CLASS II ONGOING Device recall · Reported 24 Mar 2021 · FDA Enforcement Report

Canon Interventional Angiography System recalled

Canon Medical System USA is recalling Canon Interventional Angiography System, distributed nationwide in the US. The recall was initiated on 8 Dec 2020 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesTable CAT-880B Models: INFX-8000F/AH INFX-8000F/AH INFX-8000F/AH INFX-8000F/AH INFX-8000F/AH INFX-8000F/BHUP INFX-8000F/AH INFX-8000F/AH INFX-8000F/UU INFX-8000F/AH INFX-8000F/AH INFX-8000F/AH INFX-8000F/AH INFX-8000F/AH INFX-8000F/AH INFX-8000F/BH Table Serial Numbers: 99E1582258 99F18Y2428 99C1482200 99E15Y2271 99F1862406 99G19Z2438 99E1662308 99E1762365 99A1337005 99F1852401 99E1612286 99E1712347 99E1712345 99C14Z2224 99E1592262 99G1972436 System Serial Numbers: AHA1582003 AHC18Y2013 AHA1482001 AHA15Y2005 AHC1862012 BHB19Z2002 AHB1662007 AHB1762010 UUB12X2034 AHC1852011 AHA1612006 AHB1712009 AHB1712008 AHA14Z2002 AHA1592004 BHB1972001
Quantity recalledTotal of all Systems = 145 systems (U.S.)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1271-2021
FDA event ID
87322
Recalling firm
Canon Medical System USA, Tustin, CA
Classification
Class II
Status
Ongoing
Recall initiated
8 Dec 2020
FDA classified
17 Mar 2021
Reported
24 Mar 2021
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

8 Dec 2020Recall initiated by company
17 Mar 2021Classified by FDA+99 days
24 Mar 2021Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

System table may tilt because the bolts that fasten the vertical column and the table top tilting drive unit may become loose.

Product as listed by the FDA

Canon INTERVENTIONAL ANGIOGRAPHY SYSTEM MODEL INFX-8000F Catheterization Table CAT-880B Canon CATHETERIZATION TABLE, MODEL CAT-880B INPUT 220V - Product Usage: indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen, and lower extremities.

Where it was distributed

Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NJ, NY, OH, OK, PA, SD, TN, TX, WA, WI and WV. The countries of Canada, China, Japan, Korea, and Singapore.

Nationwide AlabamaArkansasArizonaCaliforniaColoradoConnecticutDelawareFloridaGeorgiaIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMaineMichigan

Questions about this recall

Why was it recalled?

System table may tilt because the bolts that fasten the vertical column and the table top tilting drive unit may become loose.

Which lots are affected?

Table CAT-880B Models: INFX-8000F/AH INFX-8000F/AH INFX-8000F/AH INFX-8000F/AH INFX-8000F/AH INFX-8000F/BHUP INFX-8000F/AH INFX-8000F/AH INFX-8000F/UU INFX-8000F/AH INFX-8000F/AH INFX-8000F/AH INFX-8000F/AH INFX-8000F/AH INFX-8000F/AH INFX-8000F/BH Table Serial Numbers: 99E1582258 99F18Y2428 99C1482200 99E15Y2271 99F1862406 99G19Z2438 99E1662308 99E1762365 99A1337005 99F1852401 99E1612286 99E1712347 99E1712345 99C14Z2224 99E1592262 99G1972436 System Serial Numbers: AHA1582003 AHC18Y2013 AHA1482001 AHA15Y2005 AHC1862012 BHB19Z2002 AHB1662007 AHB1762010 UUB12X2034 AHC1852011 AHA1612006 AHB1712009 AHB1712008 AHA14Z2002 AHA1592004 BHB1972001

Where was it sold?

Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NJ, NY, OH, OK, PA, SD, TN, TX, WA, WI and WV. The countries of Canada, China, Japan, Korea, and Singapore.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

2 other products · FDA event 87322
ReportedClassRecallCompanyReasonWhere
CLASS IICanon Interventional Angiography System recalledCanon Medical System USAOtherNationwide
CLASS IICanon Interventional Angiography System recalledCanon Medical System USAOtherNationwide

More recalls from Canon Medical System USA

All 51

Other Heart & Vascular Devices recalls

All 6,220