CLASS II Device recall · Reported 13 Jan 2016 · FDA Enforcement Report
CNSystems Task Force Monitor 3040i, plethysmograph, impedance recalled
Partners In Medicine is recalling CNSystems Task Force Monitor 3040i, plethysmograph, impedance, distributed nationwide in the US. The recall was initiated on 1 Dec 2015 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0572-2016
- FDA event ID
- 72710
- Recalling firm
- Partners In Medicine, Saint Louis, MO
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 1 Dec 2015
- FDA classified
- 4 Jan 2016
- Reported
- 13 Jan 2016
- Type
- Voluntary: Firm initiated
- Customers notified by
- Reason category
- Other Reasons
- Device type
- Surgical Instruments
Timeline
| 1 Dec 2015 | Recall initiated by company | |
| 4 Jan 2016 | Classified by FDA | +34 days |
| 13 Jan 2016 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Device non-conformity which may lead to an incorrect oscillometric measurement of upper arm blood pressure values in some subjects - mainly in tall and slim persons (overestimation of systolic blood pressure, underestimation of diastolic blood pressure).
Product as listed by the FDA
CNSystems TASK FORCE MONITOR 3040i, plethysmograph, impedance. Noninvasively measure and display patient's hemodynamic parameters time synchronized using Impedance Cardiography (ICG), Electrocardiography (ECG), oscillometric Blood Pressure (oscBP) and continuous Blood Pressure (contBP). Furthermore Sympathic Tone and Parasympathic Tone are calculated by power spectral analysis of ECG (heart rate variability) and continuous blood pressure (blood pressure variability) as well as the Baroreceptor Reflex Sensitivity (BRRS). For use by medically trained personnel. It has been designed for diagnoses aiding and must not be used for vital sign monitoring of critically ill patients under absence of additional, suitable self-monitoring devices or medical trained personnel. The device measures continuously the subject's hemodynamic parameters without reporting any diagnosis. It is the physician's responsibility to make proper judgment based on these parameters.
Where it was distributed
Nationwide Distribution including MI and CA.
Questions about this recall
Why was it recalled?
Device non-conformity which may lead to an incorrect oscillometric measurement of upper arm blood pressure values in some subjects - mainly in tall and slim persons (overestimation of systolic blood pressure, underestimation of diastolic blood pressure).
Which lots are affected?
Serial numbers: 003040i-2014-161-014-GG-0000, 003040i-2015-101-003-GG-0000, 003040i-2014-161-001-GG-0000, 003040i-2014-161-002-GG-0000, 003040i-2014-161-003-GG-0000, 003040i-2014-161-004-GG-0000, 003040i-2014-161-005-GG-0000, 003040i-2014-161-006-GG-0000, 003040i-2014-161-007-GG-0000, 003040i-2014-161-008-GG-0000, 003040i-2014-161-009-GG-0000, 003040i-2014-161-010-GG-0000, 003040i-2014-161-011-GG-0000, 003040i-2014-161-012-GG-0000, 003040i-2014-161-013-GG-0000, 003040i-2014-161-015-GG-0000, 003040i-2014-161-016-GG-0000, 003040i-2014-161-017-GG-0000, 003040i-2014-161-018-GG-0000, 003040i-2014-161-019-GG-0000, 003040i-2014-161-020-GG-0000, 003040i-2014-161-021-GG-0000, 003040i-2014-161-022-GG-0000, 003040i-2014-161-023-GG-0000, 003040i-2014-161-024-GG-0000, 003040i-2014-161-025-GG-0000, 003040i-2015-101-001-GG-0000, and 003040i-2015-101-002-GG-0000
Where was it sold?
Nationwide Distribution including MI and CA.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other Surgical Instruments recalls
All 3,809| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Convenience Kits: IV Start KIT recalled by Medline Industries | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep Clear recalled | Medline Industries | Packaging Defects | Nationwide |
| Device | CLASS II | Medline convenience kits labeled as: Open Heart Pack, Medline Kit recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline convenience kits labeled as: GYN Robotic CDS, Medline Kit recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Linvatec Genesys CrossFT Suture Anchor with Disposable Driver recalled | ConMed | Other | Nationwide |