, Catalog #CH13000000
Where it was distributed
US: Nationwide OUS: Canada
Questions about this recall
Why was it recalled?
The firm has become aware that there is a likelihood that a missing snap ring or spanner nut may result in an insufficient mount force that causes the surgical light to fall. If the suspension of the surgical light falls, it may potentially cause serious injuries to Health Care Providers and/or patients.
Which lots are affected?
Serial Number Range: 091316 70016 through 083117-73252
Where was it sold?
US: Nationwide OUS: Canada
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Stryker Communications
All 20| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Stryker Chromophare Softlit Ring Surgical Light System recalled | Stryker Communications | Device Malfunction | Nationwide |
| Device | CLASS II | Chromophare Surgical Light System recalled by Stryker Communications | Stryker Communications | Other | Nationwide |
| Device | CLASS II | Chromophare Surgical Light System recalled by Stryker Communications | Stryker Communications | Potency & Assay | TX |
| Device | CLASS II | S-Series Equipment Management System recalled by Stryker Communications | Stryker Communications | Other | Nationwide |
| Device | CLASS II | Berchtold Chromophare F528 Surgical Lights recalled | Stryker Communications | Other | Nationwide |
Other Surgical Instruments recalls
All 3,921| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Sterile, synthetic, absorbable recalled by Henry Schein | Henry Schein | Other | Nationwide |
| Device | CLASS II | TruSystem 7500 surgical table recalled by Baxter Healthcare | Baxter Healthcare | Other | Nationwide |
| Device | CLASS II | Sterile, natural, nonabsorbable recalled by Henry Schein | Henry Schein | Other | Nationwide |
| Device | CLASS II | Sterile, natural, nonabsorbable recalled by Henry Schein | Henry Schein | Other | Nationwide |
| Device | CLASS II | Sterile, non-absorbable 7-0 blue polypropylene surgical recalled | Henry Schein | Other | Nationwide |