CLASS II ONGOING Device recall · Reported 24 Nov 2021 · FDA Enforcement Report

Bio-rad BioPlex recalled by Bio-Rad Laboratories

Bio-Rad Laboratories is recalling Bio-rad BioPlex 2200, distributed nationwide in the US. The recall was initiated on 17 Sep 2021 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesCatalog Number: 12000650 UDI Code: 03610520653012 Lot Number: 301210
Quantity recalled11,610 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0235-2022
FDA event ID
88874
Recalling firm
Bio-Rad Laboratories, Redmond, WA
Classification
Class II
Status
Ongoing
Recall initiated
17 Sep 2021
FDA classified
16 Nov 2021
Reported
24 Nov 2021
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Potency & Assay

Timeline

17 Sep 2021Recall initiated by company
16 Nov 2021Classified by FDA+60 days
24 Nov 2021Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Due to two issues: 1) Customer complaints associated with greater than expected Rapid Plasma Reagin (RPR) reactivity which does not confirm with treponemal antibody test which may be due to a positive correlation between the mass COVID vaccinations and false reactive RPR results, and 2) During routine stability monitoring, an increase in mean bias above specification was observed which can cause elevated RPR reactivity potentially due to early shelf-life failure.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

BIO-RAD BioPlex 2200 REF 12000650 SYPH T PACK Syphilis Total & RPR Reagent Pack The BioPlex Syphilis Total & RPR kit is a multiplex flow immunoassay intended for the qualitative detection of total (IgG/IgM) antibodies to Treponema pallidum and the qualitative detection and/or titer determination of nontreponemal reagin antibodies in human serum or plasma

Where it was distributed

Worldwide Distribution: U.S.(nationwide) including states of: AK, AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, and WV; O.U.S.(foreign) countries of: France, Germany, Israel, Italy, Russian Federation, South Africa, Spain, and Taiwan.

Nationwide AlaskaAlabamaArizonaCaliforniaColoradoConnecticutFloridaGeorgiaHawaiiIowaIllinoisIndianaKansasKentuckyLouisianaMassachusettsMaineMichiganMinnesotaMissouri

Questions about this recall

Is Bio-rad BioPlex 2200 recalled?

Yes. Bio-Rad Laboratories recalled Bio-rad BioPlex 2200 on 17 Sep 2021. The recall is Class II and its status is Ongoing. Recall number Z-0235-2022.

Why was it recalled?

Due to two issues: 1) Customer complaints associated with greater than expected Rapid Plasma Reagin (RPR) reactivity which does not confirm with treponemal antibody test which may be due to a positive correlation between the mass COVID vaccinations and false reactive RPR results, and 2) During routine stability monitoring, an increase in mean bias above specification was observed which can cause elevated RPR reactivity potentially due to early shelf-life failure.

Which lots are affected?

Catalog Number: 12000650 UDI Code: 03610520653012 Lot Number: 301210

Where was it sold?

Worldwide Distribution: U.S.(nationwide) including states of: AK, AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, and WV; O.U.S.(foreign) countries of: France, Germany, Israel, Italy, Russian Federation, South Africa, Spain, and Taiwan.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Bio-Rad Laboratories

All 27

Other Test Kits & Diagnostics recalls

All 2,656