CLASS II Device recall · Reported 3 Jul 2013 · FDA Enforcement Report

Coiled Tube Infusors (Portable Elastomeric Infusion Systems) recalled

Baxter Healthcare is recalling Coiled Tube Infusors (Portable Elastomeric Infusion Systems), distributed nationwide in the US. The recall was initiated on 7 Jun 2013 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesProduct Code: 2C1073KJP Lot Numbers: 08M002, 08N007, 08N029, 08N030, 08N034, 08N069, 08N070, 09A001
Quantity recalled601,230 total

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1609-2013
FDA event ID
65364
Recalling firm
Baxter Healthcare, Round Lake, IL
Classification
Class II
Status
Terminated
Recall initiated
7 Jun 2013
FDA classified
27 Jun 2013
Reported
3 Jul 2013
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Packaging Defects
Device type
Infusion Pumps

Timeline

7 Jun 2013Recall initiated by company
27 Jun 2013Classified by FDA+20 days
3 Jul 2013Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Due to an increase in complaints for leaks at the distal male Luer and Luer cap. Baxter implemented enhancements to the blue winged cap to improve the seal between the cap and the Luer.

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

Brand Name: Coiled Tube Infusors (Portable Elastomeric Infusion Systems). Baxter's Infusor SV Elastomeric Infusion Device Device Listing: D020871 Baxter's lnfusors SV Elastomeric Infusion Device is a single-use, disposable elastomeric infusion pump indicated tor patients requiring slow, continuous intravenous, intra-arterial, subcutaneous, or epidural administration of medications at a constant flow rate.

Where it was distributed

Worldwide Distribution - USA (nationwide), Puerto Rico, and Internationally to Brazil, Colombia, Panama, Mexico, Singapore Japan, Australia, New Zealand, Hong Kong, India, Thailand, Taiwan, and Korea.

Nationwide Puerto Rico

Questions about this recall

Why was it recalled?

Due to an increase in complaints for leaks at the distal male Luer and Luer cap. Baxter implemented enhancements to the blue winged cap to improve the seal between the cap and the Luer.

Which lots are affected?

Product Code: 2C1073KJP Lot Numbers: 08M002, 08N007, 08N029, 08N030, 08N034, 08N069, 08N070, 09A001

Where was it sold?

Worldwide Distribution - USA (nationwide), Puerto Rico, and Internationally to Brazil, Colombia, Panama, Mexico, Singapore Japan, Australia, New Zealand, Hong Kong, India, Thailand, Taiwan, and Korea.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 65364

More recalls from Baxter Healthcare

All 573

Other Infusion Pumps recalls

All 546