CLASS II ONGOING Device recall · Reported 4 Feb 2026 · FDA Enforcement Report
Chest Drainage Units and Accessories recalled by Cardinal Health 200
Cardinal Health 200 is recalling Chest Drainage Units and Accessories, distributed nationwide in the US. The recall was initiated on 26 Dec 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1203-2026
- FDA event ID
- 98228
- Recalling firm
- Cardinal Health 200, Waukegan, IL
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 26 Dec 2025
- FDA classified
- 28 Jan 2026
- Reported
- 4 Feb 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Other Medical Devices
Timeline
| 26 Dec 2025 | Recall initiated by company | |
| 28 Jan 2026 | Classified by FDA | +33 days |
| 4 Feb 2026 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The firm is updating the IFU to clarify the intended patient population for the identified chest drainage units and accessories as adults, 18 years and older. When used on infants, the CDU may not demonstrate tidaling/bubbling, which may confound the clinician's assessment of device functionality and lead to inadequate/inappropriate treatment/therapy, delay to treatment/therapy, and prolonged hospitalization.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Chest Drainage Units and Accessories: Product Code Description 1. 8888571562 Cardinal Health Sentinel Seal CDU 2. 8888571513 Cardinal Health Sentinel Seal CDU, Dual Drain 3. 8888571489 Cardinal Health Sentinel Seal CDU, with Easy Change Connector 4. 8888571370 Cardinal Health Altitude CDU, Dry Suction, with Easy Change Connector 5. 8888571299 Cardinal Health Aqua-Seal CDU, Wet Suction 6. 8888571406 Cardinal Health Aqua-Seal CDU, Wet Suction, Dual Drain 7. 8888571315 Cardinal Health Aqua-Seal CDU, Wet Suction, with Easy Change Connector 8. 8884713100 Cardinal Health Thora-Seal Basic CDU, One Chamber, 2000 mL 9. 1814713105 OUS¿Only - Cardinal Health Thora-Seal Basic CDU, One Chamber, 2000 mL (Bottle only) 10. 1180571570 OUS¿Only - Cardinal Health Thora-Seal CDU, Two Chamber System, 2600 mL 11. 8884713308 Cardinal Health Thora-Seal CDU, Three Chamber System, 2500 mL 12. 8884713900 Cardinal Health Thora-Seal Replacement Collection Chamber, 2500 mL
Where it was distributed
Worldwide distribution - US Nationwide and the countries of Brazil, Canada, Chile, Colombia, Dominican Republic, and Mexico.
Questions about this recall
Why was it recalled?
The firm is updating the IFU to clarify the intended patient population for the identified chest drainage units and accessories as adults, 18 years and older. When used on infants, the CDU may not demonstrate tidaling/bubbling, which may confound the clinician's assessment of device functionality and lead to inadequate/inappropriate treatment/therapy, delay to treatment/therapy, and prolonged hospitalization.
Which lots are affected?
All units affected by IFU update. UDI-DI: 1. 8888571562 Cardinal Health Sentinel Seal CDU 50192253003091 (CS) 10192253003093 (EA) 2. 8888571513 Cardinal Health Sentinel Seal CDU, Dual Drain 50192253003084 (CS) 10192253003086 (EA) 3. 8888571489 Cardinal Health Sentinel Seal CDU, with Easy Change Connector 50192253003077 (CS) 10192253003079 (EA) 4. 8888571370 Cardinal Health Altitude CDU, Dry Suction, with Easy Change Connector 50192253003039 (CS) 10192253003031 (EA) 5. 8888571299 Cardinal Health Aqua-Seal CDU, Wet Suction 50192253003015 (CS) 10192253003017 (EA) 6. 8888571406 Cardinal Health Aqua-Seal CDU, Wet Suction, Dual Drain 50192253003060 (CS) 10192253003062 (EA) 7. 8888571315 Cardinal Health Aqua-Seal CDU, Wet Suction, with Easy Change Connector 50192253003022 (CS) 10192253003024 (EA) 8. 8884713100 Cardinal Health Thora-Seal Basic CDU, One Chamber, 2000 mL 50192253002643 (CS) 10192253002645 (EA) 9. 1814713105 OUS¿Only - Cardinal Health Thora-Seal Basic CDU, One Chamber, 2000 mL (Bottle only) 50192253002308 (CS) 10192253002300 (CS) 10. 1180571570 OUS¿Only - Cardinal Health Thora-Seal CDU, Two Chamber System, 2600 mL 50192253002292 (CS) 10192253002294 (EA) 11. 8884713308 Cardinal Health Thora-Seal CDU, Three Chamber System, 2500 mL 50192253002650 (CS) 10192253002652 (EA) 12. 8884713900 Cardinal Health Thora-Seal Replacement Collection Chamber, 2500 mL 50192253002667 (CS) 10192253002669 (EA)
Where was it sold?
Worldwide distribution - US Nationwide and the countries of Brazil, Canada, Chile, Colombia, Dominican Republic, and Mexico.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Cardinal Health 200
All 283| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Presource Packs Containing CliniClean Pericare Wipes within recalled | Cardinal Health 200 | Other | Nationwide |
| Device | CLASS II | Presource Packs Containing CliniClean Pericare Wipes within recalled | Cardinal Health 200 | Other | Nationwide |
| Device | CLASS II | Presource Packs Containing CliniClean Pericare Wipes within recalled | Cardinal Health 200 | Other | Nationwide |
| Device | CLASS II | Presource Packs Containing CliniClean Pericare Wipes within recalled | Cardinal Health 200 | Other | Nationwide |
| Device | CLASS II | Presource Packs Containing CliniClean Pericare Wipes within recalled | Cardinal Health 200 | Other | Nationwide |
Other Other Medical Devices recalls
All 13,778| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Citric Complete Dry Citric Acid Concentrate recalled | Nipro Renal Soultions USA | Other | Nationwide |
| Device | CLASS I | Kit: Epidural/Pain TQC 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | Halyard IV Start Kit Non-Sterile recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS I | Kit: Single Shot Epidural 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | AMS Convenience Kit: Dsaek Pack - RX recalled | Windstone Medical Packaging | Device Malfunction | Nationwide |