CLASS II ONGOING Device recall · Reported 14 Aug 2024 · FDA Enforcement Report

Biofire Filmarray Gastrointestinal (GI) Panel recalled

BioFire Diagnostics is recalling Biofire Filmarray Gastrointestinal (GI) Panel, distributed nationwide in the US. The recall was initiated on 16 May 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesPart Number: RFIT-ASY-0116 (30 test kit) UDI-DI code: 00815381020109 0081538102011: All unexpired products are potentially impacted if used in conjunction with the Remel Cary Blair transport medium Part Number: RFIT-ASY-0104 (6 test kit) UDI-DI code: 0081538102011 All unexpired products are potentially impacted if used in conjunction with the Remel Cary Blair transport medium Remel Cary-Blair Transport Medium lot numbers: 743930 769240 712258 782425 732357 742049
Quantity recalled67,613 kits (or 2,028,309 pouches). Each pouch is used for each individual test.

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2514-2024
FDA event ID
94992
Recalling firm
BioFire Diagnostics, Salt Lake City, UT
Classification
Class II
Status
Ongoing
Recall initiated
16 May 2024
FDA classified
6 Aug 2024
Reported
14 Aug 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter

Timeline

16 May 2024Recall initiated by company
6 Aug 2024Classified by FDA+82 days
14 Aug 2024Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Due to increase of false positive Vibrio/Vibrio Cholerae results when using a certain transport medium.

Product as listed by the FDA

BIOFIRE FILMARRAY Gastrointestinal (GI) Panel REF RFIT-ASY-0116 (30 test kit) BIOFIRE FILMARRAY Gastrointestinal (GI) Panel REF RFIT-ASY-0104 (6 kit test) The BIOFIRE FILMARRAY Gastrointestinal (GI) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with BIOFIRE¿ FILMARRAY¿ Systems. The BIOFIRE GI Panel is capable of the simultaneous detection and identification of nucleic acids from multiple bacteria, viruses, and parasites directly from stool samples in Cary Blair transport media obtained from individuals with signs and/or symptoms of gastrointestinal infection.

Where it was distributed

U.S.: Nationwide - AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, Virgin Island, VT, WA, WI, WV, and WY. O.U.S.: Worldwide - AR, AU, AT, BE, BR, CA, CL, HK, CN, CO, CZ, DE, FI, FR, GR, HU, IN, IT, JP, MX, NL, NO, PH, PL, PT, SG, ZA, KR, ES, SE, CH, TW, TH, TR, GB, AL, DZ, AO, AM, BH, BY, BO, BA, BW, BG, CR, HR, CY, CD, DO, AN, EC, SV, EE, FJ, GE, GU, GT, GN, HT, HN, IS, ID, IQ, IE, IL, CI, CI, JO, KE, KW, LV, LB, LT, MO, MY, ML, MR, MD, MN, MA, MM, NP, NZ, NZ, NI, OM, PK, PA, PG, PY, PE, QA, RO, SA, RS, SI, TN, TR, UY, AE, VE, VN, and ZW.

Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusetts

Questions about this recall

Why was it recalled?

Due to increase of false positive Vibrio/Vibrio Cholerae results when using a certain transport medium.

Which lots are affected?

Part Number: RFIT-ASY-0116 (30 test kit) UDI-DI code: 00815381020109 0081538102011: All unexpired products are potentially impacted if used in conjunction with the Remel Cary Blair transport medium Part Number: RFIT-ASY-0104 (6 test kit) UDI-DI code: 0081538102011 All unexpired products are potentially impacted if used in conjunction with the Remel Cary Blair transport medium Remel Cary-Blair Transport Medium lot numbers: 743930 769240 712258 782425 732357 742049

Where was it sold?

U.S.: Nationwide - AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, Virgin Island, VT, WA, WI, WV, and WY. O.U.S.: Worldwide - AR, AU, AT, BE, BR, CA, CL, HK, CN, CO, CZ, DE, FI, FR, GR, HU, IN, IT, JP, MX, NL, NO, PH, PL, PT, SG, ZA, KR, ES, SE, CH, TW, TH, TR, GB, AL, DZ, AO, AM, BH, BY, BO, BA, BW, BG, CR, HR, CY, CD, DO, AN, EC, SV, EE, FJ, GE, GU, GT, GN, HT, HN, IS, ID, IQ, IE, IL, CI, CI, JO, KE, KW, LV, LB, LT, MO, MY, ML, MR, MD, MN, MA, MM, NP, NZ, NZ, NI, OM, PK, PA, PG, PY, PE, QA, RO, SA, RS, SI, TN, TR, UY, AE, VE, VN, and ZW.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from BioFire Diagnostics

All 44

Other Test Kits & Diagnostics recalls

All 2,661