CLASS II ONGOING Device recall · Reported 4 Nov 2026 · FDA Enforcement Report

Nalu TD2 Therapy Disc with firmware recalled by Boston Scientific

Boston Scientific Neuromodulation is recalling Nalu TD2 Therapy Disc with firmware, distributed nationwide in the US. The recall was initiated on 30 Sep 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesFirmware versions v2.0.71 and v2.0.77 REF/UDI-DI/Serial Numbers: 34014-001/00812537037134/ 100855, 100877, 100883, 100884, 100886, 100888, 100898, 100903, 100908, 100911, 100916, 100932, 101103, 100902; 74011/00812537037127, 00812537036977/101439, 100913, 100949, 101159, 101162, 101163, 101164, 101173, 101175, 101177, 101186, 101187, 101188, 101189, 101192, 101200, 101205, 101208, 101213, 101214, 101218, 101219, 101221, 101222, 101240, 101245, 101247, 101249, 101256, 101257, 101259, 101260, 101261, 101264, 101267, 101268, 101270, 101272, 101273, 101277, 101278, 101280, 101283, 101284, 101290, 101297, 101298, 101301, 101305, 101307, 101310, 101311, 101312, 101314, 101315, 101326, 101330, 101331, 101334, 101338, 101344, 101345, 101346, 101356, 101357, 101358, 101359, 101363, 101364, 101367, 101375, 101376, 101377, 101378, 101381, 101383, 101387, 101388, 101392, 101400, 101405, 101408, 101410, 101414, 101415, 101417, 101418, 101419, 101429, 101431, 101432, 101436, 101440, 101441, 101443, 101447, 101448, 101450, 101451, 101466, 101490, 101493, 101504, 101510, 101511, 101514, 101520, 101527, 101528, 101534, 101535, 101553, 101557, 101567, 101571, 101576, 101599, 101617, 101623, 101656
Quantity recalled134

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0311-2027
FDA event ID
99785
Recalling firm
Boston Scientific Neuromodulation, Valencia, CA
Classification
Class II
Status
Ongoing
Recall initiated
30 Sep 2026
FDA classified
29 Oct 2026
Reported
4 Nov 2026
Type
Voluntary: Firm initiated
Reason category
Other Reasons
Device type
Medical Software

Timeline

30 Sep 2026Recall initiated by company
29 Oct 2026Classified by FDA+29 days
4 Nov 2026Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Therapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a firmware issue that causes the TD to attempt to restart, then fail, leading it to become non-functional, and if a secondary TD is unavailable, this may lead IPG to become non-functional, thus unable to inhibit transmission of pain signals.

Product as listed by the FDA

Nalu TD2 Therapy Disc with firmware, REF: 34014-001, and TD2 Patient Kit, REF: 74011, part of the Nalu Neurostimulation System

Where it was distributed

US Nationwide distribution in the states of PA, FL, OR, AZ, OH, SC, TX, OR, OK, MI, CA, NJ.

Nationwide ArizonaCaliforniaFloridaMichiganNew JerseyOhioOklahomaOregonPennsylvaniaSouth CarolinaTexas

Questions about this recall

Why was it recalled?

Therapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a firmware issue that causes the TD to attempt to restart, then fail, leading it to become non-functional, and if a secondary TD is unavailable, this may lead IPG to become non-functional, thus unable to inhibit transmission of pain signals.

Which lots are affected?

Firmware versions v2.0.71 and v2.0.77 REF/UDI-DI/Serial Numbers: 34014-001/00812537037134/ 100855, 100877, 100883, 100884, 100886, 100888, 100898, 100903, 100908, 100911, 100916, 100932, 101103, 100902; 74011/00812537037127, 00812537036977/101439, 100913, 100949, 101159, 101162, 101163, 101164, 101173, 101175, 101177, 101186, 101187, 101188, 101189, 101192, 101200, 101205, 101208, 101213, 101214, 101218, 101219, 101221, 101222, 101240, 101245, 101247, 101249, 101256, 101257, 101259, 101260, 101261, 101264, 101267, 101268, 101270, 101272, 101273, 101277, 101278, 101280, 101283, 101284, 101290, 101297, 101298, 101301, 101305, 101307, 101310, 101311, 101312, 101314, 101315, 101326, 101330, 101331, 101334, 101338, 101344, 101345, 101346, 101356, 101357, 101358, 101359, 101363, 101364, 101367, 101375, 101376, 101377, 101378, 101381, 101383, 101387, 101388, 101392, 101400, 101405, 101408, 101410, 101414, 101415, 101417, 101418, 101419, 101429, 101431, 101432, 101436, 101440, 101441, 101443, 101447, 101448, 101450, 101451, 101466, 101490, 101493, 101504, 101510, 101511, 101514, 101520, 101527, 101528, 101534, 101535, 101553, 101557, 101567, 101571, 101576, 101599, 101617, 101623, 101656

Where was it sold?

US Nationwide distribution in the states of PA, FL, OR, AZ, OH, SC, TX, OR, OK, MI, CA, NJ.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 10 Nov 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Boston Scientific Neuromodulation

All 25

Other Medical Software recalls

All 936