CLASS II ONGOING Device recall · Reported 24 Jan 2024 · FDA Enforcement Report

Cardinal Health Medi-vac CRD Suction Canister Liner Suction recalled

Cardinal Health 200 is recalling Cardinal Health Medi-vac CRD Suction Canister Liner Suction Canister Liner Kit with 8 m. The recall was initiated on 21 Dec 2023 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codes50630140152633 (case), Lot Numbers: J308-606, J308-624, J308-628 J308-642, J308-643, J309-608, J309-610, J309-622, J309-626, J309-633, J309-645, J310-602, J310-617, J310-645
Quantity recalled203244 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0755-2024
FDA event ID
93628
Recalling firm
Cardinal Health 200, Waukegan, IL
Classification
Class II
Status
Ongoing
Recall initiated
21 Dec 2023
FDA classified
18 Jan 2024
Reported
24 Jan 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

21 Dec 2023Recall initiated by company
18 Jan 2024Classified by FDA+28 days
24 Jan 2024Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The above-listed lots of Cardinal Health Medi-Vac CRD" Suction Canister Liners 1L may experience loss of suction on low or low intermittent suction settings, due to a misalignment of the liner with the outer hard canister.

Product as listed by the FDA

Cardinal Health MEDI-VAC CRD Suction Canister Liner, 1000 mL Suction Canister Liner Kit with 5 mm 1.8 m Tube, REF 65651-516

Where it was distributed

US, Canada, Australia and New Zealand

Questions about this recall

Is Cardinal Health Medi-vac CRD Suction Canister Liner Suction Canister Liner Kit with 8 m recalled?

Yes. Cardinal Health 200 recalled Cardinal Health Medi-vac CRD Suction Canister Liner Suction Canister Liner Kit with 8 m on 21 Dec 2023. The recall is Class II and its status is Ongoing. Recall number Z-0755-2024.

Why was it recalled?

The above-listed lots of Cardinal Health Medi-Vac CRD" Suction Canister Liners 1L may experience loss of suction on low or low intermittent suction settings, due to a misalignment of the liner with the outer hard canister.

Which lots are affected?

50630140152633 (case), Lot Numbers: J308-606, J308-624, J308-628 J308-642, J308-643, J309-608, J309-610, J309-622, J309-626, J309-633, J309-645, J310-602, J310-617, J310-645

Where was it sold?

US, Canada, Australia and New Zealand

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Cardinal Health 200

All 283

Other Other Medical Devices recalls

All 13,778