CLASS II Device recall · Reported 23 Apr 2014 · FDA Enforcement Report

CMF Battery Powered Driver recalled by Synthes

Synthes is recalling Synthes CMF Battery Powered Driver, distributed nationwide in the US. The recall was initiated on 30 Jan 2014 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesPart no. 305.840, with lot nos.: 5680577, 5701123, 5713215, 5763130, 5783316, 5787131, 5828296, 6124867, 6136125, 6150803, 6167974, 6167975, 6218358, 6222322, 6258395, 6365430, 6365431, 6390225, 6425013, 6456634, 6487035, 6556704, 6624880, 6650136, 6668870, 6743859, 6844968, 6926708, 6983799, 7008819, 7038706, 7221395, 7138319.
Quantity recalled20081

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1456-2014
FDA event ID
67499
Recalling firm
Synthes, West Chester, PA
Classification
Class II
Status
Terminated
Recall initiated
30 Jan 2014
FDA classified
14 Apr 2014
Reported
23 Apr 2014
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

30 Jan 2014Recall initiated by company
14 Apr 2014Classified by FDA+74 days
23 Apr 2014Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The CMF Battery Powered Driver graphic case contains outlines of the Battery Powered Driver, taps, screwdriver blades, and a screw holding sleeve. However, the device is not intended to be used in conjunction with a tap for tapping holes in a patient's skull, upper and/or lower jaw. Therefore some outlines in the graphic case suggest a potential use of this device that is considered off-label.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Synthes CMF Battery Powered Driver, Part no. 305.840. Used to place screws in the cranialfacial bones, and to drill and/or tap holes to prepare cranialfacial bones for screw insertion.

Where it was distributed

Nationwide Distribution

Nationwide

Questions about this recall

Is CMF Battery Powered Driver recalled?

Yes. Synthes recalled CMF Battery Powered Driver on 30 Jan 2014. The recall is Class II and its status is Terminated. Recall number Z-1456-2014.

Why was it recalled?

The CMF Battery Powered Driver graphic case contains outlines of the Battery Powered Driver, taps, screwdriver blades, and a screw holding sleeve. However, the device is not intended to be used in conjunction with a tap for tapping holes in a patient's skull, upper and/or lower jaw. Therefore some outlines in the graphic case suggest a potential use of this device that is considered off-label.

Which lots are affected?

Part no. 305.840, with lot nos.: 5680577, 5701123, 5713215, 5763130, 5783316, 5787131, 5828296, 6124867, 6136125, 6150803, 6167974, 6167975, 6218358, 6222322, 6258395, 6365430, 6365431, 6390225, 6425013, 6456634, 6487035, 6556704, 6624880, 6650136, 6668870, 6743859, 6844968, 6926708, 6983799, 7008819, 7038706, 7221395, 7138319.

Where was it sold?

Nationwide Distribution

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Synthes

All 43

Other Other Medical Devices recalls

All 13,778