CLASS II ONGOING Device recall · Reported 26 Mar 2025 · FDA Enforcement Report

Comprehensive Shoulder Stem recalled by Zimmer Orthopedics

Zimmer Orthopedics Manfacturing Limited-galway is recalling Comprehensive Shoulder Stem, distributed nationwide in the US. The recall was initiated on 3 Mar 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesItem No. 113605, UDI-DI 00887868267550 Lots 66173214, 66173215; Item No. 113606, UDI-DI 00887868267567 Lots 66173216, 66173218, 66173219; Item No. 113607, UDI-DI 00887868267574 Lots 66173220, 66173221; Item No. 113627, UDI-DI 00887868267741 Lots 65901697, 65909480, 65915809
Quantity recalled90 US; 136 OUS

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1403-2025
FDA event ID
96398
Classification
Class II
Status
Ongoing
Recall initiated
3 Mar 2025
FDA classified
20 Mar 2025
Reported
26 Mar 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

3 Mar 2025Recall initiated by company
20 Mar 2025Classified by FDA+17 days
26 Mar 2025Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Five complaints received where surgeon was unable to remove and/or connect the inserter with the stem during surgery due to excess material remaining within the taper. Issue may lead to clinically significant extension of surgery to find a replacement.

Product as listed by the FDA

Comprehensive Shoulder Stem, Comprehensive Micro Stem, Comprehensive Mini Stem; uncemented shoulder prosthesis, Model Nos. 113605, 113606, 113607, 113627

Where it was distributed

US distribution nationwide. International distribution to Dominican Republic, Canada, India, Australia, Japan, EMEA, Korea, Taiwan, and Hong Kong.

Nationwide

Questions about this recall

Why was it recalled?

Five complaints received where surgeon was unable to remove and/or connect the inserter with the stem during surgery due to excess material remaining within the taper. Issue may lead to clinically significant extension of surgery to find a replacement.

Which lots are affected?

Item No. 113605, UDI-DI 00887868267550 Lots 66173214, 66173215; Item No. 113606, UDI-DI 00887868267567 Lots 66173216, 66173218, 66173219; Item No. 113607, UDI-DI 00887868267574 Lots 66173220, 66173221; Item No. 113627, UDI-DI 00887868267741 Lots 65901697, 65909480, 65915809

Where was it sold?

US distribution nationwide. International distribution to Dominican Republic, Canada, India, Australia, Japan, EMEA, Korea, Taiwan, and Hong Kong.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.