CLASS II Device recall · Reported 21 Apr 2021 · FDA Enforcement Report
Chesapeake Cervical-Ti Drill Guide Inserter- a multi-screw recalled
K2M is recalling Chesapeake Cervical-Ti Drill Guide Inserter- a multi-screw construct providing stability, distributed in Pennsylvania. The recall was initiated on 16 Mar 2021 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1410-2021
- FDA event ID
- 87565
- Recalling firm
- K2M, Leesburg, VA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 16 Mar 2021
- FDA classified
- 12 Apr 2021
- Reported
- 21 Apr 2021
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Potency & Assay
- Device type
- Spinal Implants
Timeline
| 16 Mar 2021 | Recall initiated by company | |
| 12 Apr 2021 | Classified by FDA | +27 days |
| 21 Apr 2021 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Interference condition between the 10mm Drill Guide Inserter and the inner shaft that mates with the inserters. As a result, inner shafts do not mate flush with the collet inserters, and splaying of the collet. If implant falls into the spinal cord may cause a spinal cord injury or Incomplete seating of the inserter to the implant could result in misalignment of the screws with the inserter screw guide, potentially causing malposition of the screws into the adjacent vertebral bodies
Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.
Product as listed by the FDA
Chesapeake Cervical-Ti 10mm Drill Guide Inserter- a multi-screw construct providing stability to the anterior column (cervical spine) Catalog number: 3608-90033
Where it was distributed
PA
Questions about this recall
Why was it recalled?
Interference condition between the 10mm Drill Guide Inserter and the inner shaft that mates with the inserters. As a result, inner shafts do not mate flush with the collet inserters, and splaying of the collet. If implant falls into the spinal cord may cause a spinal cord injury or Incomplete seating of the inserter to the implant could result in misalignment of the screws with the inserter screw guide, potentially causing malposition of the screws into the adjacent vertebral bodies
Which lots are affected?
Lot Number: MNAJ GTIN: 10888857082571
Where was it sold?
PA
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from K2M
All 39| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Yukon Straight Rod recalled by K2M | K2M | Other | 6 states |
| Device | CLASS II | Everest MI XT Outer Dilator recalled by K2M | K2M | Other | Nationwide |
| Device | CLASS II | Everest MI XT Inner Dilator recalled by K2M | K2M | Other | Nationwide |
| Device | CLASS II | Chesapeake Anterior Lumbar (AL) Removal Tool recalled by K2M | K2M | Other | 9 states |
| Device | CLASS II | Cascadia AN Interbody Convex 10x22x14mm recalled by K2M | K2M | Labeling Errors | Nationwide |
Other Spinal Implants recalls
All 1,161| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline surgical convenience kits labeled as: Spinal CDS recalled | Medline Industries | Packaging Defects | Nationwide |
| Device | CLASS II | Medline surgical convenience kits labeled as recalled | Medline Industries | Packaging Defects | Nationwide |
| Device | CLASS II | Extended Polyaxial Head, Attached Gimbal recalled | Seaspine Orthopedics | Other | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Anterior Spine Pack recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Neuro; Neuro; Crani; Ortho Spine Pack; Bmhm Neuro Pack recalled | Avid Medical | Sterility | Nationwide |