CLASS II Device recall · Reported 26 Sep 2012 · FDA Enforcement Report

The Bio-Rad Monolisa Anti-HBs EIA recalled by Bio-Rad Laboratories

Bio-Rad Laboratories is recalling The Bio-Rad Monolisa Anti-HBs EIA is for In Vitro Diagnostic Use and it has the, distributed nationwide in the US. The recall was initiated on 27 Jul 2012 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesAll lots distributed from May 5, 2007
Quantity recalled7,400 (approximately)- All lots since distribution date May 9, 2007

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2398-2012
FDA event ID
62850
Recalling firm
Bio-Rad Laboratories, Redmond, WA
Classification
Class II
Status
Terminated
Recall initiated
27 Jul 2012
FDA classified
17 Sep 2012
Reported
26 Sep 2012
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Potency & Assay

Timeline

27 Jul 2012Recall initiated by company
17 Sep 2012Classified by FDA+52 days
26 Sep 2012Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Prolonged exposure of the MONOLISA Anti-HBs EIA Conjugate Diluent component to light may result in elevated Optical Densities (OD) in some samples such that a negative sample could elevate to a grey zone or low positive result.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

The Bio-Rad MONOLISA Anti-HBs EIA is for In Vitro Diagnostic Use and it has the catalog number 25220. It is a qualitative and quantitative enzyme immunoassay for the detection of antibody for hepatitis B surface antigen in human serum and EDTA or citrated plasma. The Bio-Rad MONOLISA Anti-HBs EIA is a qualitative and quantitative enzyme immunoassay for the detection of antibody for hepatitis B surface antigen in human serum and EDTA or citrated plasma. The assay results may be used as an aid in the determination of susceptibility to hepatitis B virus (HBV) infection in individuals prior to or following HBV vaccination or where vaccination status is unknown. Assay results may be used with other HBV serological markers for the laboratory diagnosis of HBV disease associated with HBV infection.

Where it was distributed

Worldwide distribution: USA (nationwide) and countries of: Canada, South Africa, and Zimbabwe.

Nationwide

Questions about this recall

Why was it recalled?

Prolonged exposure of the MONOLISA Anti-HBs EIA Conjugate Diluent component to light may result in elevated Optical Densities (OD) in some samples such that a negative sample could elevate to a grey zone or low positive result.

Which lots are affected?

All lots distributed from May 5, 2007

Where was it sold?

Worldwide distribution: USA (nationwide) and countries of: Canada, South Africa, and Zimbabwe.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.