CLASS II Device recall · Reported 8 Apr 2020 · FDA Enforcement Report

Berchtold Chromophare Ceiling Mounted Surgical Light System recalled

Stryker Communications is recalling Berchtold Chromophare Ceiling Mounted Surgical Light System, distributed nationwide in the US. The recall was initiated on 11 Oct 2019 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSerial Number Range: 091316 70016 through 083117-73252
Quantity recalled1415 Surgical Lights

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1602-2020
FDA event ID
85083
Recalling firm
Stryker Communications, Flower Mound, TX
Classification
Class II
Status
Terminated
Recall initiated
11 Oct 2019
FDA classified
27 Mar 2020
Reported
8 Apr 2020
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

11 Oct 2019Recall initiated by company
27 Mar 2020Classified by FDA+168 days
8 Apr 2020Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The firm has become aware that there is a likelihood that a missing snap ring or spanner nut may result in an insufficient mount force that causes the surgical light to fall. If the suspension of the surgical light falls, it may potentially cause serious injuries to Health Care Providers and/or patients.

Product as listed by the FDA

Berchtold Chromophare Ceiling Mounted Surgical Light System, Catalog #CH13000000

Where it was distributed

US: Nationwide OUS: Canada

Nationwide

Questions about this recall

Why was it recalled?

The firm has become aware that there is a likelihood that a missing snap ring or spanner nut may result in an insufficient mount force that causes the surgical light to fall. If the suspension of the surgical light falls, it may potentially cause serious injuries to Health Care Providers and/or patients.

Which lots are affected?

Serial Number Range: 091316 70016 through 083117-73252

Where was it sold?

US: Nationwide OUS: Canada

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Stryker Communications

All 20

Other Surgical Instruments recalls

All 3,921
ReportedClassRecallCompanyReasonWhere
DeviceCLASS IISterile, synthetic, absorbable recalled by Henry ScheinHenry ScheinOtherNationwide
DeviceCLASS IITruSystem 7500 surgical table recalled by Baxter HealthcareBaxter HealthcareOtherNationwide
DeviceCLASS IISterile, natural, nonabsorbable recalled by Henry ScheinHenry ScheinOtherNationwide
DeviceCLASS IISterile, natural, nonabsorbable recalled by Henry ScheinHenry ScheinOtherNationwide
DeviceCLASS IISterile, non-absorbable 7-0 blue polypropylene surgical recalledHenry ScheinOtherNationwide