CLASS II ONGOING Device recall · Reported 19 Oct 2022 · FDA Enforcement Report

Clip Covid Rapid Antigen Test Kit recalled by Luminostics

Luminostics is recalling Clip Covid Rapid Antigen Test Kit, distributed nationwide in the US. The recall was initiated on 2 Sep 2022 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLots: CC-10094-A, CC-10115-A, CC-10123-A, CC-10129-A, CC-10136-A, CC-10144-A, CC-10148-A, CC-10166-A, CC-10168-A, CC-10168-B, CC-10189-A, CC-10193-A, CC-10300-A, CC-10328-A, CC-10329-A, CC-10330-A, CC-10361-A, CC-10364-A, CC-10365-A, CC-10368-A, CC-10369-A, CC-10395-A, CC-10396-A, CC-10397-A, CC-10398-A, CC-10399-A, CC-10439-A, CC-10440-A, CC-10440-B, CC-10441-A, CC-10441-B, CC-10609-A, CC-10609-B, CC-10611-A, CC-10613-A, CC-10615-A, CC-10654-A, CC-10667-A, CC-10710-A, CC-10710-B, CC-10712-A, CC-10714-A, CC-10714-B, CC-10725-A, CC-10727-A, CC-10729-A, CC-10766-A, CC-10768-A, CC-10809-S, CC-10811-S, CC-10874-A, CC-10874-B, CC-10874-C, CC-50002-S, CC-50003-S, CC-50004-S Expiration Dates: expiry.clipcovid.com
Quantity recalled10037 boxes of 25

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0080-2023
FDA event ID
90852
Recalling firm
Luminostics, Fremont, CA
Classification
Class II
Status
Ongoing
Recall initiated
2 Sep 2022
FDA classified
12 Oct 2022
Reported
19 Oct 2022
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Potency & Assay

Timeline

2 Sep 2022Recall initiated by company
12 Oct 2022Classified by FDA+40 days
19 Oct 2022Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

COVID test kit shelf life was periodically extended, as real-time stability test timepoints passed, without FDA concurrence.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

Clip COVID Rapid Antigen Test Kit (25-pack), Catalog Number: MDL002

Where it was distributed

US nationwide distribution

Nationwide

Questions about this recall

Is Clip Covid Rapid Antigen Test Kit recalled?

Yes. Luminostics recalled Clip Covid Rapid Antigen Test Kit on 2 Sep 2022. The recall is Class II and its status is Ongoing. Recall number Z-0080-2023.

Why was it recalled?

COVID test kit shelf life was periodically extended, as real-time stability test timepoints passed, without FDA concurrence.

Which lots are affected?

Lots: CC-10094-A, CC-10115-A, CC-10123-A, CC-10129-A, CC-10136-A, CC-10144-A, CC-10148-A, CC-10166-A, CC-10168-A, CC-10168-B, CC-10189-A, CC-10193-A, CC-10300-A, CC-10328-A, CC-10329-A, CC-10330-A, CC-10361-A, CC-10364-A, CC-10365-A, CC-10368-A, CC-10369-A, CC-10395-A, CC-10396-A, CC-10397-A, CC-10398-A, CC-10399-A, CC-10439-A, CC-10440-A, CC-10440-B, CC-10441-A, CC-10441-B, CC-10609-A, CC-10609-B, CC-10611-A, CC-10613-A, CC-10615-A, CC-10654-A, CC-10667-A, CC-10710-A, CC-10710-B, CC-10712-A, CC-10714-A, CC-10714-B, CC-10725-A, CC-10727-A, CC-10729-A, CC-10766-A, CC-10768-A, CC-10809-S, CC-10811-S, CC-10874-A, CC-10874-B, CC-10874-C, CC-50002-S, CC-50003-S, CC-50004-S Expiration Dates: expiry.clipcovid.com

Where was it sold?

US nationwide distribution

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other Test Kits & Diagnostics recalls