CLASS II ONGOING Device recall · Reported 23 Apr 2025 · FDA Enforcement Report

Cerene Cryotherapy Device, Thermal Ablation, Endometrial recalled

Channel Medsystems is recalling Cerene Cryotherapy Device, Thermal Ablation, Endometrial, distributed nationwide in the US. The recall was initiated on 20 Mar 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLot Code: Model No: FGS-7000 UDI-DI: 00850008595035 Lot Numbers: 108638953, 108645585, 108645586, 108645587, 108866674, 108866675, 109493291, 109493292 and 109895874
Quantity recalled281

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1584-2025
FDA event ID
96485
Recalling firm
Channel Medsystems, Berkeley, CA
Classification
Class II
Status
Ongoing
Recall initiated
20 Mar 2025
FDA classified
16 Apr 2025
Reported
23 Apr 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Packaging Defects

Timeline

20 Mar 2025Recall initiated by company
16 Apr 2025Classified by FDA+27 days
23 Apr 2025Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Due to evaporator gasket not sealing which can cause liquid nitrous oxide to drip/leak from the bottom of the device handle during treatment or after treatment during venting of the device

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

Brand Name: Cerene Cryotherapy Device Product Name: DEVICE, THERMAL ABLATION, ENDOMETRIAL Model/Catalog Number: FGS-7000 Product Description: The Cerene Cryotherapy Device is an endometrial ablation device that uses nitrous oxide (N2O) to freeze and ablate the endometrium to reduce future menstrual bleeding. The device is intended for use by healthcare professionals who have received appropriate training and are familiar with the principles, clinical applications, complications, side effects, and hazards commonly associated with endometrial ablation. The Cerene Device is single use, disposable, and provided sterile. The average procedure time is 7 minutes, with 2.5 minutes of active cryoablation. Ablation is achieved throughout the uterine cavity through the use of cryothermic energy. The cryothermic energy is provided by a liquid-to-gas phase change of N2O. During the 2.5-minute treatment cycle, liquid N2O (originating from a small Cylinder located in the device handle) flows through a delivery line and into an inflow line with multiple jets. This liquid N2O is infused into an ultra-thin polyurethane Liner, where it converts into gas. The gaseous N2O is exhausted through the Exhaust Hose exiting the bottom of the handle. Component: N/A

Where it was distributed

U.S Nationwide distribution in the states of AL, AZ, CA, CO, CT, DE, FL, GA, IL, IN, KY, LA, MD, MO, NC, NY, OH, PA, PR, RI, TN, TX.

Nationwide AlabamaArizonaCaliforniaColoradoConnecticutDelawareFloridaGeorgiaIllinoisIndianaKentuckyLouisianaMarylandMissouriNorth CarolinaNew YorkOhioPennsylvaniaPuerto RicoRhode Island

Questions about this recall

Why was it recalled?

Due to evaporator gasket not sealing which can cause liquid nitrous oxide to drip/leak from the bottom of the device handle during treatment or after treatment during venting of the device

Which lots are affected?

Lot Code: Model No: FGS-7000 UDI-DI: 00850008595035 Lot Numbers: 108638953, 108645585, 108645586, 108645587, 108866674, 108866675, 109493291, 109493292 and 109895874

Where was it sold?

U.S Nationwide distribution in the states of AL, AZ, CA, CO, CT, DE, FL, GA, IL, IN, KY, LA, MD, MO, NC, NY, OH, PA, PR, RI, TN, TX.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Channel Medsystems

All 2
ReportedClassRecallCompanyReasonWhere
DeviceCLASS IICerene Cryotherapy Device, Thermal Ablation, Endometrial recalledChannel MedsystemsOtherNationwide

Other Other Medical Devices recalls

All 13,778