CLASS II ONGOING Device recall · Reported 14 Aug 2024 · FDA Enforcement Report
Biofire Filmarray Gastrointestinal (GI) Panel recalled
BioFire Diagnostics is recalling Biofire Filmarray Gastrointestinal (GI) Panel, distributed nationwide in the US. The recall was initiated on 16 May 2024 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2514-2024
- FDA event ID
- 94992
- Recalling firm
- BioFire Diagnostics, Salt Lake City, UT
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 16 May 2024
- FDA classified
- 6 Aug 2024
- Reported
- 14 Aug 2024
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Bacterial Contamination
- Device type
- Test Kits & Diagnostics
Timeline
| 16 May 2024 | Recall initiated by company | |
| 6 Aug 2024 | Classified by FDA | +82 days |
| 14 Aug 2024 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Due to increase of false positive Vibrio/Vibrio Cholerae results when using a certain transport medium.
Product as listed by the FDA
BIOFIRE FILMARRAY Gastrointestinal (GI) Panel REF RFIT-ASY-0116 (30 test kit) BIOFIRE FILMARRAY Gastrointestinal (GI) Panel REF RFIT-ASY-0104 (6 kit test) The BIOFIRE FILMARRAY Gastrointestinal (GI) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with BIOFIRE¿ FILMARRAY¿ Systems. The BIOFIRE GI Panel is capable of the simultaneous detection and identification of nucleic acids from multiple bacteria, viruses, and parasites directly from stool samples in Cary Blair transport media obtained from individuals with signs and/or symptoms of gastrointestinal infection.
Where it was distributed
U.S.: Nationwide - AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, Virgin Island, VT, WA, WI, WV, and WY. O.U.S.: Worldwide - AR, AU, AT, BE, BR, CA, CL, HK, CN, CO, CZ, DE, FI, FR, GR, HU, IN, IT, JP, MX, NL, NO, PH, PL, PT, SG, ZA, KR, ES, SE, CH, TW, TH, TR, GB, AL, DZ, AO, AM, BH, BY, BO, BA, BW, BG, CR, HR, CY, CD, DO, AN, EC, SV, EE, FJ, GE, GU, GT, GN, HT, HN, IS, ID, IQ, IE, IL, CI, CI, JO, KE, KW, LV, LB, LT, MO, MY, ML, MR, MD, MN, MA, MM, NP, NZ, NZ, NI, OM, PK, PA, PG, PY, PE, QA, RO, SA, RS, SI, TN, TR, UY, AE, VE, VN, and ZW.
Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusetts
Questions about this recall
Why was it recalled?
Due to increase of false positive Vibrio/Vibrio Cholerae results when using a certain transport medium.
Which lots are affected?
Part Number: RFIT-ASY-0116 (30 test kit) UDI-DI code: 00815381020109 0081538102011: All unexpired products are potentially impacted if used in conjunction with the Remel Cary Blair transport medium Part Number: RFIT-ASY-0104 (6 test kit) UDI-DI code: 0081538102011 All unexpired products are potentially impacted if used in conjunction with the Remel Cary Blair transport medium Remel Cary-Blair Transport Medium lot numbers: 743930 769240 712258 782425 732357 742049
Where was it sold?
U.S.: Nationwide - AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, Virgin Island, VT, WA, WI, WV, and WY. O.U.S.: Worldwide - AR, AU, AT, BE, BR, CA, CL, HK, CN, CO, CZ, DE, FI, FR, GR, HU, IN, IT, JP, MX, NL, NO, PH, PL, PT, SG, ZA, KR, ES, SE, CH, TW, TH, TR, GB, AL, DZ, AO, AM, BH, BY, BO, BA, BW, BG, CR, HR, CY, CD, DO, AN, EC, SV, EE, FJ, GE, GU, GT, GN, HT, HN, IS, ID, IQ, IE, IL, CI, CI, JO, KE, KW, LV, LB, LT, MO, MY, ML, MR, MD, MN, MA, MM, NP, NZ, NZ, NI, OM, PK, PA, PG, PY, PE, QA, RO, SA, RS, SI, TN, TR, UY, AE, VE, VN, and ZW.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from BioFire Diagnostics
All 44| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Biofire Respiratory Panel 2.1 (RP2.1) recalled by BioFire Diagnostics | BioFire Diagnostics | Other | Nationwide |
| Device | CLASS II | Biofire Spotfire Respiratory/Sore Throat Panel recalled | BioFire Diagnostics | Device Malfunction | Nationwide |
| Device | CLASS II | BioFire Joint Infection (JI) Panel recalled by BioFire Diagnostics | BioFire Diagnostics | Other | Nationwide |
| Device | CLASS II | BioFire Respiratory Panel 2.1, IVD recalled by BioFire Diagnostics | BioFire Diagnostics | Device Malfunction | 10 states |
| Device | CLASS II | BioFire Respiratory Panel 2.1 plus Kit recalled by BioFire Diagnostics | BioFire Diagnostics | Device Malfunction | UT |
Other Test Kits & Diagnostics recalls
All 2,661| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | McKesson Consult hCG Urine Cassette recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |
| Device | CLASS II | Systems/units containing affected recalled | Beckman Coulter Mishima K.K | Device Malfunction | Nationwide |
| Device | CLASS II | Fisher Scientific Sure-Vue Urine hCG recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |
| Device | CLASS II | Vitek 2 Ast-N461 Test Kit, Gram Negative Susceptibility recalled | Biomerieux | Other | MO |
| Device | CLASS II | DASH Rapid PCR Instrument recalled by Nuclein | Nuclein | Device Malfunction | Nationwide |