CLASS II ONGOING Device recall · Reported 15 Mar 2023 · FDA Enforcement Report

CardioMEMS HF System PA Sensor and Delivery System recalled

St. Jude Medical is recalling CardioMEMS HF System PA Sensor and Delivery System, distributed nationwide in the US. The recall was initiated on 7 Feb 2023 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI/DI 05415067034724: M220200004, M220200005, M220200006, M220200008, M220200011, M220200014, M220200016, M220200017, M220200018, M220200019, M220200021, M220200022, M220200026, M220200028, M220200029, M220200031, M220200032, M220200034, M220200035, M220200036, M220200037, M220200039, M220200040, M220200041, M220200042, M220300002, M220300003, M220300004, M220300005, M220300007, M220300008, M220300009, M220300010, M220300011, M220300012, M220300013, M220300014, M220300016, M220300017, M220300018, M220300019, M220300021, M220300022, M220300023, M220300024, M220300025, M220300026, M220300027, M220300028, M220300029, M220300030, M220300031, M220300032, M220300035, M220300036, M220300040, M220300041, M220300042, M220300043, M220300044, M220300045, M220300046, M220300047, M220300048, M220300049, M220300050, M220300051, M220300052, M220300053, M220300054, M220300055, M220300057, M220300058, M220300059, M220300060, M220300061, M220300062, M220300063, M220300064, M220300065, M220300066, M220300067, M220300069, M220300070, M220300071, M220300073, M220300074, M220300075, M220300077, M220300078, M220300080, M220300081, M220300082, M220300083, M220300084, M220300085, M220300087, M220300088, M220300089, M220300091, M220300092, M220300093, M220300096, M220300097, M220300101, M220300102, M220300103, M220300104, M220300105, M220300107, M220300108, M220300109, M220300111, M220300112, M220300114, M220300115, M220300116, M220300118, M220300119, M220300120, M220300121, M220300122, M220300123, M220300125, M220300126, M220300127, M220300129, M220300131, M220300136, M220300138, M220300139, M220300140, M220300141, M220300142, M220300143, M220300144, M220300145, M220300146, M220300147, M220300150, M220300151, M220300152, M220300153, M220300154, M220300155, M220300156, M220300157, M220300158, M220300160, M220300161, M220300162, M220300163, M220300164, M220300166, M220300167, M220300168, M220300169, M220300170, M220300171, M220300172, M220300173, M220300174, M220300175, M220300176, M220300177, M220300178, M220300179, M220300180, M220300181, M220300182, M220300183, M220300184, M220300186, M220300187, M220300188, M220300189, M220300190, M220300191, M220300192, M220300193, M220300194, M220300195, M220300196, M220700025, M220700026, M220700027, M220700028, M220700031, M220700033, M220700034, M220700035, M220700036, M220700037, M220700041, M220700042, M220700043, M220700045, M220700051, M220700052, M220700054, M220700059, M220700061, M220700097, M220700098, M220700102, M220700104, M220700106, M220700107, M220700109, M220700110, M220700111, M220700112, M220700113, M220700114, M220700115, M220700116, M220700117, M220700118, M220700119, M220700120, M220700121, M220700122, M220700123, M220700124, M220700125, M220700126, M220700128, M220700129, M220700130, M220700131, M220700132, M220700133, M220700134, M220700135, M220700136, M220700137, M220700138, M220700139, M220700142, M220700143, M220700144, M220700145, M220700146, M220700147, M220900017, M220900018, M220900021, M220900023, M220900024, M220900025, M220900026, M220900030, M220900031, M220900032, M220900033, M220900034, M220900035, M220900037, M220900038, M220900040, M220900042, M220900044, M220900045, M220900046, M220900047, M220900048, M220900049, M220900050, M220900052, M220900063, M220900065, M220900066, M220900067, M220900068, M220900069, M220900070, M220900071, M220900072, M220900073, M220900075, M220900076, M220900078, M220900079, M220900080, M220900081, M220900086, M220900087, M220900088, M220900089, M220900090, M220900091, M220900093, M220900094, M220900095, M220900096, M220900097, M220900101, M220900104, M220900106, M220900109, M221000023, M221000030, M221000032, M221000034, M221000035, M221000038, M221000039, M221000041, M221000042, M221000044, M221000047, M221000048, M221000049, M221000050, M221000052, M221100002, M221000046
Quantity recalled317 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1225-2023
FDA event ID
91606
Recalling firm
St. Jude Medical, Atlanta, GA
Classification
Class II
Status
Ongoing
Recall initiated
7 Feb 2023
FDA classified
6 Mar 2023
Reported
15 Mar 2023
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Labeling Errors
Device type
Patient Monitors

Timeline

7 Feb 2023Recall initiated by company
6 Mar 2023Classified by FDA+27 days
15 Mar 2023Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Select CardioMEMS PA Sensors (Model CM2000) operate outside of the intended frequency range at higher elevations when readings are taken over approximately 2,000 ft. (610 meters) above sea level; operation outside of the intended radiofrequency has the potential to result in inaccurate readings or sensor signal acquisition difficulties.

Product as listed by the FDA

CardioMEMS HF System PA Sensor and Delivery System, Model Number CM3100, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and management of heart failure (HF) patients. The system measures changes in PA pressure which physicians use to initiate or modify heart failure treatment.

Where it was distributed

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Canada, Colombia, Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Kuwait, Netherlands, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, UK, United Arab Emirates.

Nationwide

Questions about this recall

Why was it recalled?

Select CardioMEMS PA Sensors (Model CM2000) operate outside of the intended frequency range at higher elevations when readings are taken over approximately 2,000 ft. (610 meters) above sea level; operation outside of the intended radiofrequency has the potential to result in inaccurate readings or sensor signal acquisition difficulties.

Which lots are affected?

UDI/DI 05415067034724: M220200004, M220200005, M220200006, M220200008, M220200011, M220200014, M220200016, M220200017, M220200018, M220200019, M220200021, M220200022, M220200026, M220200028, M220200029, M220200031, M220200032, M220200034, M220200035, M220200036, M220200037, M220200039, M220200040, M220200041, M220200042, M220300002, M220300003, M220300004, M220300005, M220300007, M220300008, M220300009, M220300010, M220300011, M220300012, M220300013, M220300014, M220300016, M220300017, M220300018, M220300019, M220300021, M220300022, M220300023, M220300024, M220300025, M220300026, M220300027, M220300028, M220300029, M220300030, M220300031, M220300032, M220300035, M220300036, M220300040, M220300041, M220300042, M220300043, M220300044, M220300045, M220300046, M220300047, M220300048, M220300049, M220300050, M220300051, M220300052, M220300053, M220300054, M220300055, M220300057, M220300058, M220300059, M220300060, M220300061, M220300062, M220300063, M220300064, M220300065, M220300066, M220300067, M220300069, M220300070, M220300071, M220300073, M220300074, M220300075, M220300077, M220300078, M220300080, M220300081, M220300082, M220300083, M220300084, M220300085, M220300087, M220300088, M220300089, M220300091, M220300092, M220300093, M220300096, M220300097, M220300101, M220300102, M220300103, M220300104, M220300105, M220300107, M220300108, M220300109, M220300111, M220300112, M220300114, M220300115, M220300116, M220300118, M220300119, M220300120, M220300121, M220300122, M220300123, M220300125, M220300126, M220300127, M220300129, M220300131, M220300136, M220300138, M220300139, M220300140, M220300141, M220300142, M220300143, M220300144, M220300145, M220300146, M220300147, M220300150, M220300151, M220300152, M220300153, M220300154, M220300155, M220300156, M220300157, M220300158, M220300160, M220300161, M220300162, M220300163, M220300164, M220300166, M220300167, M220300168, M220300169, M220300170, M220300171, M220300172, M220300173, M220300174, M220300175, M220300176, M220300177, M220300178, M220300179, M220300180, M220300181, M220300182, M220300183, M220300184, M220300186, M220300187, M220300188, M220300189, M220300190, M220300191, M220300192, M220300193, M220300194, M220300195, M220300196, M220700025, M220700026, M220700027, M220700028, M220700031, M220700033, M220700034, M220700035, M220700036, M220700037, M220700041, M220700042, M220700043, M220700045, M220700051, M220700052, M220700054, M220700059, M220700061, M220700097, M220700098, M220700102, M220700104, M220700106, M220700107, M220700109, M220700110, M220700111, M220700112, M220700113, M220700114, M220700115, M220700116, M220700117, M220700118, M220700119, M220700120, M220700121, M220700122, M220700123, M220700124, M220700125, M220700126, M220700128, M220700129, M220700130, M220700131, M220700132, M220700133, M220700134, M220700135, M220700136, M220700137, M220700138, M220700139, M220700142, M220700143, M220700144, M220700145, M220700146, M220700147, M220900017, M220900018, M220900021, M220900023, M220900024, M220900025, M220900026, M220900030, M220900031, M220900032, M220900033, M220900034, M220900035, M220900037, M220900038, M220900040, M220900042, M220900044, M220900045, M220900046, M220900047, M220900048, M220900049, M220900050, M220900052, M220900063, M220900065, M220900066, M220900067, M220900068, M220900069, M220900070, M220900071, M220900072, M220900073, M220900075, M220900076, M220900078, M220900079, M220900080, M220900081, M220900086, M220900087, M220900088, M220900089, M220900090, M220900091, M220900093, M220900094, M220900095, M220900096, M220900097, M220900101, M220900104, M220900106, M220900109, M221000023, M221000030, M221000032, M221000034, M221000035, M221000038, M221000039, M221000041, M221000042, M221000044, M221000047, M221000048, M221000049, M221000050, M221000052, M221100002, M221000046

Where was it sold?

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Canada, Colombia, Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Kuwait, Netherlands, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, UK, United Arab Emirates.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 91606
ReportedClassRecallCompanyReasonWhere
CLASS IICardioMEMS HF System PA Sensor and Delivery System recalledSt. Jude MedicalLabeling ErrorsNationwide

More recalls from St. Jude Medical

All 47

Other Patient Monitors recalls

All 1,022