CLASS I ONGOING Device recall · Reported 20 Mar 2024 · FDA Enforcement Report

Cerenovus Cerebase DA Guide Sheath, Part Numbers: a) GS9080SD recalled

Medos International Sarl is recalling Cerenovus Cerebase DA Guide Sheath, Part Numbers: a) GS9080SD. The recall was initiated on 2 Feb 2024 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesa) GS9080SD: UPDATED 4/4/2024 UDI-DI 10886704082316, Lot numbers: 31121042, 31133636, 31140739, 31140751, 31146719, 31154809, 31160736, 31174927, 31178955, 31189206, 31194710, 31194733, 31208988, 31212653, 31212661; b) GS9090SD: UPDATED 4/4/2024 UDI-DI 10886704082293, Lot numbers: 31094249, 31103843, 31108260, 31116484, 31116485, 31121039, 31121040, 31121043, 31121044, 31133637, 31133638, 31140740, 31140952, 31146720, 31150968, 31150969, 31154810, 31154811, 31160737, 31160738, 31168610, 31168611, 31174928, 31174930, 31178956, 31178957, 31178958, 31189207, 31189208, 31194711, 31194712, 31194734, 31194735, 31204436, 31208989, 31208990, 31208991, 31212654, 31212655, 31212663, 31225738; c) GS9095SD: UDI-DI 10886704082323, Lot Numbers: 31208992, 31212656
Quantity recalled1343 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1284-2024
FDA event ID
93955
Recalling firm
Medos International Sarl, Le Locle, N/A
Classification
Class I
Status
Ongoing
Recall initiated
2 Feb 2024
FDA classified
13 Mar 2024
Reported
20 Mar 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

2 Feb 2024Recall initiated by company
13 Mar 2024Classified by FDA+40 days
20 Mar 2024Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Medos has received an increase in complaints for CEREBASE DA Guide Sheath with reports of fractures at the distal end, which may result in surgical procedural delay, vascular injury and in extreme rare occasions it may result in embolism.

Product as listed by the FDA

CERENOVUS CEREBASE DA Guide Sheath, Part Numbers: a) GS9080SD; b) GS9090SD; c) GS9095SD; Vascular guide-catheter, single-use

Where it was distributed

Worldwide distribution.

Questions about this recall

Why was it recalled?

Medos has received an increase in complaints for CEREBASE DA Guide Sheath with reports of fractures at the distal end, which may result in surgical procedural delay, vascular injury and in extreme rare occasions it may result in embolism.

Which lots are affected?

a) GS9080SD: UPDATED 4/4/2024 UDI-DI 10886704082316, Lot numbers: 31121042, 31133636, 31140739, 31140751, 31146719, 31154809, 31160736, 31174927, 31178955, 31189206, 31194710, 31194733, 31208988, 31212653, 31212661; b) GS9090SD: UPDATED 4/4/2024 UDI-DI 10886704082293, Lot numbers: 31094249, 31103843, 31108260, 31116484, 31116485, 31121039, 31121040, 31121043, 31121044, 31133637, 31133638, 31140740, 31140952, 31146720, 31150968, 31150969, 31154810, 31154811, 31160737, 31160738, 31168610, 31168611, 31174928, 31174930, 31178956, 31178957, 31178958, 31189207, 31189208, 31194711, 31194712, 31194734, 31194735, 31204436, 31208989, 31208990, 31208991, 31212654, 31212655, 31212663, 31225738; c) GS9095SD: UDI-DI 10886704082323, Lot Numbers: 31208992, 31212656

Where was it sold?

Worldwide distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Medos International Sarl

All 2
ReportedClassRecallCompanyReasonWhere
DeviceCLASS IIVelys Enabling Tech Robotic-Assist Station recalledMedos International SarlDevice MalfunctionNationwide

Other Heart & Vascular Devices recalls

All 6,220