CLASS II ONGOING Device recall · Reported 25 Feb 2026 · FDA Enforcement Report

Capillary Tubes recalled by Medica

Medica is recalling Capillary Tubes. The recall was initiated on 31 Dec 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel No. 7303, UDI-DI: 00840095607575, Lot No. 25XXX.
Quantity recalled1,578 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1370-2026
FDA event ID
98315
Recalling firm
Medica, Bedford, MA
Classification
Class II
Status
Ongoing
Recall initiated
31 Dec 2025
FDA classified
13 Feb 2026
Reported
25 Feb 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

31 Dec 2025Recall initiated by company
13 Feb 2026Classified by FDA+44 days
25 Feb 2026Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

A systematic positive bias of up to 20% was identified across the reportable potassium range, resulting in false elevation results of patient samples when using Medica s capillary tubes.

Product as listed by the FDA

Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit.

Where it was distributed

International distribution to the countries of Russia, Bulgaria, Kazakhstan, Ukraine, Pakistan, Belarus.

Questions about this recall

Why was it recalled?

A systematic positive bias of up to 20% was identified across the reportable potassium range, resulting in false elevation results of patient samples when using Medica s capillary tubes.

Which lots are affected?

Model No. 7303, UDI-DI: 00840095607575, Lot No. 25XXX.

Where was it sold?

International distribution to the countries of Russia, Bulgaria, Kazakhstan, Ukraine, Pakistan, Belarus.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.