CLASS II ONGOING Device recall · Reported 30 Apr 2025 · FDA Enforcement Report

Cell-dyn Ruby recalled by Abbott Laboratories

Abbott Laboratories is recalling Cell-dyn Ruby, distributed nationwide in the US. The recall was initiated on 18 Mar 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI 00380740017170 SNs: 34006BG96 34013BG96 34022BG96 34026BG96 34045BG96 34048BG96 34050BG96 34052BG96 34053BG96 34061BG96 34068BG96 34103BG96 34109BG96 34134BG96 34154BG96 34171BG96 34174BG96 34182BG96 34188BG96 34194BG96 34229BG96 34250BG96 34254BG96 34264BG96 34285BG96 34289BG96 34317BG96 34334BG96 34338BG96 34343BG96 34348BG96 34359BG96 34383BG96 34392BG96 34422BG96 34425BG96 34426BG96 34430BG96 34431BG96 34434BG96 34445BG96 34457BG96 34479BG96 34486BG96 34503BG96 34507BG96 34532BG96 34549BG96 34556BG96 34584BG96 34591BG96 34592BG96 34595BG96 34602BG96 34603BG96 34647BG96 34650GB96 34651BG96 34655BG96 34671BG96 34683BG96 34694BG96 34698BG96 34723BG96 34733BG96 34761BG96 34774BG96 34778BG96 34814BG96 34840BG96 34891BG96 34912BG96 34917BG96 34936BG96 34962BG96 34985BG96 34986BG96 35011BG96 35035BG96 35037BG96 35040BG96 35083BG96 35084BG96 35109BG96 35129BG96 35136BG96 35162BG96 35197BG96 35210BG96 35234BG96 35238BG96 35243BG96 35266BG96 35269BG96 35289BG96 35303BG96 35458BG96 35460BG96 35464BG96 35472BG96 35519BG96 35732BG96 35743BG96 35774BG96 35849BG96 35861BG96 36158BG96 36305BG96 36338BG96 36350BG96 36351BG96 36356BG96 36599BG96 36646BG96 36659BG96 36690BG96 36763BG96 36766BG96 36820BG96 36844BG96 54017BG96 54056BG96 54083BG96 54106BG96 54118BG96 54139BG96 54145BG96 54147BG96 54148BG96 54150BG96 54158BG96 54213BG96 54254BG96 54363BG96 54400BG96 54602BG96 54932BG96 54981BG96 70474BG96 70510BG96 71057BG96 71210BG96 71481BG96 71644BG96 71794BG96 71909BG96
Quantity recalled146

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1653-2025
FDA event ID
96567
Recalling firm
Abbott Laboratories, Abbott Park, IL
Classification
Class II
Status
Ongoing
Recall initiated
18 Mar 2025
FDA classified
22 Apr 2025
Reported
30 Apr 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction
Device type
Medical Software

Timeline

18 Mar 2025Recall initiated by company
22 Apr 2025Classified by FDA+35 days
30 Apr 2025Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-03 Software Version: All software versions Product Description: CELL-DYN Ruby Component: No

Where it was distributed

Worldwide distribution: US (nationwide) to the following states: AR, AZ, CA, CO, FL, GA, IA, IL, KS, LA, MA, MD, MI, MN, MO, MS, ND, NE, NJ, NV, NY, OH, OK, PA, PR, SC, TN, TX, UT, VA, WA, WI, WV, WY and OUS (foreign) to countries of: Afghanistan, Albania, Algeria, Angola, Anguilla, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Croatia, Czech Rebublic, Ecuador, Egypt, El Salvador, Ethiopia, France, Gambia, Gaza & Jericho, Germany, Greece, Guatemala, Guana, Haiti, Honduras, Hong King, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Kyrgyzstan, Lebanon, Lithuania, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Moldova, Montenegro, Morocco, Myanmar, New Caledonia, Nicaragua, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, SRI Lanka, St. Lucia, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad & Tobago, Tunisia, Turk & Caicos, Turkey, UAE, Uganda, UK, Ukraine, Uruguay, Uzbekistan, Vietnam, and Yemen.

Nationwide ArkansasArizonaCaliforniaColoradoFloridaGeorgiaIowaIllinoisKansasLouisianaMassachusettsMarylandMichiganMinnesotaMissouriMississippiNorth DakotaNebraskaNew JerseyNevada

Questions about this recall

Why was it recalled?

When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.

Which lots are affected?

UDI-DI 00380740017170 SNs: 34006BG96 34013BG96 34022BG96 34026BG96 34045BG96 34048BG96 34050BG96 34052BG96 34053BG96 34061BG96 34068BG96 34103BG96 34109BG96 34134BG96 34154BG96 34171BG96 34174BG96 34182BG96 34188BG96 34194BG96 34229BG96 34250BG96 34254BG96 34264BG96 34285BG96 34289BG96 34317BG96 34334BG96 34338BG96 34343BG96 34348BG96 34359BG96 34383BG96 34392BG96 34422BG96 34425BG96 34426BG96 34430BG96 34431BG96 34434BG96 34445BG96 34457BG96 34479BG96 34486BG96 34503BG96 34507BG96 34532BG96 34549BG96 34556BG96 34584BG96 34591BG96 34592BG96 34595BG96 34602BG96 34603BG96 34647BG96 34650GB96 34651BG96 34655BG96 34671BG96 34683BG96 34694BG96 34698BG96 34723BG96 34733BG96 34761BG96 34774BG96 34778BG96 34814BG96 34840BG96 34891BG96 34912BG96 34917BG96 34936BG96 34962BG96 34985BG96 34986BG96 35011BG96 35035BG96 35037BG96 35040BG96 35083BG96 35084BG96 35109BG96 35129BG96 35136BG96 35162BG96 35197BG96 35210BG96 35234BG96 35238BG96 35243BG96 35266BG96 35269BG96 35289BG96 35303BG96 35458BG96 35460BG96 35464BG96 35472BG96 35519BG96 35732BG96 35743BG96 35774BG96 35849BG96 35861BG96 36158BG96 36305BG96 36338BG96 36350BG96 36351BG96 36356BG96 36599BG96 36646BG96 36659BG96 36690BG96 36763BG96 36766BG96 36820BG96 36844BG96 54017BG96 54056BG96 54083BG96 54106BG96 54118BG96 54139BG96 54145BG96 54147BG96 54148BG96 54150BG96 54158BG96 54213BG96 54254BG96 54363BG96 54400BG96 54602BG96 54932BG96 54981BG96 70474BG96 70510BG96 71057BG96 71210BG96 71481BG96 71644BG96 71794BG96 71909BG96

Where was it sold?

Worldwide distribution: US (nationwide) to the following states: AR, AZ, CA, CO, FL, GA, IA, IL, KS, LA, MA, MD, MI, MN, MO, MS, ND, NE, NJ, NV, NY, OH, OK, PA, PR, SC, TN, TX, UT, VA, WA, WI, WV, WY and OUS (foreign) to countries of: Afghanistan, Albania, Algeria, Angola, Anguilla, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Croatia, Czech Rebublic, Ecuador, Egypt, El Salvador, Ethiopia, France, Gambia, Gaza & Jericho, Germany, Greece, Guatemala, Guana, Haiti, Honduras, Hong King, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Kyrgyzstan, Lebanon, Lithuania, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Moldova, Montenegro, Morocco, Myanmar, New Caledonia, Nicaragua, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, SRI Lanka, St. Lucia, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad & Tobago, Tunisia, Turk & Caicos, Turkey, UAE, Uganda, UK, Ukraine, Uruguay, Uzbekistan, Vietnam, and Yemen.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

4 other products · FDA event 96567

This product is part of a recall event; the main page for the event is Cell-dyn Ruby recalled by Abbott Laboratories.

More recalls from Abbott Laboratories

All 153

Other Medical Software recalls

All 908