CLASS II ONGOING Device recall · Reported 15 Jun 2022 · FDA Enforcement Report

Centricity PACS RA1000 recalled by GE Healthcare

GE Healthcare is recalling Centricity PACS RA1000, distributed nationwide in the US. The recall was initiated on 28 Apr 2022 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesa. Software Version 3.0 through 3.2 SP8, No GTIN/DI (Code: 2051372-0XX, 2064678-0XX) b. Software Version 4.0 through 4.0 SP14, GTIN/DI 00840682124447 (Code: 2062264-0XX); c. 6.0 through 6.0 SP10.3, GTIN/DI: 00840682104821 (Code: 2094628-0XX); d. 7.0 through 7.0 SP0.0.4.7, GTIN/DI: 00840682145558 (Code: 5824366-0XX)
Quantity recalled676 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1233-2022
FDA event ID
90136
Recalling firm
GE Healthcare, Waukesha, WI
Classification
Class II
Status
Ongoing
Recall initiated
28 Apr 2022
FDA classified
3 Jun 2022
Reported
15 Jun 2022
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Labeling Errors

Timeline

28 Apr 2022Recall initiated by company
3 Jun 2022Classified by FDA+36 days
15 Jun 2022Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Potential for Distance and Area measurements to display inaccurate measurement values when performed on magnified images, and specific to ZFP, Distance and Area measurements can display inaccurate measurement values when performed on lossy images that are scaled down from their original resolution.

Product as listed by the FDA

Centricity PACS RA1000

Where it was distributed

Worldwide distribution. Nationwide distribution to AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. International distribution to Argentina, Australia, Austria, Bahamas, Belgium, Bermuda, Bolivia, Plurinational State of, Brazil, Canada, Chile, China, Colombia, Costa Rica, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Kazakhstan, Korea, South, Kuwait, Lebanon, Malaysia, Malta, Mexico, Namibia, Netherlands, Nigeria, Norway, Pakistan, Panama, Peru, Poland, Portugal, Puerto Rico, Qatar, Russia, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Bolivarian Republic of, Vietnam

Nationwide AlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMaryland

Questions about this recall

Why was it recalled?

Potential for Distance and Area measurements to display inaccurate measurement values when performed on magnified images, and specific to ZFP, Distance and Area measurements can display inaccurate measurement values when performed on lossy images that are scaled down from their original resolution.

Which lots are affected?

a. Software Version 3.0 through 3.2 SP8, No GTIN/DI (Code: 2051372-0XX, 2064678-0XX) b. Software Version 4.0 through 4.0 SP14, GTIN/DI 00840682124447 (Code: 2062264-0XX); c. 6.0 through 6.0 SP10.3, GTIN/DI: 00840682104821 (Code: 2094628-0XX); d. 7.0 through 7.0 SP0.0.4.7, GTIN/DI: 00840682145558 (Code: 5824366-0XX)

Where was it sold?

Worldwide distribution. Nationwide distribution to AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. International distribution to Argentina, Australia, Austria, Bahamas, Belgium, Bermuda, Bolivia, Plurinational State of, Brazil, Canada, Chile, China, Colombia, Costa Rica, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Kazakhstan, Korea, South, Kuwait, Lebanon, Malaysia, Malta, Mexico, Namibia, Netherlands, Nigeria, Norway, Pakistan, Panama, Peru, Poland, Portugal, Puerto Rico, Qatar, Russia, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Bolivarian Republic of, Vietnam

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

4 other products · FDA event 90136

This product is part of a recall event; the main page for the event is Centricity Radiology RA600 recalled by GE Healthcare.

More recalls from GE Healthcare

All 512

Other Other Medical Devices recalls